Defluorination Measurement

Alt text for the hero illustration

Overview

Our work is summarised under the following questions.

Other iGEM teams can replicate our measurement

The assay needs only two affordable chemicals that are not standard lab stock (alizarin-3-methyliminodiacetic acid and lanthanum(III) nitrate hexahydrate), plus acetate buffer, acetone and sodium fluoride. Readout is absorbance from 400 to 700 nm, or the naked eye above roughly 20 µM fluoride, so no fluoride electrode, ion chromatography or 19F-NMR is required.

Everything that determines the outcome is fixed and stated in a defined protocol: 100 µM 1:1 Ali:Lan, 84 mM acetate buffer pH 5.2, 25 v/v% acetone added after incubation, 1 hour incubation in the dark, and per-well volumes in a 96-well plate. The data analysis is written as a procedure rather than a result, so another team can reproduce the method even if their numbers differ.

One step must be repeated locally rather than copied: the isosbestic point and the optimal numerator are properties of the medium, so every team recalibrates in their own medium with their own fluoride ladder. The scan ranges, acceptance thresholds (R2≥0.95R^2 \ge 0.95, ≥40\ge 40 nm separation, ≥20%\ge 20\% signal change, positive slope) and selection rules (lowest The LOD is the lowest signal, or the lowest corresponding quantity to be determined (or extracted) from the signal, that can be observed with a sufficient degree of confidence or statistical significance. among accepted pairs) are given, so this is a defined step described on this page.

Our protocol is described

Each of the three development stages contains its own Materials & Equipment, step-by-step Protocol, Data analysis and Controls section. Thus any reader can run any stage in isolation. Stock preparation is included (1 M NaF weighed on an analytical balance, re-diluted to 1 mM in the medium being calibrated), as is a master mix table in µL/well for 100 µM, the aluminium foil light protection during incubation, and the exact point at which acetone is added.

A video of the assay steps (already optimised) is included alongside the written protocol.

The data analysis is specified down to equations. The ratio formation, isosbestic point search, numerator selection, regression, LOD =3.3σ/b= 3.3\sigma/b and The LoQ is defined as the lowest concentration of a substance that a measurement method can reliably put a number on. It is the point at which the measurement meets predefined goals for both bias and imprecision, ensuring that the analyte can be quantified with acceptable accuracy and precision =10σ/b= 10\sigma/b with the rule for which σ\sigma to use, the linear range, and interference interference analysis are described.

Any projects attempting to quantify fluoride can benefit from our measurement

The assay reports fluoride release and quantifies it, so it is useful to any project that cleaves a C-F bond, not only to our TFA work. Because biological systems rarely produce free fluoride, the signal is specific for defluorination.

We characterised the method as a tool rather than as a single experiment: four fluorochemicals (MFA, DFA, TFA, PFOA), two bacterial hosts (E. coli MG1655 and P. putida KT2440), and five media (demi water, M9, M9*(HEPES buffered), 20 mM HEPES, tap water) with a comparison card so a team can pick a medium for their own research question.

The limitations are documented as carefully as the working range, which is what makes the method adoptable for other researchers: phosphate interferes, the response saturates above 40 µM, and 5 mM fluorochemicals supressing the signal up to 60%, which is handled by dilutions. A team can therefore determine in advance whether the assay will fit their goals better, instead of finding out later.

We validated the measurement and calibrated units using defined controls.

Controls were defined per experiment and are listed in each Controls & Purpose section.

Measurement validation: a 0 µM fluoride blank defines the zero point. With 10 blank wells, we were able to get the σ\sigma behind LOD and LOQ. A 40 µM spike is the positive control for the spectral shift, and the literature condition (50 µM 1:1 reagents, 620/530 nm) was reproduced before we changed anything, so our optimisation is compared to a published method.

Matrix controls: M9 served as a negative medium control showing what phosphate interference looks like. 93 hours incubated E. coli supernatant spiked with fluoride tested the effect of incubation, while media spiked with fluorochemical quantified PFAS interference.

Biological controls: wild-type E. coli MG1655 and empty-vector P. putida KT2440 as negative strains for (spontaneous) defluorination. MFA as a substrate that fluoroacetate dehalogenases should defluorinate, and PFOA as a substrate they should not.

Units: concentrations are accurately prepared NaF standards and reported in µM fluoride. Each medium has its own calibration line in those units. Concentrations of defluorination were calculated and could be re-zeroed based on interferences. Concentrations outside of the linear range were diluted to accurately quantify defluorination.

Abstract

Introduction

After identifying or mutating and successfully producing potentially TFA degrading enzymes, our main objective became to test whether they were functionally active. Production of fluoride indicated the cleavage of the C-F bond by our enzymes. The fact that TFA, and other perfluorinated chemicals, are very difficult to defluorinate, increases the pressure for new enzymes to be found. Many starting options and introduced mutations require high-throughput screening. Current methods are prohibitive/non biocompatible (with PFAS) and not generally accessible. Improving upon a novel qualitative defluorination assay [1] currently suffering from reproducibility and characterisation issues, we aimed to present a well defined measurement technique with the following values in mind: sensitive quantification, high throughput, biocompatibility and accessibility.

