Hardware

CAPTURE-Screen is a single-cuvette fluorescence reader for GFP reporter cultures. It measures their green fluorescence directly at the bench, beside the incubator, without waiting for a shared plate reader, and we built it for our own AHL biosensor. We developed two versions of the device: a 3D-printed version and a laser-cut version, the laser-cut one in its own enclosure and the 3D-printed one on an open base plate under an opaque shield. The following sections describe the features shared by both designs, followed by a detailed overview of each version.

Background

A GFP reporter is read as green light from a liquid culture, usually on a shared plate reader. Ours reports AHL, the quorum-sensing signal, and its assay set what this reader had to do: a small sample, read every hour, a comparison with a plate reader, and an instrument cheap enough to rebuild.

200 µL
of sfGFP culture per reaction The reporter is superfolder GFP.
Every hour
for six hours, at 37 °C A shared plate reader is booked, scheduled and in one building. Either build can sit next to the incubator.
485 / 510
nm, the plate-reader benchmark Ours excites at 470 nm, and its geometry, drive current, integration time and dark baseline are ours to state.
21–65×
cheaper than an open-source plate reader At most: US$161 for the 3D-printed build and US$54 for the laser-cut one, by our own costing, against under US$3,500 for the plate reader of Szymula and colleagues5. Anyone with the source files can rebuild ours.

The need is not only ours. Green fluorescence from a GFP reporter in a liquid sample is among the commonest readings in synthetic biology: one interlaboratory study had 72 iGEM teams measure it in E. coli, on plate readers and flow cytometers6. Cell-free reactions, developed as portable sensors7, hold no cells at all. The reader is meant for samples like these: liquid, and not too dense. We tested the effect of sample density on the 3D-printed build. Both builds have read sfGFP in E. coli TOP10 culture, and both have read our AHL sensor strain without seeing its signal: the laser-cut build could not tell it from cells without sfGFP (the laser-cut build’s test), and the 3D-printed build read zero on its emission channel.

Measurement principle

Both readers keep the blue that excites the sample apart from the green it gives off in two ways: by angle, and by colour. Neither uses an emission filter: the sorting by colour is the sensor's own.

optical core — plan view, 5 px = 1 mm along X 10 mm UVette chamber, enlarged 2 mm 10 mm 90° LED holder Cree, 470 nm baffle two disk slots, empty sample UVette, 200 µL light trap angled wedge, Ø8.0 × 27 mm deep emission filter — position reserved, part not yet specified AS7341 detector, on SEN0365 board +X +Y AS7341 ten channels, one read F1415 F2445 F3480 F4515 F5555 F6590 F7630 F8680 Clear NIR 470 nm excitation sfGFP emission 1 2 3
  1. 470 nm in One blue LED lights the sample, 200 µL in a UVette4, across the cuvette's 2 mm path.
  2. 90° to the beam The detector looks at the sample from the side, along the 10 mm path. Blue that passes straight through ends in a light trap.
  3. Ten channels out The AS73411 sorts what arrives by colour, so a read gives the shape of the light and not one number. Excitation lands in F2 and F3, sfGFP2 emission in F4 and F5.
Drawn to scale as it is built in the 3D-printed build, which reads F4; the laser-cut build reads F5 over F3.
The light path inside the planned enclosure schematic
drag to rotate · scroll to zoom · shift-drag to pan

See through

View

Layers

Selection

Every part and its X extent
Parts in the cutaway and their X extents
Part X range (mm)

This view places the 3D-printed build's optical core in the two-layer laser-cut enclosure as it was drawn, optics above and electronics below. No reading on this page was taken inside it. It was built, but not adopted: of the 11 counts that leak in room light, 4 come through the mechanism and 7 straight down the open cuvette slot, and an opaque shield over that slot brings the dark read to exactly zero (Light shielding). We chose the shield over a second enclosure. It is a schematic built from the optical coordinates, not from the CAD files: the printed parts themselves are under the 3D-printed build.

The two builds

We built two readers on this principle. They share the light path and the spectral sensing described above. Each has its own mechanism, modelling and development history.

Choose a build to see it.

Counts from the two readers are not interchangeable: they run different firmware and integrate for different times.

Measured
Version 2 of the 3D-printed build, seen from above: a black printed optical core on a base plate, with a clear UVette standing in its seat.

3D-printed build

A 3D-printed optical core on a base plate, with the electronics on a breadboard beside it.

Run from
Serial commands, or LasReader (our software)
Version
Second, built 2026-10-03
Data
Three mixing sessions, a dilution series of GFP culture, a session beside a plate reader, one on AHL plates and one on how dense a culture can be. The Model page works through the mixing sessions
See the 3D-printed buildShowing below
Enclosed
The laser-cut build with its lid off: a plywood enclosure, bare outside and black inside, holding three black chambers on the deck. A USB cable runs into the reader on the left, and a small module is mounted on the front panel.

Laser-cut build

A second reader on the same principle, made from laser-cut parts in its own enclosure.

Run from
LasReader (our software)
Interface
OLED display and a button
Data
The following sections
See the laser-cut buildShowing below

The 3D-printed build

It tells a culture with GFP from one without, using no emission filter, and its reading rises with the amount of GFP culture. It is quick to load, steady when the tube is put back, cheap to build, and made of parts that can be changed one at a time.

86 vs 0
counts, with GFP and without Two cultures of similar density, read the same day with no emission filter. Dense suspensions read above zero.
3–5 s
to load Placing the cuvette, timed. Version 1 took 30–40 s.
0.76%
variation when put back Standard deviation between five re-insertions of one tube, at about 50 counts.
US$161
to build Our estimate of the cost as built, with the printing bought from a shop.

Not yet It has not resolved AHL induction, and it has not been tested with an emission filter or calibrated against a fluorescence standard.

How it got here: the same loop, twice.

Version 1

  1. Build A printed optical core.
  2. Test Three mixing sessions, and use by the wet lab.
  3. Learn Too many steps to place the cuvette.

Version 2

  1. Build A seat that takes the cuvette in one step.
  2. Test A dilution series, a plate reader beside it, and dense suspensions.
  3. Learn Dense suspensions read above zero without GFP.

Next turn An emission filter, in the drawer already on the reader.

The instrument

A 3D-printed optical core on a base plate, with the electronics on a breadboard beside it.

It has been built twice. Version 2, in use since 2026-10-03, is the one the view opens on; version 1 took the earlier readings. Each printed part is drawn from its own STL file, where the CAD assembly puts it. Pull it apart, switch versions, or lay one module's parts side by side.

The optical core, from its CAD files

Every printed part, with the mesh of its STL file · grid 10 mm

Every file 53 STL files, 66 pieces, both versions

iGEM's file server takes neither STL nor ZIP files, so the 53 STL files are kept as one glTF model, triangle for triangle: capture-screen-printed-parts-20261004.glb (1.7 MB). With JavaScript on, each file name below saves that part as an STL again.

Excitation end 14 files, 15 pieces

  • Excitation cradle ×1 36.5 × 38 × 20 mm both versions Carries the LED holder and the baffle on the base plate. excitation_cradle_v0_7.stl 195 KB
  • LED holder body ×1 20.4 × 20.4 × 21.3 mm version 1 Split in two at its clip slot when the supports were removed, and was taped back together. capture_screen_led_holder_v0_1_body.stl 259 KB
  • LED holder body, version 2 ×1 21.6 × 21.6 × 21.3 mm version 2 Replaces the old body and its clip. It has no clip slot. led_holder_body_v0_3.stl 292 KB
  • LED retainer clip ×1 14.5 × 11.5 × 1 mm version 1 Specified, but it never arrived. capture_screen_led_holder_v0_1_retainer_clip.stl 35 KB
  • Aperture baffle body ×1 16 × 16 × 15 mm version 1 Broke at both disk slots when the supports were removed, and was taped back together in three pieces. capture_screen_aperture_baffle_body.stl 247 KB
  • Aperture baffle body, version 2 ×1 15 × 16 × 13 mm version 2 Drawn to print on its side. aperture_baffle_body_v0_2.stl 146 KB
  • LED-holder cap ×1 6 × 30.6 × 12.45 mm version 1 Specified, but it never arrived. excitation_cap_LED_holder_v0_7.stl 79 KB
  • LED-holder cap, version 2 ×1 6 × 30.6 × 12.45 mm version 2 The same shape, turned over so it prints flat face down. excitation_cap_LED_holder_v0_7_roof_down.stl 93 KB
  • Baffle cap ×1 6 × 26.4 × 10.25 mm both versions excitation_cap_baffle_v0_7.stl 80 KB
  • Excitation locator key ×2 8 × 4.8 × 4 mm both versions excitation_locator_key_print_two_v0_7.stl 0.7 KB
  • Aperture disk, 1.5 mm ×1 12 × 12 × 1 mm both versions No disk was fitted for any measurement. aperture_disk_1p5mm.stl 111 KB
  • Aperture disk, 2.5 mm ×1 12 × 12 × 1 mm both versions No disk was fitted for any measurement. aperture_disk_2p5mm.stl 111 KB
  • Aperture disk, 4 mm ×1 12 × 12 × 1 mm both versions No disk was fitted for any measurement. aperture_disk_4mm.stl 99 KB
  • Aperture disk, 6 mm ×1 12 × 12 × 1 mm both versions No disk was fitted for any measurement. aperture_disk_6mm.stl 99 KB

Cuvette seat 11 files, 13 pieces

  • UVette holder body ×1 36 × 36 × 25.5 mm version 1 uvette_holder_common_body.stl 308 KB
  • UVette holder body, version 2 ×1 36 × 36 × 25.5 mm version 2 One piece in place of the old body, both gates, the collar, the gauge and the leaf. Eight ribs locate the cuvette. uvette_holder_body_v0_3.stl 215 KB
  • UVette riser ×1 43 × 42.5 × 13 mm version 1 uvette_datum_riser_v0_7.stl 172 KB
  • UVette seat riser, version 2 ×1 43 × 42.5 × 13 mm version 2 uvette_seat_riser_v0_2.stl 207 KB
  • Orientation gate A ×1 32 × 32 × 2 mm version 1 uvette_orientation_gate_A.stl 101 KB
  • Orientation gate B ×1 32 × 32 × 2 mm version 1 uvette_orientation_gate_B.stl 101 KB
  • Index collar ×1 24 × 20 × 5 mm version 1 uvette_index_collar.stl 5.0 KB
  • Collar setting gauge ×1 7 × 3 × 8.5 mm version 1 uvette_collar_setting_gauge.stl 1.4 KB
  • Preload leaf ×1 1.7 × 10 × 12 mm version 1 One was specified and received. The holder design called for two. uvette_preload_leaf.stl 2.2 KB
  • Hold-down clip, X side ×2 6.4 × 8.6 × 4.9 mm both versions uvette_x_hold_down_clip_print_two_v0_7.stl 2.6 KB
  • Hold-down clip, Y side ×2 6.4 × 7.6 × 4.9 mm both versions uvette_y_hold_down_clip_print_two_v0_7.stl 2.6 KB

Detection side 17 files, 20 pieces

  • Detector holder body ×1 40 × 30 × 40 mm both versions as7341_detector_body.stl 399 KB
  • Sensor board carrier ×1 32 × 3.2 × 32 mm both versions as7341_pcb_carrier.stl 151 KB
  • Sensor board retainer ×1 25.4 × 1.5 × 16.8 mm both versions as7341_pcb_retainer.stl 86 KB
  • Detector locator frame ×1 55 × 34.15 × 6 mm both versions The one part redrawn for version 1's print, after the print shop's slicer showed thin walls at the pin sockets. detector_locator_frame_v0_7.stl 83 KB
  • Detector locator pin ×2 3.2 × 3.2 × 4 mm both versions detector_locator_pin_print_two_v0_7.stl 20 KB
  • Detector side clamp ×2 10 × 10 × 6 mm both versions detector_side_clamp_print_two_v0_7.stl 22 KB
  • Filter cassette body ×1 22 × 22 × 8 mm both versions common_cassette_body.stl 72 KB
  • Cassette insert, open ×1 24.25 × 20.5 × 5.5 mm both versions blank_open_control_insert.stl 53 KB
  • Cassette insert, round 12.5 mm: base ×1 24.25 × 20.5 × 5.5 mm both versions round_12p5_insert_base.stl 126 KB
  • Cassette insert, round 12.5 mm: cap ×1 15 × 15 × 1.3 mm both versions round_12p5_insert_cap.stl 49 KB
  • Cassette insert, round 12.5 mm: 1.5 mm spacer for a 2 mm filter ×1 12.5 × 12.5 × 1.5 mm both versions round_12p5_2mm_spacer.stl 49 KB
  • Cassette insert, square 6 mm: base ×1 24.25 × 20.5 × 5.5 mm both versions square_6x6_insert_base.stl 55 KB
  • Cassette insert, square 6 mm: cap ×1 10 × 10 × 1.3 mm both versions square_6x6_insert_cap.stl 1.6 KB
  • Cassette clamp bar ×2 36.4 × 2.5 × 6.4 mm both versions replacement_filter_cassette_clamp_bar_print_two_v0_7.stl 42 KB
  • Alignment shim, 1.5 mm ×1 20 × 2.5 × 1.5 mm both versions alignment_shim_1p5mm.stl 0.7 KB
  • Alignment shim, 2.0 mm ×1 20 × 2.5 × 2 mm both versions alignment_shim_2p0mm.stl 0.7 KB
  • Alignment shim, 2.5 mm ×1 20 × 2.5 × 2.5 mm both versions alignment_shim_2p5mm.stl 0.7 KB

