CAPTURE-Screen is a single-cuvette fluorescence reader for GFP reporter cultures. It measures their green fluorescence directly at the bench, beside the incubator, without waiting for a shared plate reader, and we built it for our own AHL biosensor. We developed two versions of the device: a 3D-printed version and a laser-cut version, the laser-cut one in its own enclosure and the 3D-printed one on an open base plate under an opaque shield. The following sections describe the features shared by both designs, followed by a detailed overview of each version.
A GFP reporter is read as green light from a liquid culture, usually on a shared plate reader. Ours reports AHL, the quorum-sensing signal, and its assay set what this reader had to do: a small sample, read every hour, a comparison with a plate reader, and an instrument cheap enough to rebuild.
The need is not only ours. Green fluorescence from a GFP reporter in a liquid sample is among the commonest readings in synthetic biology: one interlaboratory study had 72 iGEM teams measure it in E. coli, on plate readers and flow cytometers6. Cell-free reactions, developed as portable sensors7, hold no cells at all. The reader is meant for samples like these: liquid, and not too dense. We tested the effect of sample density on the 3D-printed build. Both builds have read sfGFP in E. coli TOP10 culture, and both have read our AHL sensor strain without seeing its signal: the laser-cut build could not tell it from cells without sfGFP (the laser-cut build’s test), and the 3D-printed build read zero on its emission channel.
Both readers keep the blue that excites the sample apart from the green it gives off in two ways: by angle, and by colour. Neither uses an emission filter: the sorting by colour is the sensor's own.
| Part | X range (mm) |
|---|
This view places the 3D-printed build's optical core in the two-layer laser-cut enclosure as it was drawn, optics above and electronics below. No reading on this page was taken inside it. It was built, but not adopted: of the 11 counts that leak in room light, 4 come through the mechanism and 7 straight down the open cuvette slot, and an opaque shield over that slot brings the dark read to exactly zero (Light shielding). We chose the shield over a second enclosure. It is a schematic built from the optical coordinates, not from the CAD files: the printed parts themselves are under the 3D-printed build.
We built two readers on this principle. They share the light path and the spectral sensing described above. Each has its own mechanism, modelling and development history.
Choose a build to see it.
Counts from the two readers are not interchangeable: they run different firmware and integrate for different times.

A 3D-printed optical core on a base plate, with the electronics on a breadboard beside it.

A second reader on the same principle, made from laser-cut parts in its own enclosure.
It tells a culture with GFP from one without, using no emission filter, and its reading rises with the amount of GFP culture. It is quick to load, steady when the tube is put back, cheap to build, and made of parts that can be changed one at a time.
Not yet It has not resolved AHL induction, and it has not been tested with an emission filter or calibrated against a fluorescence standard.
How it got here: the same loop, twice.
Version 1
Version 2
Next turn An emission filter, in the drawer already on the reader.
A 3D-printed optical core on a base plate, with the electronics on a breadboard beside it.
It has been built twice. Version 2, in use since 2026-10-03, is the one the view opens on; version 1 took the earlier readings. Each printed part is drawn from its own STL file, where the CAD assembly puts it. Pull it apart, switch versions, or lay one module's parts side by side.
Every printed part, with the mesh of its STL file · grid 10 mm
Version 2 as designed: the assembly holds 38 printed pieces; the full print set is 52, the loose shims, inserts, disks and fit coupons included. The five replacement parts are in green, and all five are fitted on our reader. Pick a part for its size and its file.
Version 1 as designed: the assembly holds 45 printed pieces; the full print set is 59, the loose shims, inserts, disks and fit coupons included. The five in outline were specified but not on the instrument: a cap and a clip that never arrived, a second leaf that was never made, and two aperture disks that were never fitted. Pick a part for its size and its file.
Every file of the module, side by side as it was exported. Pick a part for its size and its file.
A drawing of the light path, slowed down.
iGEM's file server takes neither STL nor ZIP files, so the 53 STL files are kept as one glTF model, triangle for triangle: capture-screen-printed-parts-20261004.glb (1.7 MB). With JavaScript on, each file name below saves that part as an STL again.
excitation_cradle_v0_7.stl 195 KB
capture_screen_led_holder_v0_1_body.stl 259 KB
led_holder_body_v0_3.stl 292 KB
capture_screen_led_holder_v0_1_retainer_clip.stl 35 KB
capture_screen_aperture_baffle_body.stl 247 KB
aperture_baffle_body_v0_2.stl 146 KB
excitation_cap_LED_holder_v0_7.stl 79 KB
excitation_cap_LED_holder_v0_7_roof_down.stl 93 KB
excitation_cap_baffle_v0_7.stl 80 KB
excitation_locator_key_print_two_v0_7.stl 0.7 KB
aperture_disk_1p5mm.stl 111 KB
aperture_disk_2p5mm.stl 111 KB
aperture_disk_4mm.stl 99 KB
aperture_disk_6mm.stl 99 KB
uvette_holder_common_body.stl 308 KB
uvette_holder_body_v0_3.stl 215 KB
uvette_datum_riser_v0_7.stl 172 KB
uvette_seat_riser_v0_2.stl 207 KB
uvette_orientation_gate_A.stl 101 KB
uvette_orientation_gate_B.stl 101 KB
uvette_index_collar.stl 5.0 KB
uvette_collar_setting_gauge.stl 1.4 KB
uvette_preload_leaf.stl 2.2 KB
uvette_x_hold_down_clip_print_two_v0_7.stl 2.6 KB
uvette_y_hold_down_clip_print_two_v0_7.stl 2.6 KB
as7341_detector_body.stl 399 KB
as7341_pcb_carrier.stl 151 KB
as7341_pcb_retainer.stl 86 KB
detector_locator_frame_v0_7.stl 83 KB
detector_locator_pin_print_two_v0_7.stl 20 KB
detector_side_clamp_print_two_v0_7.stl 22 KB
common_cassette_body.stl 72 KB
blank_open_control_insert.stl 53 KB
round_12p5_insert_base.stl 126 KB
round_12p5_insert_cap.stl 49 KB
round_12p5_2mm_spacer.stl 49 KB
square_6x6_insert_base.stl 55 KB
square_6x6_insert_cap.stl 1.6 KB
replacement_filter_cassette_clamp_bar_print_two_v0_7.stl 42 KB
alignment_shim_1p5mm.stl 0.7 KB
alignment_shim_2p0mm.stl 0.7 KB
alignment_shim_2p5mm.stl 0.7 KB
optical_light_trap_angled_wedge_v0_1.stl 133 KB
light_trap_adapter_riser_v0_7.stl 126 KB
light_trap_clamp_plate_print_two_v0_7.stl 37 KB
optical_core_base_plate_v0_7.stl 563 KB
bench_foot_print_four_v0_7.stl 42 KB
locator_pin_print_four_v0_7.stl 20 KB
uvette_fit_coupon_v0_2.stl 293 KB
led_fit_coupon_v0_2.stl 692 KB
fit_coupon_interface_base_v0_7.stl 583 KB
fit_coupon_planar_clearance_base_v0_7.stl 212 KB
fit_coupon_slider_v0_7.stl 1.4 KB
The light is sorted without a filter, the seat places the cuvette, every read checks itself, and the parts change one at a time.
The detector looks at the sample from 90°, and its own colour bands keep the blue apart from the green. The two cultures read alike beside the source, 21.4 and 21.7 counts on F2; only the green differs. Dense suspensions are where this stops.
Measured, one session · every channel of both readsThe cuvette drops between eight ribs onto four corner tabs, which set its position and height. One tube, put back five times: 0.76% between insertions.
By design; timed and measured on version 2Every flash sits between two reads with the LED off. Shielded, those are exactly zero, so one that is not voids its cycle: eight in version 1's logs, one in the dilution series.
In the firmware; counts from our logsThat is all version 2 changed. The base plate, the detection side, the light trap and the firmware stayed as they were.
What can be changed, drawn below.
Done once, for version 2What has been changed, what is on the reader but not yet used, and what has not been tried.
Our wet lab used both versions, at least three of its members in all. What they said about the first set the brief for the second. Two external advisors also reviewed the design.
They told us Placing the cuvette took too many steps, each one a chance to seat it differently, and a measurement took a long time.
What followed Version 2's seat takes the cuvette in one step and sets its height and position itself.
Time to place the cuvette, version 1 against version 2, timed. Each range runs from quicker hands to slower ones.