Instead of an isolated application in mind, we identified key parameters, controls, various useful media types, limitations, rational data analysis, bacterial hosts and packaged it in a complete protocol for accelerated screening and quantification of defluorinating enzymes. Our development story for the measurement is structured in the following steps:

1. Validation of working principles and protocol optimisation

2. Adjustments towards biocompatibility and interferences

3. Real enzyme tests and bigger picture

Defluorination assay

Validation of working principles and protocol optimisation

Context

Our starting point for measuring fluoride was a known method first described by Belcher et al., In this approach, complexs are formed between an alizarin dye and lanthanum resulting in an absorbance change upon fluoride addition [2] (Figure 1). Bygd et al. (2022) [1] made the first efforts to utilise this chemistry for determining degradation of C-F groups by measuring fluoride released. In our project, we also implemented this approach and based the assay on the following mechanism: when TFA (PFAS) is broken down, fluoride is produced, changing the absorbance of the solution containing the complex. The common lack of fluoride production in biological systems allows for a signal purely limited to the breakdown of a C-F bond. However, team Triton and one other anonymous researcher we spoke to found that the existing methods is mainly qualitative. In other words, currently lacks optimisation for sensitivity or a clear protocol and therefore is not reproducible, high throughput, or accessible for the quantification of defluorination. We aimed to amend this.

The original paper proposed using 25 v/v% acetone for enhanced sensitivity. Bygd et al., (2022) [1] suggested using a 1:1 ratio of the dye and lanthanum at 50 µM reagent concentrations and pH 5.2 acetate buffer. We first assessed these starting concentrations and ratios as well as optimised the data analysis for quantitative measurement. It was also reported that the formation of a spectral change can take time for low fluoride concentrations, and thus we also surveyed this.


Proposed complex formation by Bygd et al., (2022) [1]
Figure 1. Proposed complex formation by Bygd et al., (2022) [1]

Aim

With the context and reasoning in place, the goal was to validate the complex formation, the effect of fluoride addition and to establish a protocol (data analysis, optimising reagent concentrations and incubation time). Our vision is to make the method as sensitive as possible and usable with common measurement devices, such as UV-VIS, for quantification and screening by eye in the visible spectrum.

Materials & Equipment
Reagents
  • 5 mM Alizarin-3-methyliminodiacetic acid (Ali);

  • 5 mM Lanthanum(III) nitrate hexahydrate (Lan);

  • Demineralised water;

  • 1.68 M acetate buffer (pH 5.2);

  • Acetone (>99.9%);

  • 1 mM and 1M sodium fluoride (NaF) in MilliQ.

Consumables
  • 96 microwellplate;

  • P200, P1000 pipette tips;

  • 1.5 mL microcentrifuge tubes;

  • 40 mL Falcon tube;

  • Multichannel pipette reservoirs;

  • Aluminium foil.

Equipment
  • 500 µL (quartz) cuvettes;

  • Micropipette (P200, P1000);

  • Multichannel pipette (P200);

  • Plate reader (400-700 nm spectral analysis);

  • pH meter;

  • Analytical balance.

WageningenUR: Quantitative fluoride assay (2026)
Protocol
Fluoride complex colourimetric change
  1. Fill 2 cuvettes with 250 µL demi water each.

  2. Add the alizarin dye till 50 µM to both.

  3. Add 25 µL acetate buffer.

  4. Add the lanthanum till 50 µM .

  5. Add 125 µL acetone.

  6. To 1 of the cuvettes, add 4.55 µL 1M NaF.

  7. Observe and compare the colour of both cuvettes.

Reagent and incubation screening
  1. Make a 1 mM sodium fluoride stock by weighing sodium fluoride on an analytical balance. First, prepare a 1 M stock to be more accurate, after which you re-dilute.

  2. Prepare fluoride standards of 0, 1, 2, 4, 5, 10, 20, 40 µM from the 1 mM NaF stock

  3. Using a multichannel pipette, add 100 µL of the standards in a 96 well microplate in pre-determined locations.

  4. Mixed reagents need to be prepared fresh before each measurement and kept out of the light for storage. Prepare pre-mixed solutions for 12.5, 25, 50, 100, 150, and 200 µM Ali:Lan reagents for 300 µL of total volume per well. Below, an example is shown for 100 µM. The volumes are multiplied for the amount of wells in the measurement (12):

    Table 1. Composition of the master mix per well. Acetone is added to the well after the incubation time has passed and is therefore not pre-mixed. Reagent concentrations are found in the Reagents section.
    Chemical Add for 100 µM Ali:Lan (µL/well)
    Alizarin dye 6
    Lanthanum 6
    Acetate buffer 15
    Acetone 75
    MilliQ 98
    Total 200
  5. Pipette 125 µL/well of the pre-mixes into the pre-defined wells.