Light trap 3 files, 4 pieces

  • Light trap, angled wedge ×1 27 × 40 × 18 mm both versions optical_light_trap_angled_wedge_v0_1.stl 133 KB
  • Light-trap riser ×1 29 × 42 × 13 mm both versions light_trap_adapter_riser_v0_7.stl 126 KB
  • Light-trap clamp plate ×2 10 × 8.7 × 2 mm both versions light_trap_clamp_plate_print_two_v0_7.stl 37 KB

Base plate 3 files, 9 pieces

  • Base plate ×1 120 × 110 × 4 mm both versions optical_core_base_plate_v0_7.stl 563 KB
  • Bench foot ×4 10 × 10 × 4.5 mm both versions bench_foot_print_four_v0_7.stl 42 KB
  • Locator pin ×4 4.8 × 4.8 × 4 mm both versions locator_pin_print_four_v0_7.stl 20 KB

Fit coupons 5 files, 5 pieces

  • UVette fit coupon ×1 70 × 21.9 × 26.7 mm version 2 Four cuvette pockets, 0.05 to 0.20 mm clear on each side. Print it before the holder body. uvette_fit_coupon_v0_2.stl 293 KB
  • LED fit coupon ×1 44 × 12 × 11.6 mm version 2 Four LED sockets, 0 to 0.30 mm tight. Print it before the LED holder body. led_fit_coupon_v0_2.stl 692 KB
  • Interface coupon, base ×1 110 × 60 × 10 mm version 1 Specified for printing first, to check holes, pins and sliding fits. fit_coupon_interface_base_v0_7.stl 583 KB
  • Planar clearance coupon, base ×1 82 × 30 × 8 mm version 1 Specified for printing first, to check holes, pins and sliding fits. fit_coupon_planar_clearance_base_v0_7.stl 212 KB
  • Coupon slider ×1 10 × 18 × 9 mm version 1 Specified for printing first, to check holes, pins and sliding fits. fit_coupon_slider_v0_7.stl 1.4 KB

Bought, not printed 3 parts, drawn as plain shapes in the assembly

  • Eppendorf UVette ×1 12.5 × 12.5 × 36 mm both versions The sample cuvette. Drawn as a plain block of the size the CAD gives it.
  • Cree 5 mm LED ×1 8.8 × 5.8 × 5.8 mm both versions The excitation LED. Drawn as a plain cylinder of the size the CAD gives it.
  • AS7341 sensor board ×1 18 × 2.4 × 14 mm both versions The detector board. Drawn as a plain block of the size the CAD gives it.
Placed as in the CAD assemblies. Screws, the LED's leads and the optional detector mask are left out. Specified in matte black PLA; drawn in grey so the shapes can be read.

Design advantages

The light is sorted without a filter, the seat places the cuvette, every read checks itself, and the parts change one at a time.

  • No emission filter 86 vs 0 86 with GFP OD600 0.86 0 without OD600 0.72

    The detector looks at the sample from 90°, and its own colour bands keep the blue apart from the green. The two cultures read alike beside the source, 21.4 and 21.7 counts on F2; only the green differs. Dense suspensions are where this stops.

    Measured, one session · every channel of both reads
  • A seat that places the cuvette 1 step, 3–5 s the tube's own mean −5% 0 +5%

    The cuvette drops between eight ribs onto four corner tabs, which set its position and height. One tube, put back five times: 0.76% between insertions.

    By design; timed and measured on version 2
  • A read that checks itself dark, lit, dark
    Shielded 0560 kept
    Leaking 1 cycle void

    Every flash sits between two reads with the LED off. Shielded, those are exactly zero, so one that is not voids its cycle: eight in version 1's logs, one in the dilution series.

    In the firmware; counts from our logs
  • Parts that change one at a time 5 parts, 1 resistor

    That is all version 2 changed. The base plate, the detection side, the light trap and the firmware stayed as they were.

    What can be changed, drawn below.

    Done once, for version 2

Built to be changed

What has been changed, what is on the reader but not yet used, and what has not been tried.

Base plate · unchanged Excitation end changed for version 2 The LED · holder drawn for a 5 mm round LED another LED not tried LED current · one resistor changed, 5.55 to about 10 mA Cuvette seat changed for version 2 Light trap unchanged Filter drawer · on the reader, empty for a 12.5 mm round or 6 mm square filter Detector The board · in a carrier of its own another board: a new carrier, a new driver not tried Ten channels · all recorded on every read only sfGFP read so far Changed for version 2 On the reader, not yet used Not tried Unchanged
Seen from above, not to scale. The CAD source is parametric and kept module by module in the repository, so one part can be redrawn without the others.

User feedback

Our wet lab used both versions, at least three of its members in all. What they said about the first set the brief for the second. Two external advisors also reviewed the design.

  • Wet-lab members, on version 1 Too many stepsto one step

    They told us Placing the cuvette took too many steps, each one a chance to seat it differently, and a measurement took a long time.

    What followed Version 2's seat takes the cuvette in one step and sets its height and position itself.

  • Wet-lab members, on version 2, in the dilution series 30–40 sto 3–5 s

    Time to place the cuvette, version 1 against version 2, timed. Each range runs from quicker hands to slower ones.

    They told us Convenient and quick: the cuvette only has to be put in, which leaves little room for operator error. Seeing the taped repairs replaced by new parts, they trusted the reader more. The arrow engraved on the seat also helped: it shows how the cuvette should face.

    What followed They made that session's five re-insertions themselves, and every reading stayed within one count of 50.

  • Two wet-lab members, each running the reader from the Run card

    They told us The card was clear to follow. One of them went through it in order and met no problem. The other met REJECTED sample_id, and the remedy the card gives for it was enough to deal with it.

    What followed The Run card stays as they used it, with its list of what to do when the reader says something else.

  • Two external advisors, reviewing the design

    They told us Prof. Tsung-Lin Liu asked what mattered most in the hardware, and suggested trying a wider range of sensors. Prof. Nan-Si Li placed the reader as one step of a path from sample to answer, behind a front-end unit that prepares the specimen.

    What followed We set three priorities for the hardware: low cost, speed and convenience. We did not add a second sensor: the AS7341 already sorts the emission by colour, so another would have meant buying a filter to do the same job. The front-end unit is not built (Human Practices).

Use turned up seven more problems, from room light getting in to a detector that lost its settings.

What version 2 changed What the operating log added

From version 1 to version 2

One step to load, a steadier reading, more light with the same zero, and fainter samples in reach.

All four come from version 2, fitted on 2026-10-03. We redrew five printed parts around what the wet lab had found slow, and around what had broken, gone missing or sat loose in version 1; and we changed one resistor to bring the LED up to its design current.

Version 1 from above. On the left the LED end is wrapped in black tape. In the middle a square collar sits around the UVette in its holder. The light-trap block is on the right and the detector block behind.
Version 1 The LED end is taped where its holder and the baffle broke, and a collar sits around the cuvette.
Version 2 from above, facing the same way. On the left the LED holder sits under an arched cap held by two screws. In the middle a clear UVette stands in a plain square seat. The light-trap block is on the right and the detector block behind.
Version 2 One seat takes the cuvette directly, and the LED holder is held down by a printed cap.
  • Placing the cuvette 4 stepsto 1
    Version 1
    1. Cuvette in
    2. Collar on
    3. Set its height
    4. Seat the holder
    ready
    Version 2
    1. Cuvette in
    ready
    By design; version 1's steps in photographs below
  • The same tube, put back 3.6–6.3%to 0.76% the tube's own mean Version 1 · test 1 Version 1 · test 2 Version 2 · test 3 −5% 0 +5%

    Each dot shows one insertion relative to the tube's own mean; the percentage is the standard deviation divided by that mean. Version 1's tubes were put back by the reader's builder, version 2's by wet-lab members.

    Observed, in different sessions
  • More light, the same zero 5.55to 10 mA 7.9 5.55 mA 18.3 10 mA 0 5.55 mA 0 10 mA beside the source (F2) emission channel (F4)

    An empty cuvette scatters about twice as much. The emission channel stays at zero.

    Measured
  • Fainter samples 0to 3 counts
    Version 1 000 zero in 3 of 4 tubes
    Version 2 3.33.03.0 above zero in all 3

    GFP culture at 0.19 on our concentration index on version 2; on version 1, at 0.19 or less. The index is the stock's displayed optical density (OD) times its dilution.

    Observed, with different culture batches

Part by part The re-insertion numbers The LED current

Placing the cuvette, step by step

The first card, played out in photographs: version 1's four steps, taken from frames of a video we filmed, and version 2's one.

Version 1 four steps
  1. The parts laid out on a table: a clear cuvette held in one hand, a square black holder with an empty pocket, a small C-shaped gauge and a square collar.

    The parts The cuvette, its holder, a collar and a small gauge.

  2. The cuvette now stands in the holder's pocket.

    1 · Cuvette in The cuvette goes into the holder's pocket.

  3. The collar sits around the cuvette on top of the holder, while two hands pick up the small gauge.

    2 · Collar on The collar is pressed down around the cuvette.

  4. The gauge is in place: a small black piece now sits against the left side of the cuvette, and the hands have just let go of the holder.

    3 · Set its height The gauge sets how high the collar sits.

  5. The holder, with the cuvette and collar in it, now sits on the reader's base plate between the LED end and the light trap.

    4 · Seat the holder The holder goes onto the reader. Ready.

Version 2 one step
  1. Version 2 from directly above, its square seat empty, with a clear cuvette lying beside the reader.

    The seat It is part of the reader. Nothing else is fitted.

  2. The same view with the cuvette standing in the seat.

    1 · Cuvette in The seat sets its height and position. Ready.

Functional testing

It tells a GFP culture from one without, and the reading rises with the amount of GFP culture; a plate reader beside it ranked the same tubes the same way. Without GFP it read zero in liquid culture, and a few counts in dense suspensions. It has not yet resolved AHL induction.

The GFP cultures read in this section are a test strain: E. coli TOP10 with a plasmid that makes sfGFP without induction. They stood in for our AHL biosensor, which we expected to be about as bright once induced. That induction has not been shown conclusively: in liquid culture both of its constructs fluoresced above an empty-vector control, with no uninduced culture beside them to compare (Engineering).

The dilution series (version 2)

Every tube without GFP read zero. From 25 to 200 µL the readings lie close to a proportional line; below 25 µL they fall below it.

Version 2 read one batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls.

0.1 1 10 100 F4, counts 6.25 12.5 25 50 100 200 GFP culture in the 200 µL tube, µL in proportion curve fitted afterwards 98 counts
Each dot is one tube; both axes are logarithmic. The grey line is proportional; the curve was fitted to all twelve tubes afterwards, and one 12.5 µL tube was read again. The four tubes without GFP read zero, which a logarithmic axis cannot show.

The smallest amount read above zero on every cycle was 12.5 µL of GFP culture in 200 µL, here and in the session beside a plate reader, on a different batch. At half that, 4 of 18 cycles read above zero.

Not yet A limit of detection: this was one session, on one batch of culture.

The dilution series, tube by tube What the wet lab said

Beside a plate reader (version 2)

Read beside a plate reader on the same tubes, it ranked them the same way, and over the three denser tubes its reading per unit of OD held within 10%.

We diluted one batch of the sfGFP test culture to five strengths, from 6.25% to 100%, and read every tube on our reader and on a SpectraMax M3 plate reader, which also measured its OD600. As the culture got stronger, our green reading rose from 2.44 to 86.44 counts, keeping close step with the M3’s OD600 (r2 = 0.993 across the five tubes). A culture without GFP, at an OD600 of 0.72, read zero on ours and close to background on the M3. Even the thinnest tube, at an OD600 of 0.07, gave a reading on every cycle; we did not try anything thinner.

0 25 50 75 100 F4 on our reader, counts sfGFP culture culture without GFP the three denser tubes: 104 counts per unit OD 0.0 0.2 0.4 0.6 0.8 1.0 OD600 on the plate reader, blank taken off
Each dot is one tube: F4 on our reader, the mean of nine cycles, against OD600 on the M3. The dashed line is the mean ratio of the three denser tubes. It is not a calibration line.

Not yet This was one culture on one day, and it is not a calibration. Whether the two instruments’ fluorescence readings are in proportion is still open.

Beside the plate reader, tube by tube

Three mixing sessions (version 1)

Zero on the emission channel for all 27 readings without GFP, above zero for 35 of the 38 tubes with it.

Three mixing sessions each combined a GFP stock, a non-GFP stock and medium at set fractions into 200 µL tubes.

Without GFP: 27 readings, none above zero With GFP: 35 of 38 tubes above zero Mixing 1 Mixing 2 Mixing 3 above zero on the emission channel zero
One mark to a reading. The three GFP tubes that read zero are all from mixing session 2, whose stock was thin, and all three sit below 0.2 on our concentration index.

