The focus of our project as a whole is to generate a chelating agent (either mono-acetylated chitobiose or mono-acetylated chitotriose) that can remove heavy metals from wastewater, aiding water treatment facilities and reducing the amount of harmful compounds in natural water bodies. Whilst synthetic biology can be used to create these agents, they must be installed into an overall piece of hardware to actually work. As such, the hardware must meet the following goals:
Loading the 3D model…
The Fusion 360 assembly, consisting of the reusable frame and disposable cartridge (connected) beside the reusable plunger.
Drag to turn it · scroll to zoom
3D viewer: three.js r160, MIT licence
Throughout the project’s development, the application of our hardware changed considerably.
Originally, the focus of our project was the development of a large-scale piece of equipment that could clean heavy metals out of industrial and mining wastewater. There were two main ideas for how this could be implemented, though they were not fully explored:
In the former design, reaction between the heavy metals and the chelating agent would create a complex of higher molecular weight that could theoretically be easier to filter. In reality, however, the change in molecular weight is insignificant, being from around 340 Da to around 570 Da. This would require nanofiltration [1], a unit operation that cannot feasibly be done at this scale.
In the latter design, the wastewater would flow continuously through a chromatography column that retains the heavy metal residues. This is significantly more feasible, though it was not fully explored due to a change in application focus.
Early into the project, it became known to the team that new EU legislation would be coming into effect that would drastically alter how pharmaceutical waste was handled. This requires pharmaceutical companies to pay 80% of the end-of-life cleanup costs of their products [2], which could potentially sway such companies from supplying their product at all. As proof of concept, cisplatin, a platinum-based chemotherapeutic drug, was selected as the target compound. Cisplatin is important for treatment of solid tumours, though it is cytotoxic and largely excreted by patients through their urine [3].
It must be noted that the potential uses of the chelating agent can still be applied to other areas, such as industrial wastewater treatment (for example using the chromatography idea mentioned above), and should not be limited to just cisplatin removal, as advised by stakeholder Dr Anne Woolridge. In this case, however, the wastewater being treated has changed origin, from industrial activity to hospital sewer systems.
The design ideas for this application are identical to the ones mentioned previously for industrial wastewater treatment, though they would be at a much smaller scale and placed within hospital waste systems, directly before discharge into sewer systems. Whilst this location was deemed optimal by medical stakeholders (Dr Kumud Kantilal and Anne O’Connor and Professor Louis Grandjean), those same stakeholders also informed us of the difficulty in changing the infrastructure of hospitals for a cause that does not save significant amounts of money. As such, it was decided that the application would not be a large-scale piece of equipment integrated into infrastructure, but rather a small device used by patients directly, ensuring that the amount of cisplatin flushed into toilets was reduced altogether.
Through both research and stakeholder interaction, many different conceptual designs were thought through. The same story is told cycle by cycle on our Engineering Success page.
In the first design, the hardware was effectively just a funnel that could be attached to the toilet. The chelating agent coating the material inside the funnel would bind to the cisplatin, trapping it within the hardware. Meanwhile the treated urine would pass through the other side and into the toilet.
After meeting with Professor Laurence Lovat, an enormous amount of insight into the technicalities of our project was provided. Some of the key points included factoring in both chelation contact time and the safety of handling the device (for example, whether there is a risk of contaminated urine splashing anywhere). Of immediate priority was the former point.
In theory, the chelation step can only occur once cisplatin has undergone aquation (has lost its chloride ions) [4]. After further research, it was discovered that the aquation step is quite long, in the magnitude of several hours [5]. There is not much literature on this, but what was found was initially taken as fact. It was also found that a higher pH is preferable for coordination to occur [6], so addition of a basic reagent was also highly considered.
A funnel coated inside with a chelating agent sits on the toilet seat. Urine runs straight through it into the toilet, and cisplatin binds.
After the first set of feedback, three separate designs were made:
A mesh bag containing the chelating agent is dropped into the toilet and later removed and disposed of.
A disposable plastic bag internally coated with the ligand. It is urinated into and, after being left to enable chelation outside the toilet, is emptied back into the toilet bowl. The bag is sent for incineration.