  6. Incubate for either 1 hour (odd columns in our case) or 2 hours (even columns). After the incubation time, add 75 µL acetone and measure those wells in the plate reader between 400-700 nm.

  7. Perform data analysis on the absorbance data.

Water calibration
  1. From the 1M standard stock, make a 1 mM stock in the calibration media used.

  2. Make analytical standards in the media wanted for 0 (blank/negative control), 1, 2, 5, 10, 20, 30, 40, 60, 80, 100 µM. 10 blanks were used and 5 technical replicates per concentration.

  3. Add 100 µL of the standards in a 96 well microplate in pre-determined locations.

  4. Make a fresh mix of reagents before each measurement and store the chemical stocks in the dark. 100 µM of Ali:Lan is used. All wells will consist of the following compositions, although the acetone is only added after incubation. Therefore excluded it in the pre-made master mix:

    Table 2. Composition of the master mix per well. Acetone is added to the well after the incubation time has passed and is therefore not pre-mixed. Reagent concentrations are found in the Reagents section.
    Chemical Add for 100 µM Ali:Lan (µL/well)
    Alizarin dye 6
    Lanthanum 6
    Acetate buffer 15
    Acetone 75
    MilliQ 98
    Total 200
  5. Add 125 µL of master mix in each well using a multichannel pipette.

  6. Incubate for 1 hour. During the incubation, cover the (black) well plate in aluminium foil to limit light exposure.

  7. Directly after 1 hour, add 75 µL acetone and measure the plate for absorbance in a microplate reader between 400-700 nm.

  8. Perform data analysis on the absorbance data.

Data analysis

Each well is reduced to a single absorbance ratio; that ratio is converted to fluoride through a media dependent calibration line, and the line is re-zeroed on the plate’s negative control before quantification. Error bars are the standard deviation of the replicate wells.

Building the dataframe
  1. Read every plate into a table with wavelength (nm) as the index and one column per well (A1…H12\mathrm{A1}\ldots\mathrm{H12}), taking the well names from row 1 of the sheet (can change depending on the plate reader type).

  2. Clean the values: convert comma decimals to points, blank cells to missing, and keep only rows whose wavelengths falls inside the measured wavelength window (400-700 nm).

Dividing wavelengths
  1. For a chosen wavelength, take the absorbance at the nearest measured point.

  2. Divide a numerator wavelength by a denominator wavelength, per well:

    R=AnumAdenR = \frac{A_{\mathrm{num}}}{A_{\mathrm{den}}}
  3. Skip wells whose denominator is missing or effectively zero, to avoid dividing by noise.

  4. Use RR, never raw absorbance, as the signal in all regressions. The division cancels out concentration and pathlength/volume differences between wells.

Finding the isosbestic point
  1. Average the replicate wells at each fluoride concentration to get one mean spectrum per concentration, A‾c(λ)\bar{A}_c(\lambda).

  2. At every wavelength, measure how far the spectra move apart from the blank:

    S(λ)=mean over concentration of |A‾c(λ)−A‾0(λ)|S(\lambda) = \text{mean over concentration of } \bigl|\bar{A}_c(\lambda) - \bar{A}_0(\lambda)\bigr|
  3. The isosbestic point is where that spread is smallest (lines "cross"), searched between 480 and 640 nm (This requires at least three different fluoride concentrations):

    λiso=argminλS(λ)\lambda_{\mathrm{iso}} = \arg\min_{\lambda} S(\lambda)
  4. Determine λiso\lambda_{\mathrm{iso}} separately for every medium; it is a property of the media, not of the assay.

Choosing the best numerator
  1. Pin the denominator to that medium’s own isosbestic point. There the absorbance does not change with fluoride, so the ratio moves only because the numerator moves, which makes it a good internal normalisation.

  2. Scan candidate numerators on a 5 nm grid between 620 and 680 nm, where the fluoride complex absorbs.

  3. Discard a pair unless it satisfies all of:

    • the two wavelengths are at least 40 nm apart;

    • the denominator carries real dye absorbance, not a near-zero baseline;

    • the ratio moves by at least 20 % across the ladder;

    • the slope is positive.

  4. Fit every pair matching the mentioned criteria over the calibration range and keep its slope, R2R^2 and LOD.

  5. Accept pairs with R2≥0.95R^2 \ge 0.95, and among those select the one with the lowest LOD. Lock that pair for the medium.

Linear regression
  1. Fit ŷi=bxi+a\hat{y}_i = bx_i + a by least squares (xx = nominal fluoride concentration in µM, yy = locked ratio; minimum four points over three concentrations), then:

    sy=SSresn−2andR2=1−SSresSStots_{y} = \sqrt{\frac{SS_{\mathrm{res}}}{n-2}}\qquad\text{and}\qquad R^{2} = 1 - \frac{SS_{\mathrm{res}}}{SS_{\mathrm{tot}}}

    with SSres=∑(yi−ŷi)2SS_{\mathrm{res}} = \sum (y_i - \hat{y}_i)^2 (scatter about the line) and SStot=∑(yi−y‾)2SS_{\mathrm{tot}} = \sum (y_i - \bar{y})^2 (total scatter about the mean). R2R^2 is only descriptive per situation, since it rises with the width of the concentration range.