They told us Convenient and quick: the cuvette only has to be put in, which leaves little room for operator error. Seeing the taped repairs replaced by new parts, they trusted the reader more. The arrow engraved on the seat also helped: it shows how the cuvette should face.
What followed They made that session's five re-insertions themselves, and every reading stayed within one count of 50.
They told us The card was clear to follow. One of them went through it in order and met no problem. The other met REJECTED sample_id, and the remedy the card gives for it was enough to deal with it.
What followed The Run card stays as they used it, with its list of what to do when the reader says something else.
They told us Prof. Tsung-Lin Liu asked what mattered most in the hardware, and suggested trying a wider range of sensors. Prof. Nan-Si Li placed the reader as one step of a path from sample to answer, behind a front-end unit that prepares the specimen.
What followed We set three priorities for the hardware: low cost, speed and convenience. We did not add a second sensor: the AS7341 already sorts the emission by colour, so another would have meant buying a filter to do the same job. The front-end unit is not built (Human Practices).
Use turned up seven more problems, from room light getting in to a detector that lost its settings.
One step to load, a steadier reading, more light with the same zero, and fainter samples in reach.
All four come from version 2, fitted on 2026-10-03. We redrew five printed parts around what the wet lab had found slow, and around what had broken, gone missing or sat loose in version 1; and we changed one resistor to bring the LED up to its design current.
Each dot shows one insertion relative to the tube's own mean; the percentage is the standard deviation divided by that mean. Version 1's tubes were put back by the reader's builder, version 2's by wet-lab members.
Observed, in different sessionsAn empty cuvette scatters about twice as much. The emission channel stays at zero.
MeasuredGFP culture at 0.19 on our concentration index on version 2; on version 1, at 0.19 or less. The index is the stock's displayed optical density (OD) times its dilution.
Observed, with different culture batchesPart by part The re-insertion numbers The LED current
The first card, played out in photographs: version 1's four steps, taken from frames of a video we filmed, and version 2's one.
It tells a GFP culture from one without, and the reading rises with the amount of GFP culture; a plate reader beside it ranked the same tubes the same way. Without GFP it read zero in liquid culture, and a few counts in dense suspensions. It has not yet resolved AHL induction.
The GFP cultures read in this section are a test strain: E. coli TOP10 with a plasmid that makes sfGFP without induction. They stood in for our AHL biosensor, which we expected to be about as bright once induced. That induction has not been shown conclusively: in liquid culture both of its constructs fluoresced above an empty-vector control, with no uninduced culture beside them to compare (Engineering).
Every tube without GFP read zero. From 25 to 200 µL the readings lie close to a proportional line; below 25 µL they fall below it.
Version 2 read one batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls.
The smallest amount read above zero on every cycle was 12.5 µL of GFP culture in 200 µL, here and in the session beside a plate reader, on a different batch. At half that, 4 of 18 cycles read above zero.
Not yet A limit of detection: this was one session, on one batch of culture.
The dilution series, tube by tube What the wet lab said
Read beside a plate reader on the same tubes, it ranked them the same way, and over the three denser tubes its reading per unit of OD held within 10%.
We diluted one batch of the sfGFP test culture to five strengths, from 6.25% to 100%, and read every tube on our reader and on a SpectraMax M3 plate reader, which also measured its OD600. As the culture got stronger, our green reading rose from 2.44 to 86.44 counts, keeping close step with the M3’s OD600 (r2 = 0.993 across the five tubes). A culture without GFP, at an OD600 of 0.72, read zero on ours and close to background on the M3. Even the thinnest tube, at an OD600 of 0.07, gave a reading on every cycle; we did not try anything thinner.
Not yet This was one culture on one day, and it is not a calibration. Whether the two instruments’ fluorescence readings are in proportion is still open.
Beside the plate reader, tube by tube
Zero on the emission channel for all 27 readings without GFP, above zero for 35 of the 38 tubes with it.
Three mixing sessions each combined a GFP stock, a non-GFP stock and medium at set fractions into 200 µL tubes.
In the two sessions with enough signal, the reading rose in proportion to the GFP culture. How it scales, and why the three sessions differ, is worked through on the Model page.
The three mixing sessions in numbers
Under a shield the dark read is exactly zero, so the zeros above are readings, not corrections.
Under the shield, every channel reads zero while the LED is off, even at 25 times the usual integration time. In room light the leak does not scale with integration time, so subtracting a dark reading would not remove it.
With the LED on, plain medium has read zero on the emission channel every time on version 2: eight blank readings across four sessions, 60 valid cycles in all.
The operating rule: Measure in a darkened room or under an opaque shield. Because a shielded dark read is exactly zero, a dark read that is not zero marks that cycle as invalid: it is dropped and its raw row kept. The firmware applies this rule itself: it voided eight cycles in version 1's logs and one in version 2's dilution series.
In a dense suspension the reader gives a few counts with no GFP there. A culture without the reporter read zero while the scatter channel stayed under about 33 counts, and no longer did from 46.
We met this on suspensions scraped from AHL plates, the samples we were given for an AHL test, in the same E. coli TOP10. The wet lab grew culture on plates containing AHL at five concentrations, 0.22 to 1.1 µM by its own estimate, scraped each plate into suspension and gave us 200 µL of it. By eye it judged every plate induced but the 1.1 µM one. The reading followed how much culture was in the cuvette, not that judgement.
Two days later we took a culture without the reporter gene, scraped from one plate, diluted it in LB and read each dilution: the emission channel for what the culture itself gives, and the scatter channel of the same read beside it. LB alone read zero. Reading zero means below what this reader can register, not that nothing is there.
Where the two days overlap, the readings from plates judged induced lie within about a tenth of a count of a curve drawn through the readings without the reporter. sfGFP added to a dense suspension was still read: 23.56 counts, against 5.22 for that suspension alone.
Not yet This was one culture on one day, so the scatter figures are not a specification.
Using version 1 changed the seat. These two sessions point to the next change: an emission filter.
What followed Every read already reports the scatter channel beside the emission channel, so the two can be read together. In the density session a culture without the reporter gave a few counts of F4 once F3 passed about 46, and none under about 33: at that scatter, a few counts alone are not a sign of GFP.
Still open An emission filter. The reader has a drawer for one in front of the detector, and it has run empty. Neither session could say whether the reading without GFP is scattered blue or green from the sample itself; a filter that passes green and blocks blue is the test that could tell them apart.
The AHL plate session, cup by cup The density session, reading by reading The background, modelled
Every channel of a read, for the states we measured. Choose a test and throw its switches.
Top view, schematic · both versions, named by session
The channel readout needs the page script. Without it, the light path is still drawn, and the sessions the readings come from are named earlier in this section.
Counts are raw sensor values. F4, outlined in green, is the emission channel: the sensor's 515 nm band, where GFP emits.
Print it, wire it, and run it under a shield. Build version 2.
Version 2 is 38 printed pieces, specified in matte black PLA; the full set of 52 adds loose shims, filter inserts, aperture disks and two fit coupons. Each can be saved as an STL file from the 3D view. The five new parts print without supports; on another printer, print the two fit coupons first.
A 2N7000 MOSFET switches the 470 nm LED at about 10 mA through a 217 Ω resistor, and the AS7341 detector sits on the ESP32's I²C bus. Keep the gate pulldown: without it the LED can light while the controller boots.
Any serial terminal at 115 200 baud, one command per line, under an opaque shield. This is one session from start to finish; the reader's answers are from the log of the dilution series on version 2, 2026-10-05.
The sketch ships with its LED gate off, and built that way the LED never lights. Build it with the gate on, then upload it.
arduino-cli compile --fqbn esp32:esp32:esp32doit-devkit-v1 --build-property "compiler.cpp.extra_flags=-DCAPTURE_LED_HARDWARE_ENABLE=1" firmware/esp32_fluorescence_reader
# hw_enable(compile)1 # sensor_ready 1 # settings_readback 1
Go on only if all three read 1.
dark 0LED off 0dark 0NO_SIGNAL;LED_DISABLED
It stops after one cycle, as it should. All 36 values must be 0.
# LED ARMED
dark 0F4 0, 0, 0dark 0OK
# sample_id=LOW_BLK_1
dark 0F4 0, 0, 0dark 0OK
dark 0F4 56, 57, 55dark 0OK
Three cycles of dark, lit, dark. This tube: 100 µL of GFP culture in its 200 µL.
# LED DISARMED
dark is all twelve values of a read, before and after each
flash: the sensor's ten channels, with Clear and NIR read twice.