A filter cup into which urine collected from patients on the cancer ward is emptied. It feeds a chelation tank, producing treated urine and a contaminated single-use bag for incineration.
All three designs now include sodium carbonate to increase the urine’s pH. They also all require a delay for adequate chelation to occur. These designs were presented to the Cancer Patient and Public Advisory Group (CPPAG) to get insight from the patients who would be using our hardware.
Almost immediately, the first and third designs were rejected. The first was rejected due to the inability to flush or defecate whilst aquation and chelation are occurring, as well as the need to actively take out the teabag, which can be considered unsafe and is too much of a burden for patients and carers. The third was rejected by Isla-Jayne Burzio, the chair of CPPAG, who stated that it was highly unlikely for there to be a large filter cup for constant addition of patient urine within the ward.
On the other hand, the second design was praised yet given critical feedback. The intention was for patients to easily remove the bag and store it with the urine inside whilst chelation occurs. Once the recommended time period was over, the bag could be emptied into the toilet, leaving behind most of the cisplatin inside for disposal. The only negative criticism given for this design was the fact that the bag would be stored with urine inside, which was considered disgusting. The patients’ suggestion was to have an absorbent material within the bag to soak up the liquid, removing the more unpleasant element of the setup.
The next design was similar to the “clip-on bag”, but the chelating agent is coated on a cellulose sponge within the bag. In addition to removing the large volume of liquid urine, the sponge also provides an extremely high surface area for chelation to occur.
In more detail, this design consists of two parts:
This was deemed the final design for a while, so the bulk of the life cycle assessment and techno-economic model outputs are actually based on this design.
During write-up of the literature, it was decided as a group that there was insufficient evidence to support the long aquation time of cisplatin stated in one piece of literature [5]. Additionally, as the hardware serves as a device for all heavy metal pharmaceutical manufacture, the much faster chelation kinetics of other heavy metals were assumed instead [7].
This meant that several aspects of the previous design could be scaled down:
The sponge has been retained so that urine doesn’t just pass straight through (it still needs a short contact time), whilst also enabling the surface area of chelating agent to remain high.
Ultimately, the design is as follows:
Using Fusion 360, the designs of these parts were visualised (Figures 6 and 7).
Consists of the frame that attaches to the toilet. It has a lever used to squash the cartridge and the sponge within.
Consists of the cartridge that is placed into the frame. It contains an empty space for the sponge. Note that the cartridge is compressible.
This too was considered a final design, and so was used to update the techno-economic model for greater accuracy. That said, this wasn’t the true final design, as an issue was discovered: too much force was required for a chemotherapy patient to use the lever at its current height, and making it any longer would cause its own problems.
The final design abandons the lever, instead adding a third part: the plunger. This is also reusable, but is separate from the frame and is stored separately, similarly to a toilet brush. Firstly, the cartridge is placed into the frame. Ten minutes after urinating, the plunger can then be used to squash the sponge by pushing against the cartridge’s press cap, releasing the treated urine into the toilet. The plunger can then be used to hook two tabs on the cartridge, removing it from the toilet easily. The cartridge should be allowed to drip and then be disposed of in an appropriate bin bag.
The measurements needed to fit the new design have been summarised in Tables 1 to 6.