LOD and LOQ
  1. With bb the slope of the calibration line and σ\sigma the standard deviation:

    LOD=3.3σbandLOQ=10σb\mathrm{LOD} = \frac{3.3\,\sigma}{b} \qquad\text{and} \qquad \mathrm{LOQ} = \frac{10\,\sigma}{b}
  2. When there are replicates for the blank, take σ\sigma as the standard deviation of the 0μ0~\muM wells. This needs more than two blank wells (>2).

  3. When there are no replicates (one well per condition, as in the reagent/time optimisation), take σ\sigma as the residual standard deviation sys_y of the fit. The limit is then a property of the curve rather than of repeated measurement.

  4. Carry forward the LOD and LOQ of the reference (fresh HEPES) curve over its validated range, and draw them uncorrected as flat horizontal lines in every figure.

Linear ceiling
  1. Sort the different concentrations form low to high, and fit only the lowest three concentrations for the linearity, then the lowest four, and so on up to all the concentrations.

  2. Record the highest concentration at which the calculated fit still holds R2≥0.98R^2 \ge 0.98. That value is the linear ceiling.

  3. Report it per condition as a descriptive ceiling. It shows where the response starts to saturate, which is also where forcing a straight line would inflate the error and the LOD with it.

Controls & Purpose

Negative control (0 µM fluoride)

Defines the zero point. Can be used to calibrate the spectra and defines the LOD and LOQ.

Positive control (40 µM fluoride spike)

Confirms that the complex responds to fluoride with an expected colour and spectral shift. High enough for easy visual determination.

50 µM Ali:Lan complex at varying fluoride concentrations

Confirm the original literature for the fluor determination (at 620/530 nm ratio).

Results

Our team investigated the change in the complex colour in Figure 2. Both cuvettes contain 50 µM alizarin (Ali) and lanthanum (Lan), 25 v/v% acetone, 84mM acetate buffer pH 5.2 in 500 µL milliQ. The cuvette on the left was spiked with sodium fluoride (NaF) and an intense colour change was observed.

Cuvettes with 50 µM 1:1 complex reagents with (left), and without (right) fluoride added
Figure 2. Cuvettes with 50 µM 1:1 complex reagents with (left), and without (right) fluoride added

Upon the addition of fluoride, the xpected colour change was observed, going from pink to purple and blue with increasing concentrations, respectively. After seeing the colour change, confirming qualitative determination of fluoride, our focus shifted towards making the measurement sensitive and quantitative. Attempting to extend the assay for fluoride concentrations as low as possible, we did this by alternating the reagent concentrations. Figure 3 shows the colouring of the complex at different reagent concentrations between 12.5 and 200 µM Ali/Lan. The higher the concentration of the reagents used, the darker the colour in the wells.

Different reagent concentrations screened for optimal conditions. Concentrations from 0 to 40 µM are screened. Incubation times and reagent concentrations were varied per column: odd columns were incubated for 1 hour and even columns for 2 hours before acetone addition and UV-VIS measuring. The Ali:Lan concentrations increase from left to right.
Figure 3. Different reagent concentrations screened for optimal conditions. Concentrations from 0 to 40 µM are screened. Incubation times and reagent concentrations were varied per column: odd columns were incubated for 1 hour and even columns for 2 hours before acetone addition and UV-VIS measuring. The Ali:Lan concentrations increase from left to right.

Spectral analysis using UV-VIS, showed increasing peak heights with increasing concentrations. For example, the spectra of 0-40 µM fluoride in 100 µM reagents after 1 hour incubation as seen in Figure 4. The proposed ratio wavelengths from literature are shown with the red ( 530nm denominator) and blue (620 nm numerator) dotted lines. After screening and dividing all wells by the previously stated wavelengths, linear regression was performed on the concentrations against the ratio signal. Linearity and the The LOD is the lowest signal, or the lowest corresponding quantity to be determined (or extracted) from the signal, that can be observed with a sufficient degree of confidence or statistical significance. were scored and shown in Figure 5. Good linearity (>0.99) and low limits of detection are found for reagent concentrations above 100 µM.

UV-VIS spectra of 100 µM complex reagents at 1 hour incubation (A) and with the blank (0 µM) subtracted. The red dotted line (530 nm) shows the denominator, while the numerator is indicated by the blue dotted line (620 nm).
Figure 4. UV-VIS spectra of 100 µM complex reagents at 1 hour incubation (A) and with the blank (0 µM) subtracted. The red dotted line (530 nm) shows the denominator, while the numerator is indicated by the blue dotted line (620 nm).
Different Ali:Lan reagent concentrations ranked at 1 and 2 hours of incubation. The linearity (A), limit of detection (LOD) (B), and linear ceiling (C) are shown. 100 µM (black box) was selected for further use.
Figure 5. Different Ali:Lan reagent concentrations ranked at 1 and 2 hours of incubation. The linearity (A), limit of detection (LOD) (B), and linear ceiling (C) are shown. 100 µM (black box) was selected for further use.