F4 is the three cycles with the LED on. Each cycle prints
one CSV row of 66 fields: keep the terminal's whole log. Ours was
kept by a script that stamps and stores every line; it is in the
repository as
firmware/host/serial_read_tool.ps1.
AMBIENT_PRESENT# REFUSED: ENABLE is not in a defined acquisition stateid
shows settings_readback 0. Reset the
controller once. If that fails, unplug the USB for five
seconds, then check the detector's wires: SDA on GPIO21,
SCL on GPIO22, and 3V3.
Seen three times on version 2, on 2026-10-03 and 2026-10-09. A reset cleared it each time.
NO_SIGNAL;LED_DISABLEDfault=NOT_ARMED: arm and
run again. fault=HW_ENABLE_COMPILED_OFF: the
sketch was built with its gate off.
# REJECTED sample_id: longer than 24 chars. _ - +.
A rejected name leaves the last one in force: read the echo
before run.
SATURATEDflag=OK with F4 at 0
The CAD sources and exports, the firmware and the parts list are published under the MIT licence in the team's repository on iGEM GitLab: capture-screen-hardware.
A second sketch runs the same reader from
LasReader, our
software, over Wi-Fi, in the protocol the laser-cut build uses:
firmware/capture_screen_3d.
It drives the LED gate from GPIO26, not GPIO25: move the gate wire, or
change LED_PIN. Run this way, beside the laser-cut
build, it read our AHL sensor strain: F4 was zero throughout, and the
samples were not dense enough to scatter into it. Those reads were
not kept, so no numbers are given here; every reading on this page
with a number was taken with the serial firmware.
Not yet The assembly order was checked by the reader's builder. Nobody has yet built a second reader from it.
It cost us about US$161 to build, most of it printing bought from a shop. With a printer of your own it is about US$38 and 180 g of PLA.
Every table behind the sections above, and what a second team needs to build and run the reader.
The page calls each experiment by a name. This is the list: what was done, on which version of the reader, and the date its records carry.
| Experiment | Reader | What was done | Date |
|---|---|---|---|
| First light | Version 1 | The LED on a paper scatterer at the sample position, and with nothing in the beam. | 2026-08-25 |
| Room-light test | Version 1 | An empty cuvette with the LED off, in room light and shielded. | 2026-08-27 |
| First fluorescence | Version 1 | A culture reported to carry sfGFP and its control, twelve cycles each. | 2026-09-03 to 09-04 |
| Stability read | Version 1 | One tube left in its seat and read 20 times over 5 minutes. | 2026-09-03 |
| Re-insertion test 1 | Version 1 | One culture, four insertions, one of them a fresh draw. | 2026-09-09 |
| Re-insertion test 2 | Version 1 | The corresponding culture five days later, read the same way. | 2026-09-14 |
| Mixing session 1 | Version 1 | A GFP stock, a non-GFP stock and medium mixed at set fractions into 200 µL tubes. | 2026-09-16 |
| Mixing session 2 | Version 1 | The same mixing, with a thinner stock. | 2026-09-21 |
| Mixing session 3 | Version 1 | The same three liquids, laid out to test a recipe difference and proportionality. | 2026-09-29 |
| LED current test | Version 2 | One empty cuvette, read before and after the current-limit resistor was changed. | 2026-10-03 |
| Dilution series | Version 2 | One batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls. Re-insertion test 3, one tube put back five times, was part of it. | 2026-10-05 |
| AHL plate session | Version 2 | Suspensions scraped from plates at five AHL concentrations, with medium, a culture without the reporter and an sfGFP culture as controls. Re-insertion test 4, one tube put back ten times, was part of it. | 2026-10-09 |
| Density session | Version 2 | A culture without the reporter, two suspensions scraped from one plate, diluted in LB: 21 readings, with LB as the blank. | 2026-10-11 |
| Plate reader comparison | Version 2 | One batch of the sfGFP test culture at five dilutions, with medium and a culture without GFP, read on our reader and on a SpectraMax M3. | 2026-10-12 |
| Session | Readings without GFP | F4 above zero | Tubes with GFP | F4 above zero |
|---|---|---|---|---|
| Mixing session 1 | 16 | 0 | 10 | 10 |
| Mixing session 2 | 8 | 0 | 12 | 9 |
| Mixing session 3 | 3 | 0 | 16 | 16 |
| All three | 27 | 0 | 38 | 35 |
Samples without GFP read zero on the emission channel every time, including undiluted non-GFP culture at a displayed OD of up to 1.42. The channels beside the excitation band do see those cells, but too variably between cuvettes and sessions to build a turbidity flag on.
| What was repeated | n | Mean F4 | Std dev | % of mean |
|---|---|---|---|---|
| One liquid, re-inserted (mixing session 1) | 3 | 6.00 | 0.33 | 5.6 % |
| One recipe, prepared three times (mixing session 1) | 3 | 6.89 | 0.51 | 7.4 % |
| One recipe, six tubes (mixing session 3) | 6 | 9.33 | 1.10 | 11.7 % |
| One culture, four insertions, one a fresh draw (re-insertion test 1) | 4 | 40.50 | 1.23 | 3.0 % |
| One culture, four insertions, one a fresh draw (re-insertion test 2) | 4 | 32.67 | 1.83 | 5.6 % |
| What went wrong | Session | What changed |
|---|---|---|
| An empty cuvette read F4 = 11 under room light. | Room-light test | Added the shield box. A dark read that is not zero voids the cycle. |
| A cuvette's number did not match what was in it, and one record had to be voided. | Mixing session 2 | Confirm what is in the cuvette before starting a run. |
| The stock's displayed OD did not scale with dilution. | Mixing sessions 1 and 3 | Recommended: read the OD at one fixed dilution, 1:4. Not yet in the operating card. |
| Washed and reused cuvettes tied “which use of this cuvette” to the order of reading. | Mixing session 3 | Recommended: one cuvette per tube. Not yet in the operating card. |
| Suspensions scraped from plates were dense enough to raise the emission channel without GFP. | AHL plate session | The density session found where a culture without the reporter stops reading zero, in the scatter channel. Recommended: check F3 on every sample. Not yet in the operating card. |
| A read stopped partway: the detector timed out once, and the bus to it failed twice. | Density session | A reset and a second read of the same cuvette. The cause has not been found. |
| The firmware refused a run because the detector had lost its settings. A reset cleared it each time. | Mixing session 3, twice on the day version 2 was fitted, and once at the end of the AHL plate session | The Run card now says what to do after a refusal. |
The last two rows are the firmware doing its job: when its state is not known it declines to measure. Four refusals point at a loose connection to the detector board. The cause has not been found. In the 28 captures of the dilution series there was no such refusal; the firmware's only stop was an opening check left unarmed on purpose. That is an observation, not a fix.
200 µL to a tube. The sensor settings were version 1's. The LED ran through the new resistor, and its current was not measured that day.
| GFP culture in the tube | Tubes | F4, each tube | Mean |
|---|---|---|---|
| None: medium alone | 2 | 0, 0 | 0 |
| None: non-GFP culture, undiluted | 2 | 0, 0 | 0 |
| 6.25 µL | 2 | 0.44, 0.11 | 0.28 |
| 12.5 µL | 3 | 3.33, 3.00, 3.00 | 3.11 |
| 25 µL | 2 | 12.67, 14.44 | 13.56 |
| 50 µL | 2 | 23.00, 26.00 | 24.50 |
| 100 µL | 2 | 56.00, 50.33 | 53.17 |
| 200 µL, undiluted | 1 | 98.33 | 98.33 |
Every tube without GFP read zero on the emission channel, the non-GFP culture included, as on version 1. From 25 to 200 µL, a range picked after the readings were seen, the seven tubes lie within −8.8 % and +14.5 % of one straight line through the origin.
Below that the reading falls short of the line: 12.5 µL read about half of what the line gives and 6.25 µL about a tenth, with a count of one or more on only 4 of its 18 cycles. That is the shape a floor in the reading would give, but one session does not tell a floor from other causes.
Version 1 had read zero for three of the four tubes at or below 0.19 on our concentration index, the stock's displayed OD scaled by its dilution in the tube, all in mixing session 2; 12.5 µL here is that same 0.19. The index leans on the displayed OD, and this session's stock read 0.77 at 1:4, above the 0.13 to 0.71 of the earlier sessions.
The 12.5 µL figure rests on the redone tube: with the first capture kept in its place, the planned analysis puts that level at 0.82 of proportional instead of 0.48, and the pooled form of that analysis no longer resolves a low-end shortfall. The 6.25 µL readings do not depend on it.