In the two sessions with enough signal, the reading rose in proportion to the GFP culture. How it scales, and why the three sessions differ, is worked through on the Model page.

The three mixing sessions in numbers

Light shielding

Under a shield the dark read is exactly zero, so the zeros above are readings, not corrections.

F4 with the LED off, empty cuvette, counts Shielded 0 Room light 11 4 through the mechanism 7 down the open cuvette

Under the shield, every channel reads zero while the LED is off, even at 25 times the usual integration time. In room light the leak does not scale with integration time, so subtracting a dark reading would not remove it.

With the LED on, plain medium has read zero on the emission channel every time on version 2: eight blank readings across four sessions, 60 valid cycles in all.

The operating rule: Measure in a darkened room or under an opaque shield. Because a shielded dark read is exactly zero, a dark read that is not zero marks that cycle as invalid: it is dropped and its raw row kept. The firmware applies this rule itself: it voided eight cycles in version 1's logs and one in version 2's dilution series.

Where it stops: dense suspensions (version 2)

In a dense suspension the reader gives a few counts with no GFP there. A culture without the reporter read zero while the scatter channel stayed under about 33 counts, and no longer did from 46.

We met this on suspensions scraped from AHL plates, the samples we were given for an AHL test, in the same E. coli TOP10. The wet lab grew culture on plates containing AHL at five concentrations, 0.22 to 1.1 µM by its own estimate, scraped each plate into suspension and gave us 200 µL of it. By eye it judged every plate induced but the 1.1 µM one. The reading followed how much culture was in the cuvette, not that judgement.

Two days later we took a culture without the reporter gene, scraped from one plate, diluted it in LB and read each dilution: the emission channel for what the culture itself gives, and the scatter channel of the same read beside it. LB alone read zero. Reading zero means below what this reader can register, not that nothing is there.

0 2 4 6 F4, counts culture without the reporter the same, added afterwards AHL plate, judged induced AHL session: not induced, or no reporter every cycle zero some cycles 1 or 2 every cycle above zero 0 20 40 60 80 100 F3, counts: the blue light the suspension scatters With sfGFP in the cuvette, F4 A dense suspension, from the 1.1 µM plate 5.22 The same suspension, 180 µL, with 20 µL of sfGFP culture 23.56 sfGFP culture, 100 µL, in 100 µL of medium 66.56
Above, each mark is one reading of one cuvette, the mean of nine cycles, placed by the scatter channel of that same read. Circles are the culture without the reporter; the rings are six readings added after the first result had been seen. Squares are the suspensions of the AHL plate session; induced and not induced are the wet lab's judgement by eye. Below, that session's sfGFP controls.

Where the two days overlap, the readings from plates judged induced lie within about a tenth of a count of a curve drawn through the readings without the reporter. sfGFP added to a dense suspension was still read: 23.56 counts, against 5.22 for that suspension alone.

Not yet This was one culture on one day, so the scatter figures are not a specification.

Using version 1 changed the seat. These two sessions point to the next change: an emission filter.

  • After the AHL plate session and the density session, on version 2

    What followed Every read already reports the scatter channel beside the emission channel, so the two can be read together. In the density session a culture without the reporter gave a few counts of F4 once F3 passed about 46, and none under about 33: at that scatter, a few counts alone are not a sign of GFP.

    Still open An emission filter. The reader has a drawer for one in front of the detector, and it has run empty. Neither session could say whether the reading without GFP is scattered blue or green from the sample itself; a filter that passes green and blocks blue is the test that could tell them apart.

The AHL plate session, cup by cup The density session, reading by reading The background, modelled

Interactive bench

Every channel of a read, for the states we measured. Choose a test and throw its switches.

Light-path bench

Top view, schematic · both versions, named by session

Y = 0 X = 0 +X excitation axis +Y detection axis room light down the open +Z slot LED holder aperture baffle UVette holder filter cassette AS7341 holder angled-wedge light trap

The channel readout needs the page script. Without it, the light path is still drawn, and the sessions the readings come from are named earlier in this section.

Counts are raw sensor values. F4, outlined in green, is the emission channel: the sensor's 515 nm band, where GFP emits.

Construction and operation

Print it, wire it, and run it under a shield. Build version 2.

  1. Print

    Version 2 is 38 printed pieces, specified in matte black PLA; the full set of 52 adds loose shims, filter inserts, aperture disks and two fit coupons. Each can be saved as an STL file from the 3D view. The five new parts print without supports; on another printer, print the two fit coupons first.

    Parts and files Assembly, step by step

  2. Wire

    A 2N7000 MOSFET switches the 470 nm LED at about 10 mA through a 217 Ω resistor, and the AS7341 detector sits on the ESP32's I²C bus. Keep the gate pulldown: without it the LED can light while the controller boots.

    The circuit

  3. Run

    Any serial terminal at 115 200 baud, one command per line, under an opaque shield. This is one session from start to finish; the reader's answers are from the log of the dilution series on version 2, 2026-10-05.

    Once, before the first session

    The sketch ships with its LED gate off, and built that way the LED never lights. Build it with the gate on, then upload it.

    arduino-cli compile --fqbn esp32:esp32:esp32doit-devkit-v1 --build-property "compiler.cpp.extra_flags=-DCAPTURE_LED_HARDWARE_ENABLE=1" firmware/esp32_fluorescence_reader
    1. Shield on Box over the reader. Reset the controller once, now, and not again.
      id
      # hw_enable(compile)1
      # sensor_ready      1
      # settings_readback 1

      Go on only if all three read 1.

    2. Empty cuvette, LED not armed The check that the shield is on.
      run

      dark 0LED off 0dark 0NO_SIGNAL;LED_DISABLED

      It stops after one cycle, as it should. All 36 values must be 0.

    3. Same cuvette, LED armed The day's empty reading, not a reference across days.
      arm ARM-LED-P1run
      # LED ARMED

      dark 0F4 0, 0, 0dark 0OK

    4. Medium blank, 200 µL The day's zero.
      sample LOW_BLK_1run
      # sample_id=LOW_BLK_1

      dark 0F4 0, 0, 0dark 0OK

    5. Each sample, 200 µL No bubble where the beam passes, 8.5 mm above the floor. Repeat for every tube.
      sample LOW_D_1run

      dark 0F4 56, 57, 55dark 0OK

      Three cycles of dark, lit, dark. This tube: 100 µL of GFP culture in its 200 µL.

    6. Done A finished run does not disarm by itself.
      disarm
      # LED DISARMED

    dark is all twelve values of a read, before and after each flash: the sensor's ten channels, with Clear and NIR read twice. F4 is the three cycles with the LED on. Each cycle prints one CSV row of 66 fields: keep the terminal's whole log. Ours was kept by a script that stamps and stores every line; it is in the repository as firmware/host/serial_read_tool.ps1.

    When the reader says something else

    AMBIENT_PRESENT
    A dark read was not zero: light is getting in. Void the cycle, fix the shield, read again. Do not subtract it, and leave the voided row in the file. Seen 2026-10-05, at 6.25 µL: one channel read 1 before one flash of nine. Read again, all nine cycles were clean.
    # REFUSED: ENABLE is not in a defined acquisition state
    The detector lost power or was reset, and id shows settings_readback 0. Reset the controller once. If that fails, unplug the USB for five seconds, then check the detector's wires: SDA on GPIO21, SCL on GPIO22, and 3V3. Seen three times on version 2, on 2026-10-03 and 2026-10-09. A reset cleared it each time.
    NO_SIGNAL;LED_DISABLED
    The LED did not light. fault=NOT_ARMED: arm and run again. fault=HW_ENABLE_COMPILED_OFF: the sketch was built with its gate off.
    # REJECTED sample_id: longer than 24 chars
    A name takes up to 24 letters, digits and . _ - +. A rejected name leaves the last one in force: read the echo before run.
    SATURATED
    A channel reached full scale, 65 535. Dilute the sample.
    flag=OK with F4 at 0
    A valid reading. Keep it: medium has read zero every time.

    Acquisition timing The dilution series, tube by tube

A session, in photographs

Two gloved hands at the reader: one lowers a clear cuvette into the square black seat, the other steadies the base plate. The LED holder is in front, the breadboard and its wires behind.
1 · Cuvette in The cuvette goes straight into the seat.
A person in a lab coat and gloves sits at a bench and lowers a cuvette into the reader with one hand while holding the base with the other. An open laptop stands behind it, connected by a cable.
At the bench A team member seats the cuvette, with the laptop beside the reader.
An upturned tray of moulded pulp covers the reader on the bench, a cable running out from under it to an open laptop whose screen shows a serial monitor.
2 · Shield on The shield goes over the reader before any read.
A laptop screen showing a serial monitor at 115200 baud. The firmware prints its identity, with hw_enable, sensor_ready and settings_readback at 1, then two refusal lines, LED ARMED, a sample name, a header row, three rows of comma-separated values and the line timepoint complete.
3 · Read One read as the serial monitor shows it: the reader's identity, then one row to a cycle. Two mistyped commands are refused above it.
A hand lifts the pulp tray to show what is under it: the black printed reader on its base plate with a cuvette in the seat, and beside it a breadboard carrying the controller, its indicator lights on.
The shield, lifted Under it, the reader and the breadboard it is wired to.
The bench from above and to one side: an open laptop showing the serial monitor, a USB cable, and to its right the breadboard and the printed reader with a cuvette in the seat. A rack with culture tubes stands behind.
The whole bench A laptop, one cable and the reader.

The CAD sources and exports, the firmware and the parts list are published under the MIT licence in the team's repository on iGEM GitLab: capture-screen-hardware.

A second sketch runs the same reader from LasReader, our software, over Wi-Fi, in the protocol the laser-cut build uses: firmware/capture_screen_3d. It drives the LED gate from GPIO26, not GPIO25: move the gate wire, or change LED_PIN. Run this way, beside the laser-cut build, it read our AHL sensor strain: F4 was zero throughout, and the samples were not dense enough to scatter into it. Those reads were not kept, so no numbers are given here; every reading on this page with a number was taken with the serial firmware.

Not yet The assembly order was checked by the reader's builder. Nobody has yet built a second reader from it.

Cost

It cost us about US$161 to build, most of it printing bought from a shop. With a printer of your own it is about US$38 and 180 g of PLA.

Cost to build one, US$, to scale Our laser-cut build $54 Our 3D-printed build $161 An open-source plate reader under $3,500 Where the $161 goes $122 printing, bought from a shop $37 boards and wiring $1 LED, switch and resistors
This compares price alone. The plate reader is the open-source design by Szymula and colleagues5, which its authors describe as a fraction of the cost of a commercial instrument. It reads a whole plate and has been validated, while ours reads a single cuvette, so the two do different jobs. Our costs are our own estimates.

The cost, line by line

Data and documentation

Every table behind the sections above, and what a second team needs to build and run the reader.

Measurements and use

The experiments, by name 14 experiments and their dates

The page calls each experiment by a name. This is the list: what was done, on which version of the reader, and the date its records carry.

The experiments on the 3D-printed build, by name
ExperimentReaderWhat was doneDate
First lightVersion 1The LED on a paper scatterer at the sample position, and with nothing in the beam.2026-08-25
Room-light testVersion 1An empty cuvette with the LED off, in room light and shielded.2026-08-27
First fluorescenceVersion 1A culture reported to carry sfGFP and its control, twelve cycles each.2026-09-03 to 09-04
Stability readVersion 1One tube left in its seat and read 20 times over 5 minutes.2026-09-03
Re-insertion test 1Version 1One culture, four insertions, one of them a fresh draw.2026-09-09
Re-insertion test 2Version 1The corresponding culture five days later, read the same way.2026-09-14
Mixing session 1Version 1A GFP stock, a non-GFP stock and medium mixed at set fractions into 200 µL tubes.2026-09-16
Mixing session 2Version 1The same mixing, with a thinner stock.2026-09-21
Mixing session 3Version 1The same three liquids, laid out to test a recipe difference and proportionality.2026-09-29
LED current testVersion 2One empty cuvette, read before and after the current-limit resistor was changed.2026-10-03
Dilution seriesVersion 2One batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls. Re-insertion test 3, one tube put back five times, was part of it.2026-10-05
AHL plate sessionVersion 2Suspensions scraped from plates at five AHL concentrations, with medium, a culture without the reporter and an sfGFP culture as controls. Re-insertion test 4, one tube put back ten times, was part of it.2026-10-09
Density sessionVersion 2A culture without the reporter, two suspensions scraped from one plate, diluted in LB: 21 readings, with LB as the blank.2026-10-11
Plate reader comparisonVersion 2One batch of the sfGFP test culture at five dilutions, with medium and a culture without GFP, read on our reader and on a SpectraMax M3.2026-10-12
Version 1: the three mixing sessions in numbers 2 tables
Does the emission channel separate GFP from no GFP?
Session Readings without GFP F4 above zero Tubes with GFP F4 above zero
Mixing session 11601010
Mixing session 280129
Mixing session 3301616
All three2703835

Samples without GFP read zero on the emission channel every time, including undiluted non-GFP culture at a displayed OD of up to 1.42. The channels beside the excitation band do see those cells, but too variably between cuvettes and sessions to build a turbidity flag on.