After initial validation and testing, we decided to continue with 100 µM complex reagents (see discussion). Further investigation and optimisation of 100 µM complex reagents with 1 hour incubation was focused on determining the best ratio wavelengths for linearity. Analysis between 0-100 µM NaF using a microplate reader resulted in the following absorbance spectra (mean ± SD) when the blank (0 µM) is subtracted (Figure 6 (A)). Low concentrations show a smooth, singular peak around 580 nm, fluoride causes the formation of a secondary peak around 620 nm.

Instead of assuming 620/530 nm is the optimal ratio for sensitivity at low fluoride concentrations, the isosbestic point (545nm, red dotted line) (point where the lines of the graphs cross) and the best numerator (630nm, blue dotted line) for linearity between the low range of 0-40 µM were calculated. The denominator is set at the isosbestic point to cancel out the effects of changes in reagent concentrations, since it is dependent on the reagent concentrations.

Spectral analysis of demi water at the optimal ratio with the blank subtracted (normalised) (A). The spiked standards are plotted against the divided wavelengths in (B).
Figure 6. Spectral analysis of demi water at the optimal ratio with the blank subtracted (normalised) (A). The spiked standards are plotted against the divided wavelengths in (B).

The optimal wavelengths divided over each other were plotted against the standards (Figure 6(B)). Widening of the data points and saturation of the signal was observed above 40 µM at the selected ratio. The limit of detection (LOD) and quantification (LOQ) are at the selected range up to 40 µM 3.8 and 11.6 µM, respectively.

Discussion & Troubleshooting

The colour change as described in literature upon fluoride addition was successfully reproduced. In Figure 3, the negative (0 µM) and positive (40 µM) controls show no and very intense colour change, respectively, indicating effects on the formed complex by fluoride too. However, the reported reagent concentrations were proven to be suboptimal for a lower fluoride concentration calibration range. We made the decision to use the 100 µM of 1:1 complex reagent because of the R2R^2 above 0.99, the best visual screening between the high concentrations while being more sustainable with the amount of chemicals used. 1 hour of incubation time before UV-VIS measuring was selected due to the minimal difference in performance and staying time efficient.

To determine the variance per concentration and upper bounds of the measurement, the calibration was repeated with technical replicates and extended range up to 100 µM (Figure 6). Saturation was observed above 40 µM, confirming a usable range of 0-40 µM in demi water. The same data was used to optimise the ratio, again using the isosbestic point to normalise the path length and reagent concentrations. Numerator optimisation was allowed between 610-700 nm, since 550-610 nm showed larger replicate standard deviations with the ratio 630/545 nm the 0-40 µM range gave R2R^2 of 0.996, LOD of 3.8 µM and LOQ of 11.6 µM.

Moving forward, after proving that the mechanism worked and sensitive measurements can be done by eye and optimised for plate readers, we could expand towards biocompatible media for enzyme screening. Since optimal ratios can shift between sources, identical optimisation iterations for different types of media will be described next.

Adjustments towards biocompatibility and interferences

Context

After we had proved that the mechanism works in water, similar approaches can be repeated in different media. The optimal medium allows enzymes to function while not interfering with the assay. This would improve the ease of use and real time monitoring. For this measurement, we tested 5 media types: demi water, M9, HEPES, M9*(HEPES), and tap water.

Defined M9 would allow for direct analysis in a growth media. However Bygd et al., (2022), determined that the phosphate will likely interfere with the complex colours (other growth media, MOPS and LB too).

Because the medium components interfere with the reaction, it is not possible to use the same medium in which the cells are grown to assess their ability to degrade TFA. Therefore, a two-step assay is required. While HEPES was identified as a functioning media for this (Bygd et al., (2022) [1] incubated cells in HEPES), cells can not grow in it. To further improve the assay, we modified the medium composition to achieve a compromise between supporting cell growth and providing suitable conditions for the analytical reaction. Removing the salts from the M9 formulation and replacing it for HEPES, while supplementing a nitrogen/phosphate source, resulted in novel M9*M9^* (HEPES) growth medium. Calibrating in tap water was done to work towards a proof of concept for water treatment in the future.

Interferences by PFAS and the effect of E. coli incubation in HEPES, were also unknown and tested as controls for upcoming testing and quantification of bacterial hosts expressing enzymes.

Aim

Determine which media types are suitable for screening and/or quantification of defluorination enzymes based on linearity, LOD and The LoQ is defined as the lowest concentration of a substance that a measurement method can reliably put a number on. It is the point at which the measurement meets predefined goals for both bias and imprecision, ensuring that the analyte can be quantified with acceptable accuracy and precision.. Finding out if TFA and other fluorinated compounds, as well as metabolic waste products from the biomass, interfere with the measurement.