One 100 µL tube was taken out and put back five times, by wet-lab members. The five means ran from 49.33 to 50.33 counts, with a standard deviation of 0.38 counts between them: about the third of a count that a three-cycle mean can resolve. With the seat empty, every channel read zero.
The sfGFP test culture, E. coli TOP10 with a plasmid that makes sfGFP without induction, diluted in medium. On our reader, 200 µL to a cuvette and nine cycles each, at the sensor settings of the dilution series; no tube was read twice. On the SpectraMax M3, OD600 and fluorescence at 485 / 515 nm.
| In the tube | F4, ours | Cycles | OD600, M3 | F4 per OD | M3 fluorescence, undiluted | M3 fluorescence, diluted 20× |
|---|---|---|---|---|---|---|
| Medium, at the start and at the end | 0, 0 | 0–0 | 0 | – | 0 | – |
| Culture without GFP | 0.00 | 0–0 | 0.721 | – | 249 | – |
| sfGFP culture, 6.25 % | 2.44 | 2–3 | 0.069 | 35 | 8,216 | not read |
| sfGFP culture, 12.5 % | 7.44 | 7–8 | 0.125 | 60 | 13,249 | 326.5 |
| sfGFP culture, 25 % | 33.22 | 33–34 | 0.302 | 110 | 25,249 | 1,147.6 |
| sfGFP culture, 50 % | 48.33 | 47–49 | 0.475 | 102 | not obtained | 1,746.2 |
| sfGFP culture, 100 % | 86.44 | 84–89 | 0.861 | 100 | not obtained | 3,802.1 |
One question was registered before the session: is our F4 in proportion to the M3's fluorescence, each tube's ratio within 20% of their mean? It could not be judged. The M3 could not read the denser tubes undiluted, so only 6.25%, 12.5% and 25% have a fluorescence reading of the registered kind. Across those three the ratio ran 0.298, 0.562 and 1.316 (×10−3): not in proportion, the thinner tubes reading low on ours, as the dilution series had led us to write down beforehand.
Worked out afterwards: F4 against the M3's OD600 gives r2 = 0.993 over five tubes, against its diluted fluorescence 0.982 over four, against its undiluted fluorescence 0.981 over three, and the tubes rank the same each time. These are a few points of one dilution series, where r2 runs high: they show one trend, not accuracy.
The M3's values were copied down by the operator; there is no export file, the number of wells and the gain were not recorded, and some values have two significant figures. The two instruments excite at different wavelengths, 470 and 485 nm, so only the trend can be compared, not the size of the numbers.
200 µL to a cuvette, nine cycles each, at the sensor settings of the dilution series. Sorted by F3. There was one cuvette to a condition and no plate without AHL, and how much was scraped was not measured: the session gives no dose–response and cannot say whether the plates were induced.
| In the cuvette | By eye | F3 | F4 |
|---|---|---|---|
| Medium, at the start and at the end | – | 0, 0 | 0, 0 |
| sfGFP culture, 100 µL, in 100 µL of medium | control | 22.89 | 66.56 |
| 1.1 µM plate, the thinner scrape | not induced | 56.78 | 1.00 |
| 0.35 µM plate, diluted by half | induced | 59.56 | 1.11 |
| Culture without the reporter | – | 64.44 | 1.78 |
| 0.22 µM plate | induced | 73.78 | 2.44 |
| 1.1 µM suspension, 180 µL, with 20 µL of sfGFP culture | control | 85.89 | 23.56 |
| 0.275 µM plate | induced | 94.00 | 4.44 |
| 1.1 µM plate, the denser scrape | not induced | 97.89 | 5.22 |
| 0.55 µM plate, first reading | induced | 102.11 | 4.89 |
| 0.35 µM plate | induced | 103.44 | 5.22 |
| 0.55 µM plate, read again | induced | 104.44 | 5.22 |
Medium read zero on the emission channel in all 18 cycles. The culture without the reporter did not: 1.78, at a density no earlier session had reached. Whether its cuvette had held culture before was not recorded.
Set side by side after the readings were seen: among the five readings with F3 from 94 to 104, the plate judged not induced read 5.22 and those judged induced 4.44 to 5.22. Among the three with F3 from 57 to 64, induced read 1.11, not induced 1.00, and the culture without the reporter 1.78.
The sfGFP culture, at a quarter of that scatter, read 66.56. Twenty microlitres of it in 180 µL of a dense suspension read 23.56, against 5.22 for that suspension alone: the dense culture did not hide it.
One test was registered before the session: that F4 rises in the order of the AHL concentrations. Read as registered, it does not hold (p = 0.64). Our script first reported that it did (p = 0.05), having used two repeat readings the registration rules out: both were asked for after the first reading was seen, and the 1.1 µM one was a second, denser scrape. Neither outcome bears on AHL. The same test on F3 gives the same p values, so what it ranks is how much was scraped.
200 µL to a cuvette, nine cycles each, at the sensor settings of the dilution series. A culture without the reporter gene, diluted in LB; per cent is the share of suspension. Sorted by F3.
| In the cuvette | F3 | Cycles above zero | F4 |
|---|---|---|---|
| LB, at the start and at the end | 0, 0 | 0 of 18 | 0, 0 |
| First suspension, 10 % | 5.33 | 0 of 9 | 0.00 |
| First suspension, 10 %, second pass | 6.44 | 0 of 9 | 0.00 |
| First suspension, 20 % | 12.33 | 0 of 9 | 0.00 |
| First suspension, 20 %, second pass | 12.44 | 0 of 9 | 0.00 |
| First suspension, 30 % | 13.89 | 0 of 9 | 0.00 |
| First suspension, 30 %, second pass | 15.11 | 0 of 9 | 0.00 |
| Second suspension, 30 %, added | 26.11 | 0 of 9 | 0.00 |
| Second suspension, 20 %, added, put back | 26.78 | 0 of 9 | 0.00 |
| Second suspension, 20 %, added | 28.78 | 0 of 9 | 0.00 |
| Second suspension, 30 %, added, read again | 29.56 | 0 of 9 | 0.00 |
| Second suspension, 35 %, added, put back | 32.33 | 0 of 9 | 0.00 |
| Second suspension, 35 %, added | 33.11 | 0 of 9 | 0.00 |
| Second suspension, 40 % | 41.78 | 0 of 9 | 0.00 |
| Second pass, the cup taken to be 50 % | 46.00 | 1 of 9 | 0.11 |
| Second pass, the cup taken to be 40 % | 50.78 | 5 of 9 | 0.56 |
| Second suspension, 50 % | 56.00 | 7 of 9 | 0.89 |
| Second suspension, 70 % | 75.11 | 9 of 9 | 2.67 |
| Second suspension, 70 %, second pass | 78.56 | 9 of 9 | 2.89 |
| First suspension, undiluted | 82.33 | 9 of 9 | 2.67 |
| Second suspension, undiluted, read again | 86.22 | 9 of 9 | 3.89 |
| Second suspension, undiluted | 88.67 | 9 of 9 | 3.89 |
The first suspension ran out after its 30 per cent tube, and the wet lab scraped a second from the same plate. The two do not line up: 30 per cent of the first scattered 14 to 15 counts, 40 per cent of the second 42 to 51. That is why the series as planned has no tube between F3 15 and 42.
The rule was registered before the session: take the densest tube whose cycles are all zero in both passes, and the thinnest with any cycle above zero. That gives 30 per cent of the first suspension, F3 14.50 over 18 cycles, and 40 per cent of the second, F3 46.28: zero at 41.78 in its first pass, 5 of 9 cycles above zero at 50.78 in its second. The six readings marked added were made after that result was seen, to fill the gap, and are not part of it. In the second pass the cups were told apart by eye, and the 40 and 50 per cent ones may have been exchanged; the table places each reading by its own F3.
Joined into a curve that is only assumed never to fall as scatter rises, these readings give the background of a culture without the reporter at any scatter up to F3 89, and the AHL plate session can be set against it. Its two readings without GFP that fall in that range lie +0.04 and +0.10 counts from the curve; the two from plates judged induced, −0.11 and −0.10. The four densest induced readings, F3 94 to 104, lie beyond the curve: if the background does not fall there, at most 0.56 to 1.33 counts are left. The cuvette with sfGFP added stands 19.76 above it. This was worked out after both sessions, with the two days' counts taken as they are; the curve is one culture on one day, not a correction to subtract. More on the Model page.