How far repeated readings scatter on version 1

How far repeated readings scatter on version 1
What was repeated n Mean F4 Std dev % of mean
One liquid, re-inserted (mixing session 1)36.000.335.6 %
One recipe, prepared three times (mixing session 1)36.890.517.4 %
One recipe, six tubes (mixing session 3)69.331.1011.7 %
One culture, four insertions, one a fresh draw (re-insertion test 1)440.501.233.0 %
One culture, four insertions, one a fresh draw (re-insertion test 2)432.671.835.6 %
Observed scatter, not a specification. Each figure mixes sampling, insertion, cuvette and time, and none of it carries over to low signal. Between re-insertion tests 1 and 2, five days apart, the same sample read 19.3 % lower; the instrument had been moved and the state of the culture is not known, so the cause is not separated.
What the operating log added 7 problems
Problems found in use and what changed
What went wrongSessionWhat changed
An empty cuvette read F4 = 11 under room light.Room-light testAdded the shield box. A dark read that is not zero voids the cycle.
A cuvette's number did not match what was in it, and one record had to be voided.Mixing session 2Confirm what is in the cuvette before starting a run.
The stock's displayed OD did not scale with dilution.Mixing sessions 1 and 3Recommended: read the OD at one fixed dilution, 1:4. Not yet in the operating card.
Washed and reused cuvettes tied “which use of this cuvette” to the order of reading.Mixing session 3Recommended: one cuvette per tube. Not yet in the operating card.
Suspensions scraped from plates were dense enough to raise the emission channel without GFP.AHL plate sessionThe density session found where a culture without the reporter stops reading zero, in the scatter channel. Recommended: check F3 on every sample. Not yet in the operating card.
A read stopped partway: the detector timed out once, and the bus to it failed twice.Density sessionA reset and a second read of the same cuvette. The cause has not been found.
The firmware refused a run because the detector had lost its settings. A reset cleared it each time.Mixing session 3, twice on the day version 2 was fitted, and once at the end of the AHL plate sessionThe Run card now says what to do after a refusal.

The last two rows are the firmware doing its job: when its state is not known it declines to measure. Four refusals point at a loose connection to the detector board. The cause has not been found. In the 28 captures of the dilution series there was no such refusal; the firmware's only stop was an opening check left unarmed on purpose. That is an observation, not a fix.

Version 2: the dilution series, tube by tube 12 tubes, 4 controls

200 µL to a tube. The sensor settings were version 1's. The LED ran through the new resistor, and its current was not measured that day.

F4 on version 2, the dilution series, one batch of culture
GFP culture in the tube Tubes F4, each tube Mean
None: medium alone20, 00
None: non-GFP culture, undiluted20, 00
6.25 µL20.44, 0.110.28
12.5 µL33.33, 3.00, 3.003.11
25 µL212.67, 14.4413.56
50 µL223.00, 26.0024.50
100 µL256.00, 50.3353.17
200 µL, undiluted198.3398.33
Each figure is one tube's mean, over three cycles or, at the three lowest levels, nine. The GFP stock had a displayed OD of 0.77 at 1:4. Two first captures were set aside. One, at 6.25 µL, had a dark read that was not zero, and the tube was read again. The other, at 12.5 µL, read 7.44; the operator reported the stock had not been mixed and redid the tube, a decision taken after the reading was seen. The third 12.5 µL tube and the 200 µL tube were added once the planned readings had been seen.

Every tube without GFP read zero on the emission channel, the non-GFP culture included, as on version 1. From 25 to 200 µL, a range picked after the readings were seen, the seven tubes lie within −8.8 % and +14.5 % of one straight line through the origin.

Below that the reading falls short of the line: 12.5 µL read about half of what the line gives and 6.25 µL about a tenth, with a count of one or more on only 4 of its 18 cycles. That is the shape a floor in the reading would give, but one session does not tell a floor from other causes.

Version 1 had read zero for three of the four tubes at or below 0.19 on our concentration index, the stock's displayed OD scaled by its dilution in the tube, all in mixing session 2; 12.5 µL here is that same 0.19. The index leans on the displayed OD, and this session's stock read 0.77 at 1:4, above the 0.13 to 0.71 of the earlier sessions.

The 12.5 µL figure rests on the redone tube: with the first capture kept in its place, the planned analysis puts that level at 0.82 of proportional instead of 0.48, and the pooled form of that analysis no longer resolves a low-end shortfall. The 6.25 µL readings do not depend on it.

One 100 µL tube was taken out and put back five times, by wet-lab members. The five means ran from 49.33 to 50.33 counts, with a standard deviation of 0.38 counts between them: about the third of a count that a three-cycle mean can resolve. With the seat empty, every channel read zero.

Version 2: beside the plate reader, tube by tube 7 tubes, 2026-10-12

The sfGFP test culture, E. coli TOP10 with a plasmid that makes sfGFP without induction, diluted in medium. On our reader, 200 µL to a cuvette and nine cycles each, at the sensor settings of the dilution series; no tube was read twice. On the SpectraMax M3, OD600 and fluorescence at 485 / 515 nm.

Our reader and the SpectraMax M3 on the same tubes
In the tube F4, ours Cycles OD600, M3 F4 per OD M3 fluorescence, undiluted M3 fluorescence, diluted 20×
Medium, at the start and at the end0, 00–00–0–
Culture without GFP0.000–00.721–249–
sfGFP culture, 6.25 %2.442–30.069358,216not read
sfGFP culture, 12.5 %7.447–80.1256013,249326.5
sfGFP culture, 25 %33.2233–340.30211025,2491,147.6
sfGFP culture, 50 %48.3347–490.475102not obtained1,746.2
sfGFP culture, 100 %86.4484–890.861100not obtained3,802.1
OD600 and the undiluted fluorescence have the medium blank taken off; the diluted reading had no blank and is as read. The two fluorescence columns were read under different conditions and cannot be joined into one.

One question was registered before the session: is our F4 in proportion to the M3's fluorescence, each tube's ratio within 20% of their mean? It could not be judged. The M3 could not read the denser tubes undiluted, so only 6.25%, 12.5% and 25% have a fluorescence reading of the registered kind. Across those three the ratio ran 0.298, 0.562 and 1.316 (×10−3): not in proportion, the thinner tubes reading low on ours, as the dilution series had led us to write down beforehand.

Worked out afterwards: F4 against the M3's OD600 gives r2 = 0.993 over five tubes, against its diluted fluorescence 0.982 over four, against its undiluted fluorescence 0.981 over three, and the tubes rank the same each time. These are a few points of one dilution series, where r2 runs high: they show one trend, not accuracy.

The M3's values were copied down by the operator; there is no export file, the number of wells and the gain were not recorded, and some values have two significant figures. The two instruments excite at different wavelengths, 470 and 485 nm, so only the trend can be compared, not the size of the numbers.

Version 2: the AHL plate session, cup by cup 12 readings, 2026-10-09

200 µL to a cuvette, nine cycles each, at the sensor settings of the dilution series. Sorted by F3. There was one cuvette to a condition and no plate without AHL, and how much was scraped was not measured: the session gives no dose–response and cannot say whether the plates were induced.

F3 and F4 on version 2, the AHL plate session
In the cuvette By eye F3 F4
Medium, at the start and at the end–0, 00, 0
sfGFP culture, 100 µL, in 100 µL of mediumcontrol22.8966.56
1.1 µM plate, the thinner scrapenot induced56.781.00
0.35 µM plate, diluted by halfinduced59.561.11
Culture without the reporter–64.441.78
0.22 µM plateinduced73.782.44
1.1 µM suspension, 180 µL, with 20 µL of sfGFP culturecontrol85.8923.56
0.275 µM plateinduced94.004.44
1.1 µM plate, the denser scrapenot induced97.895.22
0.55 µM plate, first readinginduced102.114.89
0.35 µM plateinduced103.445.22
0.55 µM plate, read againinduced104.445.22

Medium read zero on the emission channel in all 18 cycles. The culture without the reporter did not: 1.78, at a density no earlier session had reached. Whether its cuvette had held culture before was not recorded.

Set side by side after the readings were seen: among the five readings with F3 from 94 to 104, the plate judged not induced read 5.22 and those judged induced 4.44 to 5.22. Among the three with F3 from 57 to 64, induced read 1.11, not induced 1.00, and the culture without the reporter 1.78.

The sfGFP culture, at a quarter of that scatter, read 66.56. Twenty microlitres of it in 180 µL of a dense suspension read 23.56, against 5.22 for that suspension alone: the dense culture did not hide it.

One test was registered before the session: that F4 rises in the order of the AHL concentrations. Read as registered, it does not hold (p = 0.64). Our script first reported that it did (p = 0.05), having used two repeat readings the registration rules out: both were asked for after the first reading was seen, and the 1.1 µM one was a second, denser scrape. Neither outcome bears on AHL. The same test on F3 gives the same p values, so what it ranks is how much was scraped.

Version 2: the density session, reading by reading 21 readings, 2026-10-11

200 µL to a cuvette, nine cycles each, at the sensor settings of the dilution series. A culture without the reporter gene, diluted in LB; per cent is the share of suspension. Sorted by F3.

F3 and F4 on version 2, the density session
In the cuvette F3 Cycles above zero F4
LB, at the start and at the end0, 00 of 180, 0
First suspension, 10 %5.330 of 90.00
First suspension, 10 %, second pass6.440 of 90.00
First suspension, 20 %12.330 of 90.00
First suspension, 20 %, second pass12.440 of 90.00
First suspension, 30 %13.890 of 90.00
First suspension, 30 %, second pass15.110 of 90.00
Second suspension, 30 %, added26.110 of 90.00
Second suspension, 20 %, added, put back26.780 of 90.00
Second suspension, 20 %, added28.780 of 90.00
Second suspension, 30 %, added, read again29.560 of 90.00
Second suspension, 35 %, added, put back32.330 of 90.00
Second suspension, 35 %, added33.110 of 90.00
Second suspension, 40 %41.780 of 90.00
Second pass, the cup taken to be 50 %46.001 of 90.11
Second pass, the cup taken to be 40 %50.785 of 90.56
Second suspension, 50 %56.007 of 90.89
Second suspension, 70 %75.119 of 92.67
Second suspension, 70 %, second pass78.569 of 92.89
First suspension, undiluted82.339 of 92.67
Second suspension, undiluted, read again86.229 of 93.89
Second suspension, undiluted88.679 of 93.89

The first suspension ran out after its 30 per cent tube, and the wet lab scraped a second from the same plate. The two do not line up: 30 per cent of the first scattered 14 to 15 counts, 40 per cent of the second 42 to 51. That is why the series as planned has no tube between F3 15 and 42.

The rule was registered before the session: take the densest tube whose cycles are all zero in both passes, and the thinnest with any cycle above zero. That gives 30 per cent of the first suspension, F3 14.50 over 18 cycles, and 40 per cent of the second, F3 46.28: zero at 41.78 in its first pass, 5 of 9 cycles above zero at 50.78 in its second. The six readings marked added were made after that result was seen, to fill the gap, and are not part of it. In the second pass the cups were told apart by eye, and the 40 and 50 per cent ones may have been exchanged; the table places each reading by its own F3.

Joined into a curve that is only assumed never to fall as scatter rises, these readings give the background of a culture without the reporter at any scatter up to F3 89, and the AHL plate session can be set against it. Its two readings without GFP that fall in that range lie +0.04 and +0.10 counts from the curve; the two from plates judged induced, −0.11 and −0.10. The four densest induced readings, F3 94 to 104, lie beyond the curve: if the background does not fall there, at most 0.56 to 1.33 counts are left. The cuvette with sfGFP added stands 19.76 above it. This was worked out after both sessions, with the two days' counts taken as they are; the curve is one culture on one day, not a correction to subtract. More on the Model page.

One plate, one day, one set of detector settings and no OD: the F3 figures are not a specification and do not carry to another reader or another day. How much GFP signal is left in a sample that thin was not measured.

The firmware stopped three captures partway, once on a detector time-out and twice on a bus error, each before a reading had been shown. The same cuvette was read again after a reset.

Version 1 against version 2: putting one tube back 6 series

Version 1 held the cuvette against one spring leaf where the design called for two, which left it free across about 0.6 mm on one axis. Version 2's eight ribs leave 0.05 mm a side, as drawn. Taking one tube out and putting it back is the test of that. On version 2 the five insertions were made by wet-lab members, not by the member who built the reader.

How far one tube's reading moves
What was done Reader n Mean F4 Spread (SD) % of mean
Left in its seat, read 20 times over 5 minutes (stability read)Version 12021.950.291.3 %
Taken out and put back (re-insertion test 1)Version 1340.331.453.6 %
Taken out and put back (re-insertion test 2)Version 1333.002.086.3 %
Taken out and put back (re-insertion test 3)Version 2549.870.380.76 %
Left in its seat, read 20 times (AHL plate session)Version 22062.951.131.8 %
Taken out and put back (re-insertion test 4)Version 21068.371.842.7 %
Spread is the standard deviation between insertions, each one the mean of three cycles; for the tube left alone it is the standard deviation of its 20 reads. That tube sat at 22 counts, where one step of a three-cycle mean is already 1.5 % of the reading, so its percentage does not compare with the others. Fresh draws of the same culture are left out. On version 1 the reading also climbed with each insertion, by 1.4 and 2.1 % a minute, against 0.6 % on version 2. The days differ in more than the seat: the culture batch, the cuvette, the LED current and how long the tube had stood. In the dilution series the wet lab also used ice, which it had not before; what was chilled, and for how long, was not recorded, and cooling could slow that climb. This is what was observed, not a controlled test of the seat, and the printed rib spacing was not measured.