Materials & Equipment
Reagents
  • 5 mM Alizarin-3-methyliminodiacetic acid (Ali);

  • 5 mM Lanthanum(III) nitrate hexahydrate (Lan);

  • Demineralised water;

  • 1.68 M acetate buffer (pH 5.2);

  • Acetone (>99.9%);

  • Trace elements (100x);

  • 2 M glycerol;

  • 600 mM HEPES;

  • Sodium phosphate (monobasic);

  • Ammonium chloride;

  • Sodium chloride;

  • 1 mM sodium fluoride (NaF) in M9, M9*, 20 mM HEPES, and tap water;

  • 5 mM monofluoroacetate (MFA), difluoroacetate (DFA), trifluoroacetate (TFA), and PFOA in 20 mM HEPES;

  • E. coli MG1655.

Consumables
  • 96 microwellplate;

  • P200, P1000 pipette tips;

  • 1.5 mL microcentrifuge tubes;

  • 40 mL Falcon tube;

  • Multichannel pipette reservoirs;

  • Aluminium foil;

  • 50 mL syringe;

  • Sterile 0.2 µm filter.

Equipment
  • Micropipette (P200, P1000);

  • Multichannel pipette (P200);

  • Plate reader (400-700 nm spectral analysis);

  • pH meter;

  • Analytical balance.

Protocol
Making M9*
  1. Produce M9 media as usual. Dilute 100x trace elements for a final concentration of 1x. 2 mM magnesium sulphate and 0.1 mM calcium chloride;

  2. Use 40 mM glycerol as a carbon source;

  3. Add 20 mM HEPES buffer. Use 2 mM phosphate (monobasic) and 1 g/L ammonium chloride as a phosphate and nitrogen source, respectively;

  4. Supplement 0.5 g/L sodium chloride;

  5. Adjust the pH till 7.00;

  6. Filter sterilise and fill till final volume with sterile MilliQ.

Calibration lines for different media types
  1. From the 1M standard stock, make a 1 mM stock in the calibration media used.

  2. Make analytical standards in the media wanted for 0 (blank/negative control), 1, 2, 5, 10, 20, 30, 40, 60, 80, 100 µM. 10 blanks were used and 5 technical replicates per concentration.

  3. Add 100 µL of the standards in a 96 well microplate in pre-determined locations.

  4. Make a fresh mix of reagents before each measurement and store the chemical stocks in the dark. 100 µM of Ali:Lan is used. All wells will consist of the following compositions, although the acetone is only added after incubation. Therefore, it is excluded from the pre-made master mix:

    Table 3. Composition of the master mix per well. Acetone is added to the well after the incubation time has passed and is therefore not pre-mixed. Reagent concentrations are found in the Reagents section.
    Chemical Add for 100 µM Ali:Lan (µL/well)
    Alizarin dye 6
    Lanthanum 6
    Acetate buffer 15
    Acetone 75
    MilliQ 98
    Total 200
  5. Add 125 µL of master mix in each well using a multichannel pipette.

  6. Incubate for 1 hour. During the incubation, cover the (black) well plate in aluminium foil to limit light exposure.

  7. Directly after 1 hour, add 75 µL acetone and measure the plate for absorbance in a microplate reader between 400-700 nm.

  8. Perform data analysis on the absorbance data.

Incubation interference testing
  1. Grow E. coli MG1655 in M9 overnight.

  2. Spin down the overnight culture at 3000 g for 10 minutes

  3. Resuspend the pellet in 3 mL 20 mM HEPES. This step is performed to wash the cell pellet and minimise interference from the growth media.

  4. Spin down the overnight culture at 3000 g for 10 minutes

  5. Resuspend the pellet in 3 mL 20 mM HEPES. Incubate the E. coli MG1655 for 93 hours.

  6. Spin down the overnight culture at 3000 g for 10 minutes and harvest the cell supernatant.

  7. Spike the supernatant with fluoride, making a calibration of 5, 10, 20, 30, 40 µM fluoride. Load in a plate with triplicates.

  8. Measure the samples between 400-700 nm.

  9. Follow the data analysis protocol for incubation interference determination against the calibration line of HEPES. The different types of observable interferences can be found there.

PFAS interference testing
  1. The MFA, DFA, TFA and PFOA in 20 mM HEPES is spiked with both 20 and 40 µM fluoride from the 1 mM NaF stock. Load the samples on plate in duplicates.

  2. Measure the samples between 400-700 nm.

  3. Follow the data analysis protocol for PFAS interference determination against the calibration line of HEPES.

  4. Analyse the obtained recovery value. The lower the recovery, the larger the interference.

Data analysis

Each well is reduced to a single absorbance ratio, that ratio is converted to fluoride through a media dependent calibration line, and the line is re-zeroed on the plate’s negative control before quantification. Error bars are the standard deviation of the replicate wells.

Building the dataframe
  1. Read every plate into a table with wavelength (nm) as the index and one column per well (A1…H12\mathrm{A1}\ldots\mathrm{H12}), taking the well names from row 1 of the sheet (can change depending on the plate reader type).