One plate, one day, one set of detector settings and no OD: the F3 figures are not a specification and do not carry to another reader or another day. How much GFP signal is left in a sample that thin was not measured.
The firmware stopped three captures partway, once on a detector time-out and twice on a bus error, each before a reading had been shown. The same cuvette was read again after a reset.
Version 1 held the cuvette against one spring leaf where the design called for two, which left it free across about 0.6 mm on one axis. Version 2's eight ribs leave 0.05 mm a side, as drawn. Taking one tube out and putting it back is the test of that. On version 2 the five insertions were made by wet-lab members, not by the member who built the reader.
| What was done | Reader | n | Mean F4 | Spread (SD) | % of mean |
|---|---|---|---|---|---|
| Left in its seat, read 20 times over 5 minutes (stability read) | Version 1 | 20 | 21.95 | 0.29 | 1.3 % |
| Taken out and put back (re-insertion test 1) | Version 1 | 3 | 40.33 | 1.45 | 3.6 % |
| Taken out and put back (re-insertion test 2) | Version 1 | 3 | 33.00 | 2.08 | 6.3 % |
| Taken out and put back (re-insertion test 3) | Version 2 | 5 | 49.87 | 0.38 | 0.76 % |
| Left in its seat, read 20 times (AHL plate session) | Version 2 | 20 | 62.95 | 1.13 | 1.8 % |
| Taken out and put back (re-insertion test 4) | Version 2 | 10 | 68.37 | 1.84 | 2.7 % |
In counts, putting a tube back on version 1 moved the reading five to seven times as much as leaving it alone. On version 2 it moved it little more than leaving it alone had: 0.38 counts on a mean of 49.87, or 0.76 %.
Test 4, in the AHL plate session, put an sfGFP culture back ten times: 2.7 %, against 1.8 % for the same tube left alone. That culture settles. Left alone its reading fell, and handling stirred it up again, so the wider spread cannot be put down to the seat. A first resting series, which the operator may have bumped, spread 3.3 %.
Version 1 ran the LED at 5.553 mA, under the 8 to 12 mA the design aims for. On 2026-10-03 one 220 Ω resistor replaced the series pair, which brings it to about 10 mA. With an empty cuvette, the two channels beside the source then read about twice what they had, and the emission channel still read zero. Two days later the same held with samples in the beam: medium, a non-GFP culture and the empty seat all read zero on the emission channel, while that culture scattered up to 40 counts into the channel next to the source. The scatter off an empty cuvette roughly doubled and the emission channel stayed at zero in every one of these reads, which is what the 90° layout and the light trap are for.
| Quantity | R1 = 434 Ω | R1 = 217 Ω |
|---|---|---|
| In use | 2026-08-25 to 2026-10-03: every version 1 reading | Since 2026-10-03 |
| Made of | 218 Ω + 216 Ω in series | One 220 Ω resistor, 217 Ω measured |
| LED current | 5.553 mA, measured 2026-08-25 | 9.2 to 9.9 mA by meter; at most 11.3 mA by calculation |
| If the LED shorted | 11.89 mA. The resistor alone holds it under our 15 mA ceiling | About 24 mA. It no longer does |
| Empty cuvette, F2 / F3 / F4 | 7.89 / 6.89 / 0 | 18.33 / 13.33 / 0 |
Limits It is scattered excitation on an empty cuvette, not a fluorescence result, and it does not show that 10 mA is the right current. Readings at the two resistances cannot be pooled: the empty-cuvette ratio is not a conversion for fluorescence. The meter readings were taken during 0.55 s pulses, so they are a lower estimate.
| Reader | Read | F2 | F3 | F4 |
|---|---|---|---|---|
| Version 1 | Seven sessions, first fluorescence to mixing session 3 | 7.33 to 11.89 | 6.33 to 10.22 | 0 |
| Version 2, as first fitted | Four insertions | 6.58 | 5.42 | 0 |
| Version 2, after an adjustment | Four insertions | 7.75 | 6.71 | 0 |
| In version 1 | In version 2 |
|---|---|
| In use from 2026-08-25 to 2026-10-03. 57 of its 59 printed pieces arrived, and two of those broke. | In use since 2026-10-03, with all five replacement parts fitted. |
| The LED holder split along its clip slot and the baffle broke at both disk slots as the print supports came off. Both were taped. | No clip slot, and a baffle that prints on its side, so no layer loses most of its area. The weakest layer keeps 0.82 of the fullest section within 3 mm of it; version 1 broke at 0.63 and 0.32. |
| The clip that retains the LED never arrived. Friction and its own leads held it. | Three ribs in the holder grip the LED, 0.10 mm tight. No clip. |
| The cap that clamps the LED holder never arrived. The holder lay in its saddle. | The cap is printed, turned over so that it needs no support. |
| One preload leaf instead of two left the cuvette free on one axis, by up to about 0.6 mm. | One body with eight ribs, 0.05 mm clear on each side, replaces six pieces. The cuvette lands on four corner tabs, 8.5 mm below the beam. |
| The LED ran at 5.55 mA, under the 8 to 12 mA the design aims for. | About 10 mA, with one resistor in place of two. See the circuit. |
| Function | Part | State |
|---|---|---|
| Excitation LED | Cree C503B-BCN-CV0Z0461 | Received; bench characterised |
| Detector | DFRobot SEN0365 (AS7341) | Received; bench bring-up passed |
| Controller | ESP32-WROOM-32, DOIT DevKit V1 30-pin | In bench use |
| LED switch | 2N7000 N-channel MOSFET | Conditional bench acceptance, single unit |
| Current limit / gate network | 217 Ω current limit (434 Ω for every version 1 reading), 217 Ω gate, 9.9 kΩ pulldown | Measured in place |
| Sample vessel | Eppendorf UVette | Engineering benchmark, not the final vessel |
| Optical core | Printed PLA: version 1's geometry, with five replacement parts for version 2 | Rebuilt 2026-10-03 as version 2, with all five fitted. How closely they fit was not measured |
| Fasteners | 36 M3 screws, 8 M3 nuts and 4 washers | Lengths under Assembly, step by step |
| Shield | Opaque cardboard box | Interim; every measurement needs it or a darkened room |
Build version 2. Version 1 was never a complete set, for the reasons under From version 1 to version 2, and its alignment was shown to work but never quantified.
The printed parts are in the 3D view. Each part there carries its size and can be saved as an STL file.
| Part | Replaces | Size (mm) | On the print bed |
|---|---|---|---|
| UVette holder body | The old holder body, two gates, collar, gauge and preload leaf | 36 × 36 × 25.5 | Square base down |
| UVette seat riser | Version 1's riser | 43 × 42.5 × 13 | Large face down |
| LED holder body | The old holder body and its retainer clip | 21.6 × 21.6 × 21.3 | Flanged end down, cylinder upright |
| Aperture baffle body | The old baffle body | 15 × 16 × 13 | Flattened face down, bore horizontal |
| LED-holder cap | The cap version 1 never received | 6 × 30.6 × 12.45 | Flat face down, arch opening up |
All five print without supports in the orientation given. The print was specified as matte black PLA, which has to be opaque, on a 0.4 mm nozzle with 0.20 mm layers and 100 % infill, and nothing scaled or drilled afterwards. The two fit values came from coupons printed on one machine. On another printer, print the coupons first.
| Item | Cost (US$) | Basis |
|---|---|---|
| Version 1's print set, test pieces included | $36.30 | Paid to a print shop |
| Version 2's two fit coupons | $26.52 | Paid to a print shop |
| Version 2's five parts | $59.66 | Paid to a print shop |
| ESP32 development board | $6.79 | Our costing sheet |
| AS7341 sensor board | $27.15 | Our costing sheet |
| Breadboard and jumper wires | $3.00 | Our costing sheet |
| LED, MOSFET and three resistors | $1.20 | Distributor list prices, 2026-10-07 |
| Total, estimated | $160.62 | Of which printing is $122.48 |
The mechanical source is parametric CadQuery Python with STEP and STL exports; the firmware is an ESP32 sketch with its own minimal AS7341 driver, and needs no detector library.
| Artefact | Format |
|---|---|
| Optical core geometry | Parametric CadQuery Python: version 1, and the five version 2 parts |
| Printable parts | STL: version 1's print set, 46 files for 59 pieces (the assembly's parts plus loose pieces such as shims, inserts and fit coupons); five files for version 2; two version 2 fit coupons |
| Assembly geometry | STEP |
| Acquisition firmware | ESP32 sketch with its own AS7341 driver, P1-ACQ-1.0.0 |
| Bill of materials | CSV |
| Bench circuit | Schematic with measured values |
| Optical coordinates | Every part and its X extent |
| Working sensor settings | AGAIN 512×, ATIME 255, ASTEP 280, about 200 ms per integration |
| Operating rule | Darkened room or opaque shield |
The circuit, the coordinates, the sensor settings and the operating rule are on this page, and every printed part can be saved as an STL file from the 3D view. The CadQuery source, the STEP and STL exports, the firmware and the bill of materials are in the team's repository on iGEM GitLab, capture-screen-hardware, under the MIT licence. It holds every part the 3D view draws, for both versions. The order of assembly is in the next row.