In counts, putting a tube back on version 1 moved the reading five to seven times as much as leaving it alone. On version 2 it moved it little more than leaving it alone had: 0.38 counts on a mean of 49.87, or 0.76 %.

Test 4, in the AHL plate session, put an sfGFP culture back ten times: 2.7 %, against 1.8 % for the same tube left alone. That culture settles. Left alone its reading fell, and handling stirred it up again, so the wider spread cannot be put down to the seat. A first resting series, which the operator may have bumped, spread 3.3 %.

Version 1 against version 2: the LED current 2 tables

Version 1 ran the LED at 5.553 mA, under the 8 to 12 mA the design aims for. On 2026-10-03 one 220 Ω resistor replaced the series pair, which brings it to about 10 mA. With an empty cuvette, the two channels beside the source then read about twice what they had, and the emission channel still read zero. Two days later the same held with samples in the beam: medium, a non-GFP culture and the empty seat all read zero on the emission channel, while that culture scattered up to 40 counts into the channel next to the source. The scatter off an empty cuvette roughly doubled and the emission channel stayed at zero in every one of these reads, which is what the 90° layout and the light trap are for.

The excitation branch at the two values of R1
QuantityR1 = 434 ΩR1 = 217 Ω
In use2026-08-25 to 2026-10-03: every version 1 readingSince 2026-10-03
Made of218 Ω + 216 Ω in seriesOne 220 Ω resistor, 217 Ω measured
LED current5.553 mA, measured 2026-08-259.2 to 9.9 mA by meter; at most 11.3 mA by calculation
If the LED shorted11.89 mA. The resistor alone holds it under our 15 mA ceilingAbout 24 mA. It no longer does
Empty cuvette, F2 / F3 / F47.89 / 6.89 / 018.33 / 13.33 / 0
The last row is nine consecutive cycles on each side of the change, on version 2, with the same empty cuvette left in its seat in the dark. It is excitation scattered by an empty cuvette, not fluorescence.

Limits It is scattered excitation on an empty cuvette, not a fluorescence result, and it does not show that 10 mA is the right current. Readings at the two resistances cannot be pooled: the empty-cuvette ratio is not a conversion for fluorescence. The meter readings were taken during 0.55 s pulses, so they are a lower estimate.

The same empty cuvette on version 2, as first fitted

The same empty cuvette, LED on, at the version 1 current
Reader Read F2 F3 F4
Version 1Seven sessions, first fluorescence to mixing session 37.33 to 11.896.33 to 10.220
Version 2, as first fittedFour insertions6.585.420
Version 2, after an adjustmentFour insertions7.756.710
The same empty cuvette throughout, LED on, 434 Ω resistor. Each version 2 figure is the mean of four insertion means, and the version 1 range runs over seven session means, counting only cycles the firmware flagged valid, which requires a zero dark read before the flash. F2 and F3 are the two channels either side of the 470 nm source, so they measure how much excitation an empty cuvette scatters toward the detector; F4 is the emission channel. As first fitted the rebuilt reader read low, and after an adjustment it read inside the range version 1 had covered. The adjustment itself was not written down.
Version 1 against version 2: part by part 6 rows
What was wrong in version 1 and what version 2 does about it
In version 1In version 2
In use from 2026-08-25 to 2026-10-03. 57 of its 59 printed pieces arrived, and two of those broke. In use since 2026-10-03, with all five replacement parts fitted.
The LED holder split along its clip slot and the baffle broke at both disk slots as the print supports came off. Both were taped. No clip slot, and a baffle that prints on its side, so no layer loses most of its area. The weakest layer keeps 0.82 of the fullest section within 3 mm of it; version 1 broke at 0.63 and 0.32.
The clip that retains the LED never arrived. Friction and its own leads held it. Three ribs in the holder grip the LED, 0.10 mm tight. No clip.
The cap that clamps the LED holder never arrived. The holder lay in its saddle. The cap is printed, turned over so that it needs no support.
One preload leaf instead of two left the cuvette free on one axis, by up to about 0.6 mm. One body with eight ribs, 0.05 mm clear on each side, replaces six pieces. The cuvette lands on four corner tabs, 8.5 mm below the beam.
The LED ran at 5.55 mA, under the 8 to 12 mA the design aims for. About 10 mA, with one resistor in place of two. See the circuit.
All five replacement parts are fitted on our reader. Between its first and last rows the right-hand column gives design values: the printed parts were not measured. The last row is measured. The two fits were chosen with printed coupons, described under Parts and files. Unchanged: the base plate, the detector side, the light trap and the firmware. Still absent: aperture disks and an emission filter. An enclosure was drawn and assembled, but not adopted — a shield over the open cuvette slot already brings the dark read to zero.

Building and running

Parts and files: what to buy, what to print, what it cost 4 tables

To buy

Core parts
FunctionPartState
Excitation LEDCree C503B-BCN-CV0Z0461Received; bench characterised
DetectorDFRobot SEN0365 (AS7341)Received; bench bring-up passed
ControllerESP32-WROOM-32, DOIT DevKit V1 30-pinIn bench use
LED switch2N7000 N-channel MOSFETConditional bench acceptance, single unit
Current limit / gate network217 Ω current limit (434 Ω for every version 1 reading), 217 Ω gate, 9.9 kΩ pulldownMeasured in place
Sample vesselEppendorf UVetteEngineering benchmark, not the final vessel
Optical corePrinted PLA: version 1's geometry, with five replacement parts for version 2Rebuilt 2026-10-03 as version 2, with all five fitted. How closely they fit was not measured
Fasteners36 M3 screws, 8 M3 nuts and 4 washersLengths under Assembly, step by step
ShieldOpaque cardboard boxInterim; every measurement needs it or a darkened room

Build version 2. Version 1 was never a complete set, for the reasons under From version 1 to version 2, and its alignment was shown to work but never quantified.

To print

The printed parts are in the 3D view. Each part there carries its size and can be saved as an STL file.

The five version 2 parts
PartReplacesSize (mm)On the print bed
UVette holder bodyThe old holder body, two gates, collar, gauge and preload leaf36 × 36 × 25.5Square base down
UVette seat riserVersion 1's riser43 × 42.5 × 13Large face down
LED holder bodyThe old holder body and its retainer clip21.6 × 21.6 × 21.3Flanged end down, cylinder upright
Aperture baffle bodyThe old baffle body15 × 16 × 13Flattened face down, bore horizontal
LED-holder capThe cap version 1 never received6 × 30.6 × 12.45Flat face down, arch opening up
Two dimensions cannot be taken from a drawing, because they depend on the printer: how far the ribs stand off the cuvette, and how tightly the LED is gripped. Two fit coupons were printed first, a block of four cuvette pockets from 0.05 to 0.20 mm clear on each side and a block of four LED sockets from 0 to 0.30 mm tight. The 0.05 mm pocket fitted and the three looser ones were too loose; the 0.10 mm socket held the LED and the next was slightly too tight. Those two values went into the parts, on a judgement by hand, one try per position, with no caliper.

All five print without supports in the orientation given. The print was specified as matte black PLA, which has to be opaque, on a 0.4 mm nozzle with 0.20 mm layers and 100 % infill, and nothing scaled or drilled afterwards. The two fit values came from coupons printed on one machine. On another printer, print the coupons first.

What it cost

Estimated cost of the 3D-printed build as we built it
ItemCost (US$)Basis
Version 1's print set, test pieces included$36.30Paid to a print shop
Version 2's two fit coupons$26.52Paid to a print shop
Version 2's five parts$59.66Paid to a print shop
ESP32 development board$6.79Our costing sheet
AS7341 sensor board$27.15Our costing sheet
Breadboard and jumper wires$3.00Our costing sheet
LED, MOSFET and three resistors$1.20Distributor list prices, 2026-10-07
Total, estimated$160.62Of which printing is $122.48
An estimate, not an account: the printing is what we paid, the boards are from our costing sheet, and the small parts are list prices, not receipts. Converted from New Taiwan dollars at NT$31.68 to US$1. Printing was bought from a print shop: a team with its own printer pays for filament. Not priced: the screws and nuts and the cardboard shield. The cuvette is a consumable, sold in boxes of 80. The laser-cut build is costed in its own section.

The files

The mechanical source is parametric CadQuery Python with STEP and STL exports; the firmware is an ESP32 sketch with its own minimal AS7341 driver, and needs no detector library.

What you need to build one
ArtefactFormat
Optical core geometryParametric CadQuery Python: version 1, and the five version 2 parts
Printable partsSTL: version 1's print set, 46 files for 59 pieces (the assembly's parts plus loose pieces such as shims, inserts and fit coupons); five files for version 2; two version 2 fit coupons
Assembly geometrySTEP
Acquisition firmwareESP32 sketch with its own AS7341 driver, P1-ACQ-1.0.0
Bill of materialsCSV
Bench circuitSchematic with measured values
Optical coordinatesEvery part and its X extent
Working sensor settingsAGAIN 512×, ATIME 255, ASTEP 280, about 200 ms per integration
Operating ruleDarkened room or opaque shield

The circuit, the coordinates, the sensor settings and the operating rule are on this page, and every printed part can be saved as an STL file from the 3D view. The CadQuery source, the STEP and STL exports, the firmware and the bill of materials are in the team's repository on iGEM GitLab, capture-screen-hardware, under the MIT licence. It holds every part the 3D view draws, for both versions. The order of assembly is in the next row.

Assembly, step by step 12 steps

This is the order the design gives, and the reader's builder has checked it against how ours was put together. In this order a screwdriver can reach each of the 26 screws on the base, which the CAD record checks one by one. Pull the 3D view apart to see each part named here.

Every screw is an M3. Most cut their own thread in a printed pilot hole 2.7 mm wide: stop when the screw is snug, because PLA strips. Eight pass through clearance holes in both the part and the plate, and take a nut underneath.

The screws, by where they go
WhereScrewsWhat sets the length
Feet4 × M3×8From above, through the 4 mm plate into the foot
Risers4 × M3×12From under the plate, 8 mm into a 9.5 mm pilot
Cuvette holder clips4 × M3×12Through the 4.9 mm clip into an 8 mm pilot
Light-trap clamp plates2 × M3×12Through the clamp plate and the trap's flange into an 8 mm pilot
LED-holder cap2 × M3×16Through the 12.45 mm cap into a 5 mm pilot
Baffle cap2 × M3×14Through the 10.25 mm cap into a 5 mm pilot
Cradle; detector frame ears; detector side clamps8 × M3×10, each with a nut3 mm of part and 4 mm of plate, and it must stop short of the bench, 4.5 mm below
Inside the detector holder4 × M3×16; 4 × M3×8 to 10 with washers; 2 × M3×6 to 8The cassette's clamp bars, the board carrier and the board retainer, as the detector's own notes give them
The lengths for the feet, the risers and the clips are the ones our records name. The others are worked out from the drawings, so that no screw bottoms in its hole, and were not checked on the reader.
  1. Clean the prints. Clear strings from the pilot holes, the sockets and the slots. The open-sided sockets on the detector frame, the slots in the baffle and the arrow engraved on the cuvette holder are meant to be there: do not fill them.
  2. Feet. One under each corner of the base plate, held by an M3×8 from above through the corner slot.
  3. Pins and keys. Into the plate's blind sockets: four round pins along the excitation axis, two for each riser; two narrow pins for the detector frame; two rectangular keys for the cradle.
  4. Risers. Seat the cuvette seat riser and the light-trap riser on their pins, and fix each with two M3×12 from under the plate.
  5. Excitation cradle. Onto its two keys, which set it across the beam and let it slide along it. Four screws through its slots, two in front and two behind, with nuts under the plate, only clamp it.
  6. Baffle and LED holder. Lay the baffle body and the LED holder body in the cradle's two saddles, facing as the fitting notes say, with a cap over each and two screws to a cap.
  7. Cuvette holder. Onto the seat riser, against its two edge stops. Four clips hold it down, two on the LED side and two on the side away from the detector, each with an M3×12.
  8. Detector frame. Onto its two pins, with a screw and nut through each of its two ears. Its rail and rear stop are what locate the detector holder.
  9. Sensor board. Into the carrier, sensor facing the sample, under the U-shaped retainer: two screws, the open side to the connector, only tight enough to keep the board in. The carrier goes into the back of the detector holder on four screws with large washers through its oversize holes. Leave them loose, slide the carrier to where the reading is highest, then tighten crosswise.
  10. Filter cassette. Against the front of the detector holder, its drawer opening towards the LED end, held by a clamp bar along its top edge and another along its bottom, two screws to a bar. Slide in an insert: ours holds no filter.
  11. Detector holder. Stand it against the frame's rail and rear stop and close the two side clamps on it, a screw and nut to each.
  12. Light trap. Onto its riser, with a D-shaped clamp plate over each of its two keyholes and a screw through each. Then the LED, its wires and the cuvette, as in the fitting notes, and the circuit.
Fitting the version 2 parts 5 steps

These are the notes that went with the parts. They were written at design time and are not yet a record of an assembly that was checked step by step.