  2. Clean the values: convert comma decimals to points, blank cells to missing, and keep only rows whose wavelengths fall inside the measured window (400-700 nm).

Dividing wavelengths
  1. For a chosen wavelength, take the absorbance at the nearest measured point.

  2. Divide a numerator wavelength by a denominator wavelength, per well:

    R=AnumAdenR = \frac{A_{\mathrm{num}}}{A_{\mathrm{den}}}
  3. Skip wells whose denominator is missing or effectively zero, to avoid dividing by noise.

  4. Use RR, never raw absorbance, as the signal in all regressions. The division cancels out concentration and pathlength/volume differences between wells.

Finding the isosbestic point
  1. Average the replicate wells at each fluoride concentrations to get one mean spectrum per concentration, A‾c(λ)\bar{A}_c(\lambda).

  2. At every wavelength, measure how far the spectra move apart from the blank:

    S(λ)=mean over concentration of |A‾c(λ)−A‾0(λ)|S(\lambda) = \text{mean over concentration of } \bigl|\bar{A}_c(\lambda) - \bar{A}_0(\lambda)\bigr|
  3. The isosbestic point is where that spread is smallest (lines "cross"), searched between 480 and 640 nm (This requires at least three different fluoride concentrations):

    λiso=argminλS(λ)\lambda_{\mathrm{iso}} = \arg\min_{\lambda} S(\lambda)
  4. Determine λiso\lambda_{\mathrm{iso}} separately for every medium; it is a property of the media, not of the assay.

Choosing the best numerator
  1. Pin the denominator to that medium’s own isosbestic point. There the absorbance does not change with fluoride, so the ratio moves only because the numerator moves, which makes it a good internal normalisation.

  2. Scan candidate numerators on a 5 nm grid between 620 and 680 nm, where the fluoride complex absorbs.

  3. Discard a pair unless it satisfies all of:

    • the two wavelengths are at least 40 nm apart;

    • the denominator carries real dye absorbance, not a near-zero baseline;

    • the ratio moves by at least 20 % across the ladder;

    • the slope is positive.

  4. Fit every pair matching the mentioned criteria over the calibration range and keep its slope, R2R^2 and LOD.

  5. Accept pairs with R2≥0.95R^2 \ge 0.95, and among those select the one with the lowest LOD. Lock that pair for the medium.

Linear regression
  1. Fit ŷi=bxi+a\hat{y}_i = bx_i + a by least squares (xx = nominal fluoride concentration in µM, yy = locked ratio; minimum four points over three concentrations), then:

    sy=SSresn−2andR2=1−SSresSStots_{y} = \sqrt{\frac{SS_{\mathrm{res}}}{n-2}}\qquad\text{and} \qquad R^{2} = 1 - \frac{SS_{\mathrm{res}}}{SS_{\mathrm{tot}}}

    with SSres=∑(yi−ŷi)2SS_{\mathrm{res}} = \sum (y_i - \hat{y}_i)^2 (scatter about the line) and SStot=∑(yi−y‾)2SS_{\mathrm{tot}} = \sum (y_i - \bar{y})^2 (total scatter about the mean). R2R^2 is only descriptive per situation, since it rises with the width of the concentration range.

LOD and LOQ
  1. With bb the slope of the calibration line and σ\sigma the standard deviation:

    LOD=3.3σbandLOQ=10σb\mathrm{LOD} = \frac{3.3\,\sigma}{b}\qquad\text{and}\qquad \mathrm{LOQ} = \frac{10\,\sigma}{b}
  2. When there are replicates for the blank, take σ\sigma as the standard deviation of the 0μ0~\muM wells. This needs more than two blank wells (>2).

  3. When there are no replicates (one well per condition, as in the reagent/time optimisation), take σ\sigma as the residual standard deviation sys_y of the fit. The limit is then a property of the curve rather than of repeated measurement.

  4. Carry forward the LOD and LOQ of the reference (fresh HEPES) curve over its validated range, and draw them uncorrected as flat horizontal lines in every figure.

Linear ceiling
  1. Sort the different concentrations from low to high, and fit only the lowest three concentrations for the linearity, then the lowest four, and so on up to all the concentrations.

  2. Record the highest concentration at which the calculated fit still holds R2≥0.98R^2 \ge 0.98. That value is the linear ceiling.

  3. Report it per condition as a descriptive ceiling. It shows where the response starts to saturate, which is also where forcing a straight line would inflate the error and the LOD with it.

Interfering samples against the calibration line
  1. Spike the medium under test (5 mM of each PFAS individually, or an incubated sample containing cells) with known fluoride concentrations.

  2. Read these wells at the optimised wavelength pair and back-calculate them against the unshifted calibration line, without re-zeroing, dilution or thresholding.