Table 1: Sourced design parameters
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Urine void volume | 245 mL mean, 425 mL max | Human physiology | [8] [9] |
| Toilet envelope | 350–380 mm bowl, 400–450 mm seat, 100–180 mm fixings | UK standard | [10] [11] |
| Mounting precedent | 330 × 203 mm (0.61 mm wall); 68 g polypropylene rim lip and lowered seat | Clinical nun’s cap | [12] [13] |
| Pulp wall range | 3–5 mm for transfer-moulded trays | Process limit | [14] |
| Toilet seat hole | 213 × 285 mm | UK standard | [10] [11] |
Table 2: Assumed design parameters
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Dwell time | 10 minutes (drain in place) | Aquation and coordination kinetics | [7] |
| Cartridge volume | At least one void | Short dwell time | – |
| Cavity target | 600 mL | Above max void volume | – |
Table 3: Parameters impacting shape design
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Sponge capillary limit | Pore < 0.23 mm at 0.050 N/m surface tension of urine | Jurin’s law | [15] [16] |
| Cavity height | 45 mm | Shallow enough for a sponge to retain fluid | – |
| Press pressure | 15–30 kPa | Assumed | – |
| Cavity width | 112 mm floor; 116.7 mm mouth | A wider cavity requires more force; a shorter cavity can cause technical problems for users | – |
| Cavity | 589 mL total; assumed ~10% sponge fraction, usable volume of ~530 mL | Cavity heights and widths | – |
Table 4: Disposable cartridge design
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Tray wall | 3 mm, 3° draft | Pulp wall range | [17] |
| Base / rim / flange | 117.7 / 122.7 / 152.2 mm | Derived from cavity volume, tray wall thickness and draft angle. Also adds 8.6 mm width going up the cartridge (tapered design) | – |
| Body at flange | 15 mm tall, nests in frame cone, 45° | Derived from tray wall thickness, draft angle and base/rim/flange dimensions | – |
| Cartridge funnel | 201 × 274 mm; 120.5 mm tall | Frame height | – |
| Press cap | Pulp dish, 111.3 mm square base, 15 mm tall, 3 mm wall (25 g) | Cartridge dimensions | – |
| Tabs | 2 × (44 × 32 mm), 22 × 27 mm window, 5 mm bar | Typical | – |
| Floor holes | 36 × (0.6 mm), 17 mm pitch | Typical / assumed | – |
| Sponge mass | 30 g | Cavity volume | [18] |
| Tray and skirt masses | 147 g | Derived from base/rim/flange dimensions and tray wall thickness | [19] |
| Sodium carbonate | 3 g | pH buffer | [20] [21] |
| Total pulp mass | 171 g | – | – |
Note: masses do not go into the overall design of the hardware, but they do feed directly into the techno-economic and life cycle assessment models.
Table 5: Reusable frame design
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Funnel throat | 155.2 mm inner width, 45° cone to 125.2 mm | 5 mm clearance of body at flange | – |
| Funnel mouth | 199 × 269 mm | Seat hole minus 5 mm per side | – |
| Funnel depth / slope | 55 mm, 68° walls, 44° from horizontal | Steep enough not to pool | – |
| Funnel mass | 103 g | – | – |
| Rim bands and ribs | 260 × 320 × 2.5 mm (65 g) | – | – |
| Stack below toilet rim | 132 mm | Must clear bowl water | – |
| Sleeve and rib grid | 5 × 5 open rib grid, 3 × 14 mm, ~77% open (101 g) | Must carry 600 N without a plate | – |
Note: masses do not go into the overall design of the hardware, but they do feed directly into the techno-economic and life cycle assessment models.
Table 6: Mechanism design (squeezing the sponge)
| Parameter | Value | Set by | Source |
|---|---|---|---|
| Press force | 250 N; limit 290 N for patients | Medical standard | [22] |
| Press plate | 104.1 mm², 5 mm width, ribs and hub (89 g) | Must fit within the cap, pressing down on it without going beyond its perimeter | – |
| Shaft | 32 mm outer diameter, 26 mm inner diameter, 450 mm overall (75 g) | Typical | – |
| Grip and palm cap | 38 mm diameter, 120 mm length (68 g) | Medical equipment standard | [22] |
| Hooks | 2 × spring fingers and barbs (10 g) | Standard | – |
| Polypropylene mass total | 532 g | – | – |
Note: masses do not go into the overall design of the hardware, but they do feed directly into the techno-economic and life cycle assessment models.
The designs listed in Tables 1 to 6 have been used to create the overarching mechanical design, put together in Fusion 360.
The complete frame placed into the toilet prior to urination.
The instrument used to squeeze the cartridge and lift it out of the frame without exposing the user to urine.
Functionally a disposable, smaller version of the frame. It contains a sponge (not shown) and is placed within the frame before use.
Urine doesn’t make contact with the reusable parts at any point except steps 5 and 6. Here, it should only make contact with the tiny grates at the bottom.