There are no photographs of the assembly: the reader was already put together when this page was written. The 3D view can be pulled apart to show where each part goes.

  1. Cuvette. Line up the arrow marking the UVette's 10 mm path with the double-headed arrow on the holder's top face, which runs along the detector axis. Lower it in vertically and push it to the bottom. Keep one orientation through a series.
  2. If the cuvette is tight. Sand the ribs in the pocket lightly, eight of them and each 1.6 mm wide. Do not sand the cuvette. A tight pocket can be eased; a loose one cannot.
  3. LED holder. The straight groove on its outer wall faces up, on the centre line of the baffle's flat top, and the wire notch at the back faces down. Push the LED in from behind until its flange meets the shoulder.
  4. Wires. Out through the notch, tied between the two flanges with a cable tie no wider than 2.5 mm.
  5. Cap. Two M3 screws into the cradle, holding the LED holder down.
The circuit, with its measured values 434 Ω and 217 Ω

The excitation branch turns one LED3 on and off at a current that cannot damage it: a resistor-limited LED with an N-channel MOSFET on the low side, switched from one pin of the controller. It is a breadboard circuit that was measured, not a product design. The gate pulldown, R3, is the part that is not optional: without it the LED can emit while the controller boots, during what the firmware takes for a dark baseline.

Excitation branch and sensor bus

Hover or focus a labelled node for its measured value

ESP32-WROOM-32 DOIT DevKit V1, 30-pin SEN0365 AS7341 detector R1 434 Ω 470 nm LED 2N7000 R2 217 Ω R3 9.9 kΩ J0 GPIO21 SDA · 0x39 GPIO22 SCL GPIO25 VIN
Measured 2026-08-25 on breadboard. Total draw ≈ 77 mA. ILED = 5.553 mA, measured 2026-08-25: every version 1 reading.
Acquisition timing 551–552 ms of a 1.0 s ceiling

One reading is not instantaneous. The LED turns on, the optics and the sensor are given time to settle, the sensor integrates twice to cover all its channels, and the LED turns off again. The firmware holds a hard 1.0 s ceiling on that window.

One acquisition, against the 1.0 s ceiling

Measured 2026-08-25 with a diagnostic sketch, one integration per channel set

settle 100 ms
all-channel read 451.6 ms
remaining margin 448 ms
0.5 s target
LED off · 551–552 ms
0250 ms 500 ms750 ms 1.0 s hard timeout

The window fits the 1.0 s hard timeout with 448 ms of margin and misses our own 0.5 s target by about 52 ms, which we have recorded as a missed target. That was measured with a diagnostic sketch. The project firmware has since reported 550 to 552 ms in every session, by its own timer and not by an edge measured on the pin, and the 1.0 s cut-off itself has never been exercised.

History

The development log 12 steps, 2026-08-14 to 10-12

The same story, step by step. Each step in the log was a test whose result changed what we did next.

Development log

Twelve steps, 2026-08-14 to 2026-10-12

  1. 2026-08-14 to 08-17

    The printed parts arrive

    The printer flagged walls too thin to print on one part, which was redrawn first. Of 59 pieces, 57 arrived; two were missing, and two broke as their supports were removed.

    What it changed We built with what we had, taped the two breaks, and wrote down where they broke for the next revision.

  2. 2026-08-25

    First light

    With the LED at 5.553 mA, a paper scatterer at the sample position read F2 = 87 and F3 = 81 against dark reads of zero. With nothing in the beam, stray excitation stayed under one count.

    What it changed The light path works, taped joints included. The UVette's orientation was fixed from here on.

  3. 2026-08-27

    The room-light test: room light gets in

    An empty UVette under ordinary room lighting read F4 = 11 with the LED off. In a darkened room, or under an opaque box, every channel read zero.

    What it changed A shield became mandatory, and “the dark read must be zero” became the rule that voids a cycle.

  4. 2026-09-03 to 09-04

    First fluorescence

    Running the project firmware, a culture reported to carry sfGFP read F4 = 19.58 ± 0.79 and its control read 0, over twelve cycles each. Twelve cycles are not twelve samples.

    What it changed The emission channel is usable. We had called that culture “induced”; no AHL had been added, and the word was withdrawn.

  5. 2026-09-09 and 09-14

    Two re-insertion tests, five days apart

    A GFP culture read 40.50 ± 1.23 over four insertions, with its control at zero. Five days later the corresponding culture read 32.67 ± 1.83, lower by 19.3 %. The instrument had been moved in between and the state of the culture was not known.

    What it changed A difference across days cannot be explained afterwards, so every comparison now carries its reference in the same session.

  6. 2026-09-16, 09-21 and 09-29

    Three mixing sessions

    GFP stock, non-GFP stock and medium, mixed at set fractions. Samples without GFP read zero every time, and in the two sessions with enough signal the reading rose in proportion to the GFP culture. A thin stock in session 2 left three tubes at zero.

    What it changed The stock has to be dense enough to clear the floor, and one session's slope cannot be carried to another as it stands. The Model page takes it from there.

  7. 2026-09-23 to 10-02

    The weak parts, redrawn

    The cuvette seat and the excitation end were redrawn: a pocket the cuvette drops into and lands in, an LED holder that grips the LED without a clip, and a baffle that prints on its side. With the cap that never arrived, that makes five parts to print. Two fit coupons were printed first, and the pocket and the socket that fitted by hand set the two dimensions a drawing cannot settle.

    What it changed The fractures became a check: each printed layer's cross-section against the fullest section within 3 mm of it. Where the parts broke that ratio was 0.63 and 0.32; the five new parts keep at least 0.82, against a threshold of 0.75. It is a geometric screen tuned on those two breaks, not a strength calculation.

  8. 2026-10-03

    Version 2 on the bench

    New printed parts were fitted, and one resistor was changed to take the LED from 5.55 mA to about 10 mA. With the same empty cuvette the rebuilt reader read F2 = 7.75 and F3 = 6.71 after an adjustment, inside the range version 1 had read, and about twice that once the current was raised. The emission channel stayed at zero throughout.

    What it changed The reader is no longer the instrument that produced version 1's bench and culture readings. Its fit, its alignment and its response to a culture have to be measured again before the two can be put side by side.

  9. 2026-10-05

    The dilution series: version 2 reads a culture

    One batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls. The controls read zero on the emission channel. Looked at afterwards, the reading stayed close to one straight line from 25 to 200 µL of culture and fell short of it below that.

    What it changed Version 2 has culture readings of its own. They are one session with one batch, so they stand beside version 1's and are not yet a comparison with them.

  10. 2026-10-09

    The AHL plate session: the reading follows the culture

    Suspensions scraped from plates at five AHL concentrations, with medium, a culture without the reporter and an sfGFP culture as controls. The reading rose with how much culture was in the cuvette, whether or not the wet lab had seen the plate glow, and sfGFP added to a dense suspension was still read.

    What it changed Zero without GFP now carries a condition: it holds in liquid culture, not in a suspension this dense. Reading induction needs samples of matched density and one without AHL.

  11. 2026-10-11

    The density session: where zero ends

    A culture without the reporter, diluted in LB and read from thin to dense. It read zero while the scatter channel stayed under about 33 counts, and no longer did from 46.

    What it changed A few counts on the emission channel can be set against the scatter channel of the same read. An emission filter remains untested.

  12. 2026-10-12

    Beside a plate reader

    Five dilutions of the sfGFP test culture, read on our reader and on a SpectraMax M3. The two ranked the tubes alike, and over the three denser tubes our reading per unit of OD held within 10%.

    What it changed The reader has now been set beside a plate reader once. Whether the two fluorescence readings are in proportion is still open: the plate reader needs every tube read undiluted.

The week-by-week record is in the Notebook.
Decisions that narrowed the first prototype 5 decisions

Recorded project decisions from 2026-08-09. The marked ones removed capability from the prototype on purpose.

The turbidity check is out of this prototype's scope narrows the scope
Red-light transmission and the turbidity check are removed. The far side of the excitation axis is occupied by the light trap, so there is no transmission path and the measurement is not physically implementable in this geometry.
470 nm excitation for the first round
Three independent design documents agree on the nominal 470 nm Cree LED, and the LED holder was dimensioned from that part's datasheet.
The UVette is a benchmark, not the product vessel
The reaction cartridge is a joint wet–dry decision that has not been made. Future containers will be handled with adapters rather than by modifying approved optical modules.
PlateDock is out of this prototype's scope narrows the scope
Petri-dish imaging and swarming analysis are held as separate future work. No CAD for it exists in any revision, and the effort belongs on validating one reader properly.
Angled-wedge light trap selected for manufacture
The stepped-baffle variant is deferred, not rejected. Both validate digitally and no bench evidence separates them, so the selection is a manufacturing choice and explicitly not a claim that the alternative is worse.

See the laser-cut build Back to the two builds

The laser-cut build

Every reading in this section was taken on this build, with its lid closed.

Measured on E. coli cultures with and without sfGFP, this build's green-to-blue ratio rose in a straight line with the amount of sfGFP (r² ≈ 0.97), with a 3σ detection limit of about 4% sfGFP culture; the tube with no sfGFP varied by 1.5% in that ratio over eight batches. Across its eight colour channels the added signal has the spectral shape of fluorescence, not of scattering.

r² ≈ 0.97
signal against the amount of sfGFP, seven tubes
≈ 4%
sfGFP culture, the 3σ detection limit
1.5%
variation (CV) in F5/F3 of the tube with no sfGFP over eight batches, four placements, linearity test

The instrument

A blue LED, a 90° sensor and a light trap, in a plywood box cut from one sheet.

A blue LED lights a UVette from the side, and an AS7341 spectral sensor looks at it from 90°. Everything sits in a laser-cut plywood enclosure in two layers: the optics above, blackened with matte black marker, and the ESP32 and its wiring below, with a deck between them. An OLED shows each result and one button starts a measurement.

The laser-cut reader closed, on a wooden table: a bare plywood box with finger-jointed edges and a small square cover on its lid. Wires come out of an opening low on the front panel to a lit OLED display and a button module.
The laser-cut build, closed, with the OLED and button.
Top view into the open enclosure, roofs off, with the blue LED lit. On the left, the light-trap chamber with a black square sloping inside it; in the middle, a clear UVette glowing blue in its pocket; on the right, the LED chamber with the LED taped in place; below the cuvette, the sensor board in its slot, with red, black, blue and green wires running down the detection chamber.
The optical layer with the roofs off and the LED on: the light trap with its sloped square on the left, the UVette in its pocket, the LED in the right-hand chamber, and the sensor below the cuvette with its four wires.

The reader in 3D

Every laser-cut part, from the generator at 3.0 mm

The deck carries three chambers around the cuvette pocket: the LED chamber, the light trap straight across from it, and the detection chamber at right angles. Click a part for its size.
Excitation
Cree C503B blue LED, 470 nm, driven from GPIO25 through 220 Ω
Detector
AS7341, ten channels (DFRobot Gravity SEN0364 board), I²C 0x39
Sensor settings
Gain 512×, ATIME 59, ASTEP 999: 166.8 ms per integration, full scale 60,000 counts. Unchanged throughout
Cuvette
Eppendorf UVette, 200 µL. The 2 mm path faces the LED, the 10 mm path faces the sensor
Light trap
A deep, dark chamber straight across from the LED, entered through an 8 mm hole in the wall of the cuvette holder
Controller
ESP32 DevKit (ESP32-WROOM-32), with an SSD1306 OLED and a push button
Enclosure
Laser-cut 3 mm plywood, matte black marker inside the optical layer; two layers

The light trap is designed to swallow the excitation that passes straight through the sample so that it does not bounce back toward the sensor.

The sensor sits at 90° to the beam, so the excitation does not shine straight into it. Some blue still reaches it: with an empty chamber the blue channel F3 (480 nm) read about 312 counts, and a cuvette of plain medium raised it to about 730 (both in the fluorescence test). That the reading goes up when the sample goes in is consistent with the sensor collecting light scattered sideways out of the sample.

Design advantages

Cheap to make from one sheet of plywood, cut to whatever board you have, and steady in its readings.

  • Low cost $53.70

    Everything as costed below; the jointed parts come from one 450 × 300 mm sheet.

    As costed below; the UVette was not priced
  • Any 3 mm board 2.8–3.2 mm

    The generator redraws every joint for the board's measured thickness, so any batch of nominal 3 mm plywood will do; only the jointed parts need to come from the same batch.

    Checked in the generator, below
  • Steady 1.5%

    Eight batches of the tube with no sfGFP varied by 1.5% in F5/F3.

    Linearity test

User feedback

Wet-lab members used a first version, and two changes came from what they found; every measurement on this page was taken on the version that followed.

  • The lid's opening 16to 25 mm

    They told us it was too small to lift the cuvette out. What followed: it was enlarged from 16 × 16 to 25 × 25 mm, with a matching lid.