This is the order the design gives, and the reader's builder has checked it against how ours was put together. In this order a screwdriver can reach each of the 26 screws on the base, which the CAD record checks one by one. Pull the 3D view apart to see each part named here.
Every screw is an M3. Most cut their own thread in a printed pilot hole 2.7 mm wide: stop when the screw is snug, because PLA strips. Eight pass through clearance holes in both the part and the plate, and take a nut underneath.
| Where | Screws | What sets the length |
|---|---|---|
| Feet | 4 × M3×8 | From above, through the 4 mm plate into the foot |
| Risers | 4 × M3×12 | From under the plate, 8 mm into a 9.5 mm pilot |
| Cuvette holder clips | 4 × M3×12 | Through the 4.9 mm clip into an 8 mm pilot |
| Light-trap clamp plates | 2 × M3×12 | Through the clamp plate and the trap's flange into an 8 mm pilot |
| LED-holder cap | 2 × M3×16 | Through the 12.45 mm cap into a 5 mm pilot |
| Baffle cap | 2 × M3×14 | Through the 10.25 mm cap into a 5 mm pilot |
| Cradle; detector frame ears; detector side clamps | 8 × M3×10, each with a nut | 3 mm of part and 4 mm of plate, and it must stop short of the bench, 4.5 mm below |
| Inside the detector holder | 4 × M3×16; 4 × M3×8 to 10 with washers; 2 × M3×6 to 8 | The cassette's clamp bars, the board carrier and the board retainer, as the detector's own notes give them |
These are the notes that went with the parts. They were written at design time and are not yet a record of an assembly that was checked step by step.
There are no photographs of the assembly: the reader was already put together when this page was written. The 3D view can be pulled apart to show where each part goes.
The excitation branch turns one LED3 on and off at a current that cannot damage it: a resistor-limited LED with an N-channel MOSFET on the low side, switched from one pin of the controller. It is a breadboard circuit that was measured, not a product design. The gate pulldown, R3, is the part that is not optional: without it the LED can emit while the controller boots, during what the firmware takes for a dark baseline.
Hover or focus a labelled node for its measured value
Current-limit resistor R1
One reading is not instantaneous. The LED turns on, the optics and the sensor are given time to settle, the sensor integrates twice to cover all its channels, and the LED turns off again. The firmware holds a hard 1.0 s ceiling on that window.
Measured 2026-08-25 with a diagnostic sketch, one integration per channel set
The window fits the 1.0 s hard timeout with 448 ms of margin and misses our own 0.5 s target by about 52 ms, which we have recorded as a missed target. That was measured with a diagnostic sketch. The project firmware has since reported 550 to 552 ms in every session, by its own timer and not by an edge measured on the pin, and the 1.0 s cut-off itself has never been exercised.
The same story, step by step. Each step in the log was a test whose result changed what we did next.
Twelve steps, 2026-08-14 to 2026-10-12
The printer flagged walls too thin to print on one part, which was redrawn first. Of 59 pieces, 57 arrived; two were missing, and two broke as their supports were removed.
What it changed We built with what we had, taped the two breaks, and wrote down where they broke for the next revision.
With the LED at 5.553 mA, a paper scatterer at the sample position read F2 = 87 and F3 = 81 against dark reads of zero. With nothing in the beam, stray excitation stayed under one count.
What it changed The light path works, taped joints included. The UVette's orientation was fixed from here on.
An empty UVette under ordinary room lighting read F4 = 11 with the LED off. In a darkened room, or under an opaque box, every channel read zero.
What it changed A shield became mandatory, and “the dark read must be zero” became the rule that voids a cycle.
Running the project firmware, a culture reported to carry sfGFP read F4 = 19.58 ± 0.79 and its control read 0, over twelve cycles each. Twelve cycles are not twelve samples.
What it changed The emission channel is usable. We had called that culture “induced”; no AHL had been added, and the word was withdrawn.
A GFP culture read 40.50 ± 1.23 over four insertions, with its control at zero. Five days later the corresponding culture read 32.67 ± 1.83, lower by 19.3 %. The instrument had been moved in between and the state of the culture was not known.
What it changed A difference across days cannot be explained afterwards, so every comparison now carries its reference in the same session.
GFP stock, non-GFP stock and medium, mixed at set fractions. Samples without GFP read zero every time, and in the two sessions with enough signal the reading rose in proportion to the GFP culture. A thin stock in session 2 left three tubes at zero.
What it changed The stock has to be dense enough to clear the floor, and one session's slope cannot be carried to another as it stands. The Model page takes it from there.
The cuvette seat and the excitation end were redrawn: a pocket the cuvette drops into and lands in, an LED holder that grips the LED without a clip, and a baffle that prints on its side. With the cap that never arrived, that makes five parts to print. Two fit coupons were printed first, and the pocket and the socket that fitted by hand set the two dimensions a drawing cannot settle.
What it changed The fractures became a check: each printed layer's cross-section against the fullest section within 3 mm of it. Where the parts broke that ratio was 0.63 and 0.32; the five new parts keep at least 0.82, against a threshold of 0.75. It is a geometric screen tuned on those two breaks, not a strength calculation.
New printed parts were fitted, and one resistor was changed to take the LED from 5.55 mA to about 10 mA. With the same empty cuvette the rebuilt reader read F2 = 7.75 and F3 = 6.71 after an adjustment, inside the range version 1 had read, and about twice that once the current was raised. The emission channel stayed at zero throughout.
What it changed The reader is no longer the instrument that produced version 1's bench and culture readings. Its fit, its alignment and its response to a culture have to be measured again before the two can be put side by side.
One batch of GFP culture, undiluted and at five dilutions, with medium and a non-GFP culture as controls. The controls read zero on the emission channel. Looked at afterwards, the reading stayed close to one straight line from 25 to 200 µL of culture and fell short of it below that.
What it changed Version 2 has culture readings of its own. They are one session with one batch, so they stand beside version 1's and are not yet a comparison with them.
Suspensions scraped from plates at five AHL concentrations, with medium, a culture without the reporter and an sfGFP culture as controls. The reading rose with how much culture was in the cuvette, whether or not the wet lab had seen the plate glow, and sfGFP added to a dense suspension was still read.
What it changed Zero without GFP now carries a condition: it holds in liquid culture, not in a suspension this dense. Reading induction needs samples of matched density and one without AHL.
A culture without the reporter, diluted in LB and read from thin to dense. It read zero while the scatter channel stayed under about 33 counts, and no longer did from 46.
What it changed A few counts on the emission channel can be set against the scatter channel of the same read. An emission filter remains untested.
Five dilutions of the sfGFP test culture, read on our reader and on a SpectraMax M3. The two ranked the tubes alike, and over the three denser tubes our reading per unit of OD held within 10%.
What it changed The reader has now been set beside a plate reader once. Whether the two fluorescence readings are in proportion is still open: the plate reader needs every tube read undiluted.
Recorded project decisions from 2026-08-09. The marked ones removed capability from the prototype on purpose.
Every reading in this section was taken on this build, with its lid closed.
Measured on E. coli cultures with and without sfGFP, this build's green-to-blue ratio rose in a straight line with the amount of sfGFP (r² ≈ 0.97), with a 3σ detection limit of about 4% sfGFP culture; the tube with no sfGFP varied by 1.5% in that ratio over eight batches. Across its eight colour channels the added signal has the spectral shape of fluorescence, not of scattering.
A blue LED, a 90° sensor and a light trap, in a plywood box cut from one sheet.
A blue LED lights a UVette from the side, and an AS7341 spectral sensor looks at it from 90°. Everything sits in a laser-cut plywood enclosure in two layers: the optics above, blackened with matte black marker, and the ESP32 and its wiring below, with a deck between them. An OLED shows each result and one button starts a measurement.