    From wet-lab use
  • Placing the cuvette 13 × 13 mm

    They told us that with nothing to hold it, the cuvette had to be placed with care, watching where it sat and keeping clear of the chamber walls. What followed: a plain plate with a 13 × 13 mm opening now lies on top of the chambers and holds the cuvette in place.

    From wet-lab use
Top view into the open enclosure: three black chambers with their roofs on, and a clear cuvette standing in the narrow pocket where they meet, with nothing above it to hold it.
Without the plate: the cuvette stands in its pocket between the chambers, with nothing to hold it upright.
The same enclosure with a plain plywood plate laid over the chambers; a square opening in its centre sits over the cuvette pocket.
With the plate: its 13 × 13 mm opening sits over the pocket and holds the cuvette in place.

How a reading is made

Each reading subtracts LED-off from LED-on, and the result is the ratio F5/F3, which cancels changes in how much light reaches the sensor.

For each reading the firmware takes eight samples with the LED off, then eight with it on, and subtracts the averages channel by channel, which removes ambient light and the sensor's dark current; averaging eight samples also lowers the sensor's own noise by about √8, or 2.8 times. One press of the button records a batch of five readings, about 22 seconds each. The first is discarded and the other four are averaged.

The figure we report is F5/F3: the green channel at 555 nm divided by the blue channel at 480 nm. Dividing by the blue channel largely cancels changes in how much light reaches the sensor, whether from where the cuvette sits or from changes within a batch: over readings two to five, F3 and F5 each fall by about 2.7%, while F5/F3 changes by less than 0.1% on average. Net signal is the F5/F3 of a tube minus that of a tube with no sfGFP cells, measured in the same session.

Why F5, and not F4

F4, centred nearer the sfGFP peak, collects about 2.2 times as much sfGFP light as F5, but it also collects about three times the background from the blue LED, so its detection limit is slightly worse (about 4.4% against 3.6%). On this build F5/F3 is the better measure of sfGFP; F4/F3 supports the same conclusions with more noise.

F5 against F4, in full the same data, both ratios

sfGFP emits most strongly near 510 nm. F4 and F5 each respond over a band 39 nm wide at half maximum (AS7341 datasheet): F4 from about 496 to 535 nm, F5 from about 536 to 575 nm. The band centred closest to the sfGFP peak is F4. This build was set up from the start to read F5 (555 nm): the on-screen signal of the firmware used for these measurements is built on it. To check that choice we recomputed the main results with F4/F3 alongside, using the same data and the same method.

Both ratios from the same readings and the same method; the experiments are those described under Validation.
MeasureF5/F3F4/F3
Detection limit, share of sfGFP culture (3σ), linearity test≈ 0.036≈ 0.044
Linearity r², linearity test0.970.96
Dose response r², fluorescence test0.900.82
Background relative to F3, tube with no sfGFP cells0.0550.160
Shift when a blank is lifted out and put back2.9%0.7%

F4/F3 supports the same conclusions, linear and with an intercept indistinguishable from zero. Where it does better is stability: it shifts less when the same tube is taken out and put back, and drifts less over an hour, because F4 and F3 are close in wavelength. For sensitivity, F5/F3 is the better measure on this build.

The other number in our records: SIG5 and k5 not used for any result

For its on-screen reading the firmware also computes SIG5 = F5 − k5 × F3. Part of what F5 sees is not sfGFP fluorescence, such as the blue LED's long-wave tail scattered into the sensor; k5 is that part as a fraction of F3, measured on a tube with no sfGFP cells. It is set by sending the k command with such a tube in place, and stored in the ESP32's flash.

k5 changes from one experiment to the next, so SIG5 is not compared across batches. Every result on this page uses F5/F3, which does not depend on k5.

Validation

The LED holds steady, and with the sensor aligned the reader detects sfGFP down to about 4% sfGFP culture; our AHL sensor strain's signal is too faint for it.

The density test: does the reading follow the sample?

Aim Check that the reading follows what is in the cuvette.
Method Cultures of different density (OD600 0.15 to 1.53) were read one after another, and the blue channel F3 of the 28 batches with an OD600 record was compared with their OD600.

Result Culture density explained almost none of the variation in the blue channel F3 (R² = 0.01): the reading was not following what was in the cuvette, consistent with a sensor mounted too low to see the sample well. The reading also fell by about 40% in a single step between the seventh and eighth batches, about 20 minutes in, and stayed low. Allowing for culture density, a single step fits the data much better than a gradual drift (R² 0.88 against 0.67). The cause of the step was not identified, but a sudden change like this is more consistent with the sensor shifting than with the LED fading. The test gave no usable culture data. The next test checked whether the LED was at fault; after that, the sensor was aligned with the cuvette's measuring window and fixed in place for all later tests.

The blue channel F3 in the density test, for the 28 batches with an OD600 record, in the order they were measured and corrected for culture density. The line is the best single-step fit: a drop of about 40% between the seventh and eighth batches.

The LED stability test: an hour with nothing moved

Aim Find out whether the LED itself fades, which could have caused the fall seen in the density test.
Method With the sensor still in its original position and the chamber empty, the firmware repeated batches automatically for an hour: five readings taking about two minutes, then a one-minute pause with the LED off. Twenty-one batches in all.

F3 with the chamber empty and nothing moved, the mean of readings two to five of each batch, relative to the first batch. All 21 batches lie within a 1% range (shaded); the last batch is 0.5% below the first.

Result and conclusion Over an hour the reading stayed within a 1% range. The LED and electronics are stable, so the fall in the density test was not the LED fading. The sensor was then aligned with the cuvette's measuring window and fixed in place for every test that followed.

The fluorescence test: is the signal sfGFP?

Aim Show that the signal comes from sfGFP fluorescence and scales with the amount of sfGFP.
Method Twelve tubes of 200 µL in a grid: culture density d of 0.25, 0.5, 0.75 and 1 (the fraction of each tube that is culture, the rest medium) crossed with a fraction f of 0, 0.5 and 1 of cells carrying sfGFP. The two stock cultures had nearly the same density (OD600 0.84 with sfGFP, 0.85 without), so the share of culture taken from the sfGFP stock is close to the share of sfGFP cells. Each tube was read in at least two batches.

Result Net F5/F3 rose in proportion to the amount of sfGFP, d × f (r² ≈ 0.90, eight tubes), and the line passes through zero (intercept −0.5 ± 2.2, in net F5/F3 × 1000).

Conclusion The signal behaves as sfGFP fluorescence: it is proportional to the amount of sfGFP and has the spectral shape of its emission.

Why this is not scattering the spectral check

A signal that rises with the amount of sfGFP does not on its own prove fluorescence. Each net value already compares tubes of the same density, so turbidity is subtracted, but the two strains could still scatter light differently. The spectral shape can tell scattering from fluorescence, and it is shown under What ten channels add: the difference between tubes with and without sfGFP has the shape expected of sfGFP fluorescence.

The linearity test: a straight line and a detection limit

Aim Keep the total culture density nearly constant, vary only the share of sfGFP culture, and measure the detection limit.
Method Tubes of 200 µL mixed to shares f of 0, 0.10, 0.15, 0.20, 0.30, 0.50, 0.75 and 1, where f is the share of the volume taken from the sfGFP culture, which is proportional to the number of sfGFP cells. The OD600 of every tube in the fit was measured, from 1.30 (no sfGFP) to 1.45 (all sfGFP). Each tube was read in at least two batches. The tube with no sfGFP was placed four times during the session and read in two consecutive batches each time, eight batches in all.

Net signal (F5/F3 of each tube minus that of the tube with no sfGFP) against the share of sfGFP culture, f. The fit uses the seven tubes with sfGFP; the tube without it is the zero that is subtracted, so it is not part of the fit. Relative to their size, the two lowest fitted tubes, f = 0.10 and 0.15, lie furthest from the line. Open markers: three further low-share tubes, shown but not fitted.

Result Signal and the share of sfGFP culture are linear. Without forcing the line through the origin, the intercept is 1.6 ± 2.6 (standard error, in net F5/F3 × 1000), not different from zero, with r² ≈ 0.97; a line through the origin has slope 68.7 and r² ≈ 0.97. Across its eight batches, the tube with no sfGFP varied by 1.5% in F5/F3 (standard deviation 0.00083 divided by a mean of 0.055). The detection limit, three times that standard deviation divided by the slope in the same units (0.0687 per unit f), is 0.036: about 4% sfGFP culture (from F5/F3).

The low end why the lowest shares scatter

At the lowest shares the points scatter most relative to their size: the two lowest tubes in the fit are off by −23% (f = 0.10) and +48% (f = 0.15), while from f = 0.20 up every tube is within about ±20%. At f = 0.05 a 200 µL tube holds the equivalent of only 10 µL of sfGFP culture, so an error of 2 µL is already 20% of f; the low tubes are the most sensitive to how they were mixed. The reading itself is not the cause: repeated batches of the same tube agree, the zero is stable, and F4, F5 and F6 are off in the same direction and by similar amounts in each tube (correlation 0.95–0.98), so the deviation lies in how much sfGFP the tube holds, not in one channel. Three further low-share tubes (two at f = 0.05, one at f = 0.10) scatter the same way, from −54% to +276%, and are shown but left out of the fit. Including them changes little: the slope becomes 68.8 and r² 0.95, and the detection limit, which comes from the blank tube, does not change. Mixing low shares by serial dilution should reduce this scatter.

The sensor-strain test: our AHL sensor on this reader

Aim Read the AHL-sensing strain our wet lab built, E. coli TOP10 carrying ProD-LasR · pLasRV-sfGFP, across a range of AHL.
Method The strain was given AHL at six concentrations, 10 nM to 1 mM, with two or three readings at each, 16 in all, and three tubes of medium alone as blanks. It was read with the current firmware rather than the batch firmware of the tests above.

F5/F3 of each reading of the sensor strain, and of medium alone. The two dashed lines are from the linearity test, a different session with the batch firmware: a culture without sfGFP, and the undiluted sfGFP test culture.

Result F5/F3 stayed between 0.048 and 0.060 (mean 0.055) across the whole range, at the level of the culture without sfGFP in the linearity test (0.055) and far below the undiluted sfGFP test culture (0.127); medium alone read 0.026 to 0.033. The differences between concentrations were no larger than those between tubes at one concentration. F4 rose and fell with the red channel F7 and the clear channel (r = 0.99 and 0.92): the readings followed the light scattered by the cells.

Conclusion This reader detects and quantifies a strongly expressing sfGFP culture down to about 4%, but our sensor strain's F5/F3 could not be told apart from a culture without sfGFP read in another session. No culture without sfGFP was read alongside it, so the result cannot be stated against the 4% detection limit. In the wet lab's plate reader the strain did glow above its control (Engineering), so the signal exists but is too faint for this reader. The 3D-printed build saw the same pattern with dense suspensions: readings followed the amount of culture, not the AHL (Where it stops).

What ten channels add

The extra channels confirm that the signal is fluorescence, and that the ratio cancels where the cuvette sits.

Every reading records all ten channels, while the results above use only F5 and F3. Recomputing the later tests with the other channels checks those results from a different angle, without another experiment. The conclusions hold up; these are the two checks that matter most.

The added signal has the shape of fluorescence

For each culture density in the fluorescence test we took the tube with all cells carrying sfGFP and the tube with none, divided each by its own F3 to remove brightness differences, and subtracted. What is left is the spectrum that sfGFP adds. Because every channel is divided by F3, the F3 point is zero by construction; each curve is then scaled to its own F4 value so that the four densities can be compared by shape.

The fluorescence test. Difference spectra, each scaled to its own F4 value; F3 is zero by construction. Across the four densities F5 to F8 agree to a standard deviation of 0.03 or less. F2 (445 nm, shown hollow) sits next to the excitation and is very sensitive to turbidity (standard deviation 0.53), so it is not read.

The four curves almost coincide: a single peak at F4 (515 nm), falling steadily toward the red, and near zero at 415 nm. That is the shape expected of sfGFP emission. Scattering would be strongest at short wavelengths, and a change in turbidity would move every channel the same way; neither would produce this curve.

The ratio cancels where the cuvette sits

If only the amount of light reaching the sensor changes, all channels change in proportion and the ratio is unaffected. The figure shows a real case from the fluorescence test: a blank was read, taken out while another tube was read, then put back and read again.

Each channel's change divided by the mean change of the nine channels, so changes in proportion lie on a flat line at 1.00. The near-infrared channel is left out: its signal is only about 20 to 40 counts.

When the blank went back in, the nine channels rose together by 14 to 20%, at most 6% apart, while F5/F3 moved by 2.9%: the jump came from where the cuvette sat, and the ratio largely cancelled it.

Limits and next steps

  • It does not measure turbidity. The 90° geometry suits fluorescence and does not track culture density; no choice of channels recovers it. Cell density has to be measured separately, with a spectrophotometer.
  • Low shares scatter more. Below about 20% sfGFP culture, tubes lie further from the line (see The low end); mixing them by serial dilution should help but has not been tested.
  • Experiments are not directly comparable. The reader was moved between sessions, so absolute counts differ from one experiment to the next (an empty chamber read about 310 in the fluorescence test and 160 to 190 in the linearity test). Each experiment is therefore referenced to its own tube with no sfGFP, measured in the same session.
  • The AHL sensor strain is too faint for this reader. Our sensor strain glows above its control in a plate reader, but on this reader its F5/F3 stayed at the level of cells without sfGFP, read in another session, from 10 nM to 1 mM AHL; no culture without sfGFP was read alongside it (see the sensor-strain test). A culture without sfGFP read in the same session would settle whether the strain reads at zero or only close to it.
  • No reference comparison yet. This reader has not been compared with a plate reader, so its readings are checked only against themselves: linearity, the ratio’s stability on reinserting a tube, and the shape of the spectrum. Reading the linearity tubes on a plate reader alongside it is the comparison we would run first.

Building one

  1. Build

    Generate the parts below for your board's thickness and cut the jointed parts from one 450 × 300 mm sheet of the same board. Colour the parts that face the optical layer with a matte black marker; fit the chamber walls into the deck by their tabs, set the deck on its rails inside the enclosure, and fit the roofs onto the walls.

    The laser-cut parts Assembly, step by step

  2. Wire

    The AS7341 sensor and the OLED share the ESP32's I²C bus; the LED runs from GPIO25 through 220 Ω, and the button sits on GPIO13. Everything runs from 3.3 V, and the ESP32 takes its power over USB from any supply.

    The circuit Wiring

  3. Run

    Seat the UVette with its 2 mm path toward the LED, close the lid, and press the button: five readings of about 22 seconds each, the first discarded and the other four averaged. In each session, also read a tube with no sfGFP; results are its F5/F3 subtracted from each sample's.

    How a reading is made Operating steps

The laser-cut parts

The whole set of parts is generated in the browser from the measured thickness of the plywood. The optical layout stays fixed: the cuvette pocket, the heights and positions of the light holes, and the inside of every chamber do not change with thickness, and thicker board makes the walls grow outward. Tabs are 8 mm wide; slots and notches are made 0.1 mm smaller on each side to take up the laser's kerf. The jointed parts all fit one 450 × 300 mm sheet and must be cut from one batch of board, because every joint is sized from a single measured thickness. The small parts without joints fill the gaps around them; those that do not fit go on a second sheet of any board.

Assembly, step by step 10 steps

This is the order our reader was put together in. Open the chambers in the 3D view to see each part named here.

  1. Cut and colour. Cut the parts for your board's thickness, the jointed ones from one sheet. Blacken every surface facing the optical layer with matte black marker; the electronics layer can stay bare.
  2. Chambers. Press the walls' feet into the deck slots, the 12 mm wall included; friction holds them.
  3. Light trap. Glue the two triangles against its long walls and the square onto their slopes, lower edge against the wall with the 8 mm opening.
    The deck with its chambers pressed in, roofs off. Left, the light trap with its sloped square; right, the LED chamber; bottom, the detection chamber; centre, the cuvette pocket.
    The deck with its chambers pressed in, roofs off. Left, the light trap with its sloped square; right, the LED chamber; bottom, the detection chamber; centre, the cuvette pocket.
  4. LED. Push the LED into the 4 mm hole in the cuvette wall, tape it in place, and take its leads down through the round wire hole.
  5. Sensor. Glue the support plate under the deck's sensor slot, stand the sensor board in the slot facing the cuvette, and take its wires down through the rectangular wire hole.
  6. Enclosure and rails. Glue the base and sides together, glue the four rails at about half height, and tape the breadboard with the ESP32 to the base.
  7. Deck. Connect all wires to the ESP32 as in the wiring table, then lower the deck onto the rails.
    The deck in the enclosure, roofs off. The sensor stands in its slot facing the cuvette pocket, its four wires going down through the rectangular wire hole; the LED is in the right-hand chamber and the light trap on the left.
    The deck in the enclosure, roofs off. The sensor stands in its slot facing the cuvette pocket, its four wires going down through the rectangular wire hole; the LED is in the right-hand chamber and the light trap on the left.
  8. Display and button. Lead their wires out through the long side's opening, connect them, and tape both over the opening.
    The electronics layer, with the deck lifted out and stood up behind it, underside facing out: the ESP32 on its breadboard, the bare-wood rails around the walls, and the OLED and button led out through the opening in the long side. The sensor's four wires and the LED's two come down through the deck to the breadboard.
    The electronics layer, with the deck lifted out and stood up behind it, underside facing out: the ESP32 on its breadboard, the bare-wood rails around the walls, and the OLED and button led out through the opening in the long side. The sensor's four wires and the LED's two come down through the deck to the breadboard.
  9. Roofs. Press the three roofs into the wall-top notches.
    The deck in the enclosure with the three roofs on, before the plate and the lid.
    The deck in the enclosure with the three roofs on, before the plate and the lid.
  10. Plate and lid. Lay the plate over the chambers, its 13 × 13 mm opening over the cuvette pocket, and put the lid on; neither is glued. Glue the 25 mm square under the 27 mm one to make the cover for the lid's opening.
    The plate with the 13 × 13 mm opening resting on the chambers, before the lid goes on.
    The plate with the 13 × 13 mm opening resting on the chambers, before the lid goes on.

Assembly tip: the deck and the enclosure panels have symmetric outlines, so each also fits when turned around, but their holes are not symmetric and the parts carry no orientation marks. Before gluing, check that every hole lines up with what it serves: the LED, the cuvette, the sensor and the wire holes.

Operating steps 6 steps

These are the steps the first four tests under Validation were taken with, using the batch firmware described under Firmware and data.

  1. Connect. Plug the ESP32 into a computer by USB and open a serial monitor at 115200 baud. The display shows the reader's status and “Type ID, press button”.
  2. Load. Fill a UVette with the sample and lower it through the lid opening into its pocket, the 2 mm path toward the LED. Close the opening with the small cover.
  3. Calibrate. Start each session with a tube with no sfGFP. Load it and run one batch first, then type k into the serial monitor. The display shows CAL OK with k5, the share of the blue signal that reaches the green channel. This only sets the on-screen SIG; it does not affect F5/F3.
  4. Name and run. Type the tube's ID into the serial monitor and press the button. The reader takes five readings, about 22 seconds each; leave it alone while the display says “do not touch”.
  5. Read. The display shows DONE with the mean F3 and SIG of readings two to five. Every reading also goes to the serial monitor as a line of CSV with all ten channels; F5/F3, the figure used on this page, is computed from those lines.
  6. Reference. Read the tube with no sfGFP again during the session. Every result on this page is a tube's F5/F3 minus that tube's.

The circuit

Everything runs from the ESP32's 3.3 V supply. The AS7341 and the OLED share one I²C bus at different addresses (0x39 and 0x3C); both boards carry their own pull-up resistors. The LED is driven directly: GPIO25 to 220 Ω to the LED to ground, switched by the firmware. The button connects GPIO13 to ground and uses the ESP32's internal pull-up. None of the pins used (13, 21, 22, 25) is a strapping pin that affects boot. The ESP32 itself is powered over USB, through the opening in one short side of the enclosure: we use a power bank, or the computer when the batch firmware sends its readings, but any USB supply will do.

Breadboard wiring diagram. An ESP32 DevKit sits across the middle of a breadboard. The blue LED's long leg runs through a 220 ohm resistor to GPIO25 and its short leg to ground. A push button connects GPIO13 to ground. The SSD1306 OLED and the AS7341 sensor board each take 3.3 V and ground from the top rails, and share the I2C lines from the bottom rails: SDA to GPIO21 in green, SCL to GPIO22 in blue.
Breadboard wiring, redrawn from the wiring table below. The blue LED runs from GPIO25 through 220 Ω, the button sits on GPIO13, and the AS7341 and the OLED share the I²C lines.
Circuit schematic. The ESP32 DevKit, U1, connects through GPIO25 and a 220 ohm resistor R1 to the blue LED D1, and through GPIO13 to push button SW1, both returning to ground. Its 3V3, GND, GPIO21 (SDA) and GPIO22 (SCL) go to both the AS7341 Gravity board U2, at I2C address 0x39, and the 0.96 inch SSD1306 OLED U3, at 0x3C.
The same circuit as a schematic: the ESP32 (U1), the AS7341 (U2, DFRobot Gravity SEN0364) and the SSD1306 OLED (U3) on one I²C bus, all from the 3.3 V rail.
Wiring. All parts run from 3.3 V.
SignalESP32 pinConnects to
Power3V3AS7341 VCC, OLED VCC
GroundGNDAS7341 GND, OLED GND, LED cathode, the button's other terminal
I²C SDAGPIO21AS7341 SDA, OLED SDA
I²C SCLGPIO22AS7341 SCL, OLED SCL
LEDGPIO25220 Ω to the LED anode; cathode to GND. HIGH turns it on
ButtonGPIO13One terminal of the button; the other terminal to GND. Internal pull-up, reads LOW when pressed

Firmware and data

The readings on this page were taken with the reader's batch firmware (v5.5), which prints every reading over the serial port. The current firmware builds on it and adds the link to our LasReader software: a measurement still takes five readings and keeps the last four, and the data processing is done in the software rather than on the reader. It is published in the team's GitLab repository; the software is described on the Software page.

What it costs

The costing below is for the laser-cut build, including its enclosure, display and button. All figures are in US dollars.

Cost of one laser-cut build $53.70 as costed
Cost of one laser-cut build
ItemQty Cost (US$)
Electronics
ESP32 development board1$6.79
AS7341 spectral sensor board1$27.15
Push button1$0.51
OLED display1$4.73
Breadboard1$1.42
Jumper wiresseveral$1.58
LED and 220 Ω resistor1 each$0.16
Electronics subtotal$42.34
Enclosure, laser-cut
Wood board1$1.89
Laser cutting1$9.47
Enclosure subtotal$11.36
Total, as costed$53.70

Not in the costing: the Eppendorf UVette, which the sheet never priced, and the USB cable and power supply, which any USB supply can be; we have not guessed at them. The few small parts that do not fit on the main sheet were cut from leftover board, so they add nothing to the cost.

See the 3D-printed build Back to the two builds

How we got here

We built it for our biosensor, rehearsed on GFP cultures, and ended with a platform that other teams can tune to their own samples.

We expected an AHL-induced culture to be about as bright as a GFP culture, so the first stage ran on a GFP test strain as a rehearsal. What we had read beforehand put the two in the same range, as one LuxR circuit is8. What we had not noticed is that circuits differ: a Las device, the kind our biosensor is, stayed well below a constitutive reference9.

Along the way we found that how dense a sample is changes the reading, and worked on describing it with a model and with the reader’s other channels (Model 3.5).

The samples our wet lab then gave us were denser still, and AHL-dependent induction had not been shown conclusively (Engineering): not what we had prepared for, and a different thing to measure. A plate reader has its place, and we read ours beside one. What the two builds offer is a steady platform to read from. Both record ten channels on every read, and both run from the same software. Under its shield the 3D-printed build reads zero in the dark, so its counts are used as they come; the laser-cut build is cheaper and easier to assemble, and removes its background in the data. A team can choose which bands to read, and how to process them, to suit its own samples.

References

  1. ams OSRAM. AS7341 11-Channel Multi-Spectral Digital Sensor, datasheet DS000504, v3-00, 2020. ams-osram.com. The 3D-printed build uses the DFRobot SEN0365 board (wiki.dfrobot.com/sen0365), the laser-cut build the DFRobot Gravity SEN0364 (wiki.dfrobot.com/sen0364).
  2. Pédelacq J-D, Cabantous S, Tran T, Terwilliger TC, Waldo GS. Engineering and characterization of a superfolder green fluorescent protein. Nature Biotechnology 24, 79–88 (2006). doi:10.1038/nbt1172
  3. Cree LED. Cree 5-mm Blue and Green Round LED, C503B-BAS/BAN/BCS/BCN/GAS/GAN/GCS/GCN, product family data sheet CLD-CT1094. PDF, hosted by Farnell
  4. Eppendorf. Eppendorf UVette, product page. eppendorf.com
  5. Szymula KP, Magaraci MS, Patterson M, Clark A, Mannickarottu SG, Chow BY. An open-source plate reader. Biochemistry 58, 468–473 (2019). doi:10.1021/acs.biochem.8b00952
  6. Beal J, Haddock-Angelli T, Baldwin G, Gershater M, Dwijayanti A, Storch M, et al. Quantification of bacterial fluorescence using independent calibrants. PLOS ONE 13(6), e0199432 (2018). doi:10.1371/journal.pone.0199432
  7. Silverman AD, Karim AS, Jewett MC. Cell-free gene expression: an expanded repertoire of applications. Nature Reviews Genetics 21, 151–170 (2020). doi:10.1038/s41576-019-0186-3
  8. Ayukawa S, Kobayashi A, Nakashima Y, Takagi H, Hamada S, Uchiyama M, et al. Construction of a genetic AND gate under a new standard for assembly of genetic parts. BMC Genomics 11 (Suppl 4), S16 (2010). doi:10.1186/1471-2164-11-S4-S16
  9. Kylilis N, Tuza ZA, Stan G-B, Polizzi KM. Tools for engineering coordinated system behaviour in synthetic microbial consortia. Nature Communications 9, 2677 (2018). doi:10.1038/s41467-018-05046-2

Every measured number on this page comes from our own dated engineering records.