Every laser-cut part, from the generator at 3.0 mm
The light trap is designed to swallow the excitation that passes straight through the sample so that it does not bounce back toward the sensor.
The sensor sits at 90° to the beam, so the excitation does not shine straight into it. Some blue still reaches it: with an empty chamber the blue channel F3 (480 nm) read about 312 counts, and a cuvette of plain medium raised it to about 730 (both in the fluorescence test). That the reading goes up when the sample goes in is consistent with the sensor collecting light scattered sideways out of the sample.
Cheap to make from one sheet of plywood, cut to whatever board you have, and steady in its readings.
Everything as costed below; the jointed parts come from one 450 × 300 mm sheet.
As costed below; the UVette was not pricedThe generator redraws every joint for the board's measured thickness, so any batch of nominal 3 mm plywood will do; only the jointed parts need to come from the same batch.
Checked in the generator, belowEight batches of the tube with no sfGFP varied by 1.5% in F5/F3.
Linearity testWet-lab members used a first version, and two changes came from what they found; every measurement on this page was taken on the version that followed.
They told us it was too small to lift the cuvette out. What followed: it was enlarged from 16 × 16 to 25 × 25 mm, with a matching lid.
From wet-lab useThey told us that with nothing to hold it, the cuvette had to be placed with care, watching where it sat and keeping clear of the chamber walls. What followed: a plain plate with a 13 × 13 mm opening now lies on top of the chambers and holds the cuvette in place.
From wet-lab use
Each reading subtracts LED-off from LED-on, and the result is the ratio F5/F3, which cancels changes in how much light reaches the sensor.
For each reading the firmware takes eight samples with the LED off, then eight with it on, and subtracts the averages channel by channel, which removes ambient light and the sensor's dark current; averaging eight samples also lowers the sensor's own noise by about √8, or 2.8 times. One press of the button records a batch of five readings, about 22 seconds each. The first is discarded and the other four are averaged.
The figure we report is F5/F3: the green channel at 555 nm divided by the blue channel at 480 nm. Dividing by the blue channel largely cancels changes in how much light reaches the sensor, whether from where the cuvette sits or from changes within a batch: over readings two to five, F3 and F5 each fall by about 2.7%, while F5/F3 changes by less than 0.1% on average. Net signal is the F5/F3 of a tube minus that of a tube with no sfGFP cells, measured in the same session.
F4, centred nearer the sfGFP peak, collects about 2.2 times as much sfGFP light as F5, but it also collects about three times the background from the blue LED, so its detection limit is slightly worse (about 4.4% against 3.6%). On this build F5/F3 is the better measure of sfGFP; F4/F3 supports the same conclusions with more noise.
sfGFP emits most strongly near 510 nm. F4 and F5 each respond over a band 39 nm wide at half maximum (AS7341 datasheet): F4 from about 496 to 535 nm, F5 from about 536 to 575 nm. The band centred closest to the sfGFP peak is F4. This build was set up from the start to read F5 (555 nm): the on-screen signal of the firmware used for these measurements is built on it. To check that choice we recomputed the main results with F4/F3 alongside, using the same data and the same method.
| Measure | F5/F3 | F4/F3 |
|---|---|---|
| Detection limit, share of sfGFP culture (3σ), linearity test | ≈ 0.036 | ≈ 0.044 |
| Linearity r², linearity test | 0.97 | 0.96 |
| Dose response r², fluorescence test | 0.90 | 0.82 |
| Background relative to F3, tube with no sfGFP cells | 0.055 | 0.160 |
| Shift when a blank is lifted out and put back | 2.9% | 0.7% |
F4/F3 supports the same conclusions, linear and with an intercept indistinguishable from zero. Where it does better is stability: it shifts less when the same tube is taken out and put back, and drifts less over an hour, because F4 and F3 are close in wavelength. For sensitivity, F5/F3 is the better measure on this build.
For its on-screen reading the firmware also computes
SIG5 = F5 − k5 × F3. Part of
what F5 sees is not sfGFP fluorescence, such as the blue LED's
long-wave tail scattered into the sensor; k5 is that part as a
fraction of F3, measured on a tube with no sfGFP cells. It is set by
sending the k command with such a tube in place, and
stored in the ESP32's flash.
k5 changes from one experiment to the next, so SIG5 is not compared across batches. Every result on this page uses F5/F3, which does not depend on k5.
The LED holds steady, and with the sensor aligned the reader detects sfGFP down to about 4% sfGFP culture; our AHL sensor strain's signal is too faint for it.
Aim Check that the reading follows what is in the cuvette.
Method Cultures of different density (OD600 0.15 to
1.53) were read one after another, and the blue channel F3 of the 28
batches with an OD600 record was compared with their
OD600.
Result Culture density explained almost none of the variation in the blue channel F3 (R² = 0.01): the reading was not following what was in the cuvette, consistent with a sensor mounted too low to see the sample well. The reading also fell by about 40% in a single step between the seventh and eighth batches, about 20 minutes in, and stayed low. Allowing for culture density, a single step fits the data much better than a gradual drift (R² 0.88 against 0.67). The cause of the step was not identified, but a sudden change like this is more consistent with the sensor shifting than with the LED fading. The test gave no usable culture data. The next test checked whether the LED was at fault; after that, the sensor was aligned with the cuvette's measuring window and fixed in place for all later tests.
Aim Find out whether the LED itself fades, which could have
caused the fall seen in the density test.
Method With the sensor still in its original position and the
chamber empty, the firmware repeated batches automatically for an
hour: five readings taking about two minutes, then a one-minute pause
with the LED off. Twenty-one batches in all.
Result and conclusion Over an hour the reading stayed within a 1% range. The LED and electronics are stable, so the fall in the density test was not the LED fading. The sensor was then aligned with the cuvette's measuring window and fixed in place for every test that followed.
Aim Show that the signal comes from sfGFP fluorescence and
scales with the amount of sfGFP.
Method Twelve tubes of 200 µL in a grid: culture
density d of 0.25, 0.5, 0.75 and 1 (the fraction of each tube
that is culture, the rest medium) crossed with a fraction f of
0, 0.5 and 1 of cells carrying sfGFP. The two stock cultures had nearly
the same density (OD600 0.84 with sfGFP, 0.85 without), so
the share of culture taken from the sfGFP stock is close to the share
of sfGFP cells. Each tube was read in at least two batches.
Result Net F5/F3 rose in proportion to the amount of sfGFP, d × f (r² ≈ 0.90, eight tubes), and the line passes through zero (intercept −0.5 ± 2.2, in net F5/F3 × 1000).
Conclusion The signal behaves as sfGFP fluorescence: it is proportional to the amount of sfGFP and has the spectral shape of its emission.
A signal that rises with the amount of sfGFP does not on its own prove fluorescence. Each net value already compares tubes of the same density, so turbidity is subtracted, but the two strains could still scatter light differently. The spectral shape can tell scattering from fluorescence, and it is shown under What ten channels add: the difference between tubes with and without sfGFP has the shape expected of sfGFP fluorescence.
Aim Keep the total culture density nearly constant, vary only
the share of sfGFP culture, and measure the detection limit.
Method Tubes of 200 µL mixed to shares f of
0, 0.10, 0.15, 0.20, 0.30, 0.50, 0.75 and 1, where f is the share
of the volume taken from the sfGFP culture, which is proportional to the
number of sfGFP cells. The OD600 of every tube in the fit was
measured, from 1.30 (no sfGFP) to 1.45 (all sfGFP). Each tube was read
in at least two batches. The tube with no sfGFP was placed four times
during the session and read in two consecutive batches each time, eight
batches in all.
Result Signal and the share of sfGFP culture are linear. Without forcing the line through the origin, the intercept is 1.6 ± 2.6 (standard error, in net F5/F3 × 1000), not different from zero, with r² ≈ 0.97; a line through the origin has slope 68.7 and r² ≈ 0.97. Across its eight batches, the tube with no sfGFP varied by 1.5% in F5/F3 (standard deviation 0.00083 divided by a mean of 0.055). The detection limit, three times that standard deviation divided by the slope in the same units (0.0687 per unit f), is 0.036: about 4% sfGFP culture (from F5/F3).
At the lowest shares the points scatter most relative to their size: the two lowest tubes in the fit are off by −23% (f = 0.10) and +48% (f = 0.15), while from f = 0.20 up every tube is within about ±20%. At f = 0.05 a 200 µL tube holds the equivalent of only 10 µL of sfGFP culture, so an error of 2 µL is already 20% of f; the low tubes are the most sensitive to how they were mixed. The reading itself is not the cause: repeated batches of the same tube agree, the zero is stable, and F4, F5 and F6 are off in the same direction and by similar amounts in each tube (correlation 0.95–0.98), so the deviation lies in how much sfGFP the tube holds, not in one channel. Three further low-share tubes (two at f = 0.05, one at f = 0.10) scatter the same way, from −54% to +276%, and are shown but left out of the fit. Including them changes little: the slope becomes 68.8 and r² 0.95, and the detection limit, which comes from the blank tube, does not change. Mixing low shares by serial dilution should reduce this scatter.
Aim Read the AHL-sensing strain our wet lab
built, E. coli TOP10 carrying ProD-LasR ·
pLasRV-sfGFP, across a range of AHL.
Method The strain was given AHL at six
concentrations, 10 nM to 1 mM, with two or three readings at
each, 16 in all, and three tubes of medium alone as blanks. It was read
with the current firmware rather than the batch firmware of the tests
above.
Result F5/F3 stayed between 0.048 and 0.060 (mean 0.055) across the whole range, at the level of the culture without sfGFP in the linearity test (0.055) and far below the undiluted sfGFP test culture (0.127); medium alone read 0.026 to 0.033. The differences between concentrations were no larger than those between tubes at one concentration. F4 rose and fell with the red channel F7 and the clear channel (r = 0.99 and 0.92): the readings followed the light scattered by the cells.
Conclusion This reader detects and quantifies a strongly expressing sfGFP culture down to about 4%, but our sensor strain's F5/F3 could not be told apart from a culture without sfGFP read in another session. No culture without sfGFP was read alongside it, so the result cannot be stated against the 4% detection limit. In the wet lab's plate reader the strain did glow above its control (Engineering), so the signal exists but is too faint for this reader. The 3D-printed build saw the same pattern with dense suspensions: readings followed the amount of culture, not the AHL (Where it stops).
The extra channels confirm that the signal is fluorescence, and that the ratio cancels where the cuvette sits.
Every reading records all ten channels, while the results above use only F5 and F3. Recomputing the later tests with the other channels checks those results from a different angle, without another experiment. The conclusions hold up; these are the two checks that matter most.
For each culture density in the fluorescence test we took the tube with all cells carrying sfGFP and the tube with none, divided each by its own F3 to remove brightness differences, and subtracted. What is left is the spectrum that sfGFP adds. Because every channel is divided by F3, the F3 point is zero by construction; each curve is then scaled to its own F4 value so that the four densities can be compared by shape.
The four curves almost coincide: a single peak at F4 (515 nm), falling steadily toward the red, and near zero at 415 nm. That is the shape expected of sfGFP emission. Scattering would be strongest at short wavelengths, and a change in turbidity would move every channel the same way; neither would produce this curve.
If only the amount of light reaching the sensor changes, all channels change in proportion and the ratio is unaffected. The figure shows a real case from the fluorescence test: a blank was read, taken out while another tube was read, then put back and read again.
When the blank went back in, the nine channels rose together by 14 to 20%, at most 6% apart, while F5/F3 moved by 2.9%: the jump came from where the cuvette sat, and the ratio largely cancelled it.
Generate the parts below for your board's thickness and cut the jointed parts from one 450 × 300 mm sheet of the same board. Colour the parts that face the optical layer with a matte black marker; fit the chamber walls into the deck by their tabs, set the deck on its rails inside the enclosure, and fit the roofs onto the walls.
The AS7341 sensor and the OLED share the ESP32's I²C bus; the LED runs from GPIO25 through 220 Ω, and the button sits on GPIO13. Everything runs from 3.3 V, and the ESP32 takes its power over USB from any supply.
Seat the UVette with its 2 mm path toward the LED, close the lid, and press the button: five readings of about 22 seconds each, the first discarded and the other four averaged. In each session, also read a tube with no sfGFP; results are its F5/F3 subtracted from each sample's.
The whole set of parts is generated in the browser from the measured thickness of the plywood. The optical layout stays fixed: the cuvette pocket, the heights and positions of the light holes, and the inside of every chamber do not change with thickness, and thicker board makes the walls grow outward. Tabs are 8 mm wide; slots and notches are made 0.1 mm smaller on each side to take up the laser's kerf. The jointed parts all fit one 450 × 300 mm sheet and must be cut from one batch of board, because every joint is sized from a single measured thickness. The small parts without joints fill the gaps around them; those that do not fit go on a second sheet of any board.
This is the order our reader was put together in. Open the chambers in the 3D view to see each part named here.
Assembly tip: the deck and the enclosure panels have symmetric outlines, so each also fits when turned around, but their holes are not symmetric and the parts carry no orientation marks. Before gluing, check that every hole lines up with what it serves: the LED, the cuvette, the sensor and the wire holes.
These are the steps the first four tests under Validation were taken with, using the batch firmware described under Firmware and data.
Everything runs from the ESP32's 3.3 V supply. The AS7341 and the OLED share one I²C bus at different addresses (0x39 and 0x3C); both boards carry their own pull-up resistors. The LED is driven directly: GPIO25 to 220 Ω to the LED to ground, switched by the firmware. The button connects GPIO13 to ground and uses the ESP32's internal pull-up. None of the pins used (13, 21, 22, 25) is a strapping pin that affects boot. The ESP32 itself is powered over USB, through the opening in one short side of the enclosure: we use a power bank, or the computer when the batch firmware sends its readings, but any USB supply will do.
| Signal | ESP32 pin | Connects to |
|---|---|---|
| Power | 3V3 | AS7341 VCC, OLED VCC |
| Ground | GND | AS7341 GND, OLED GND, LED cathode, the button's other terminal |
| I²C SDA | GPIO21 | AS7341 SDA, OLED SDA |
| I²C SCL | GPIO22 | AS7341 SCL, OLED SCL |
| LED | GPIO25 | 220 Ω to the LED anode; cathode to GND. HIGH turns it on |
| Button | GPIO13 | One terminal of the button; the other terminal to GND. Internal pull-up, reads LOW when pressed |
The readings on this page were taken with the reader's batch firmware (v5.5), which prints every reading over the serial port. The current firmware builds on it and adds the link to our LasReader software: a measurement still takes five readings and keeps the last four, and the data processing is done in the software rather than on the reader. It is published in the team's GitLab repository; the software is described on the Software page.
The costing below is for the laser-cut build, including its enclosure, display and button. All figures are in US dollars.
| Item | Qty | Cost (US$) |
|---|---|---|
| Electronics | ||
| ESP32 development board | 1 | $6.79 |
| AS7341 spectral sensor board | 1 | $27.15 |
| Push button | 1 | $0.51 |
| OLED display | 1 | $4.73 |
| Breadboard | 1 | $1.42 |
| Jumper wires | several | $1.58 |
| LED and 220 Ω resistor | 1 each | $0.16 |
| Electronics subtotal | $42.34 | |
| Enclosure, laser-cut | ||
| Wood board | 1 | $1.89 |
| Laser cutting | 1 | $9.47 |
| Enclosure subtotal | $11.36 | |
| Total, as costed | $53.70 | |
Not in the costing: the Eppendorf UVette, which the sheet never priced, and the USB cable and power supply, which any USB supply can be; we have not guessed at them. The few small parts that do not fit on the main sheet were cut from leftover board, so they add nothing to the cost.
We built it for our biosensor, rehearsed on GFP cultures, and ended with a platform that other teams can tune to their own samples.
We expected an AHL-induced culture to be about as bright as a GFP culture, so the first stage ran on a GFP test strain as a rehearsal. What we had read beforehand put the two in the same range, as one LuxR circuit is8. What we had not noticed is that circuits differ: a Las device, the kind our biosensor is, stayed well below a constitutive reference9.
Along the way we found that how dense a sample is changes the reading, and worked on describing it with a model and with the reader’s other channels (Model 3.5).
The samples our wet lab then gave us were denser still, and AHL-dependent induction had not been shown conclusively (Engineering): not what we had prepared for, and a different thing to measure. A plate reader has its place, and we read ours beside one. What the two builds offer is a steady platform to read from. Both record ten channels on every read, and both run from the same software. Under its shield the 3D-printed build reads zero in the dark, so its counts are used as they come; the laser-cut build is cheaper and easier to assemble, and removes its background in the data. A team can choose which bands to read, and how to process them, to suit its own samples.
Every measured number on this page comes from our own dated engineering records.