  3. Fit the spiked matrix separately, giving slope b′b' and intercept a′a', and compare it with the interference free calibration line on two measures: the relative response b′/bb'/b, and the intercept offset converted to concentration units,

    Δ=a′−ab[μM]\Delta = \frac{a' - a}{b} \quad [\mu\mathrm{M}]
  4. Plot back-calculated against pipetted concentration, with the identity line y=xy = x for reference (everything on this line is perfectly in agreement, thus no interference is present)

  5. Read the two measures of the slope and intercept separately, as they point to different problems.

    • Slope close to the original calibration line, but all points sit above or below the y=xy = x line: The matrix shifts the blank, not the response. The bias is constant across the whole range and equal to Δ\Delta. re-zeroing of the blanks (intercept) in the media removes it.

    • Slope clearly not parallel (near 100%): The interference affects the response (complex) itself, so the error grows with concentration. Re-zeroing does not help here. A separate calibration of the medium is needed instead.

Controls & Purpose

Negative fluoride control (0 µM fluoride)

Defines the zero point. Can be used to calibrate the spectra and define the LOD and LOQ.

Positive fluoride control (40 µM fluoride spike)

Confirms that the complex responds to fluoride with an expected colour and spectral shift. High enough for easy visual determination.

Negative media control (M9)

M9 gives large interferences. See the interference of phosphate in the media.

Incubated spiked HEPES media

To determine if incubating with cells gives rise to interference and to quantify the extent.

Fluorochemical in spiked HEPES media

To quantify the interference of fluorochemicals in the HEPES media.

Results

M9 medium was measured with concentrations between 0 and 20 µM of fluoride. Even with the optimised ratio of A645/A640, a clear linear trend is not visible. The variance within each concentration is large. The HEPES buffered M9* does show the formation of the peaks around 620 nm, as described for demi water in Figure 6. We proved that E. coli MG1655 is able to grow in the M9* medium by a microplate reader growth experiment.

Growth on M9*

E. coli grows wells in M9. However, this was unknown for M9*, after making a novel medium, the growth was checked by a microplate reader experiment using E. coli MG1655 expressing 6 different enzymes. Absorbance at 600 nm resulted in peak values between 0.20 and 0.40 (Figure 7).

E. coli expressing different enzymes grown on the novel M9(HEPES) medium
Figure 7. E. coli expressing different enzymes grown on the novel M9(HEPES) medium

The growth media were analysed between 0-20 µM and 0-100 µM for M9 and M9*M9^*, respectively. For HEPES and tap water, the full fluoride range was tested as well. All the spectra with the blank subtracted are shown in Figure 8.

When analysing the spectra of the different media, the HEPES and tap water share similar UV-VIS spectra (like in demi/MilliQ water) with less variability within the same concentration compared to the defined growth media.

The UV-VIS spectra of the 4 media tested are shown: M9 (A), M9^* (B), 20 mM HEPES (C), and tap water (D). The spectra is normalised with the blank. The red dotted lines show the isosbestic point (denominator) while the optimised numerator for each media is indicated by the blue dotted lines.
Figure 8. The UV-VIS spectra of the 4 media tested are shown: M9 (A), M9*M9^* (B), 20 mM HEPES (C), and tap water (D). The spectra is normalised with the blank. The red dotted lines show the isosbestic point (denominator) while the optimised numerator for each media is indicated by the blue dotted lines.



We plotted the ratio signals, optimised between 0-40 µM like before, against the spiked concentrations (Figure 9;, saturation of the signal is observed above 40 µM of fluoride. Within the 0-40 µM range, good linearity is shown for the non-growth media, while the M9* is a more qualitative measurement media above the LOD.

The divided optimal ratio’s of the 4 media tested are shown: M9 (A), M9^* (B), 20 mM HEPES (C), and tap water (D).
Figure 9. The divided optimal ratio’s of the 4 media tested are shown: M9 (A), M9*M9^* (B), 20 mM HEPES (C), and tap water (D).
Interferences

Interference of a 93 hours spiked incubated medium was tested. The spiked samples are compared against the HEPES calibration curve and is shown in Figure 10. The spike slope is a factor 0.973 the calibration slope with an intercept offset of -0.0193.

The effects of 93 hours of incubation of MG1655 are shown. The slope is almost 100% (parallel) to the ideal line, the intercept is dropped by -0.0193 (b). Interference does not change a lot based on the concentration of fluoride.
Figure 10. The effects of 93 hours of incubation of MG1655 are shown. The slope is almost 100% (parallel) to the ideal line, the intercept is dropped by -0.0193 (b). Interference does not change a lot based on the concentration of fluoride.

There was a chance that PFAS could interfere with the measurement signal, influencing the screening and quantification. The interference of four different fluorochemicals, MFA, DFA, TFA, PFOA, is shown in Figure 11 by spiking 20 mM HEPES with 5 mM of fluorochemical. 20 µM fluoride results in roughly 60% recovery of the signal, while 40 µM results in  70%. Higher fluoride concentrations cause less interference when, again, comparing the difference with the HEPES calibration line.

The recovery of the signal compared to the HEPES calibration line. All PFAS interfere based on the fluoride concentration in the media.
Figure 11. The recovery of the signal compared to the HEPES calibration line. All PFAS interfere based on the fluoride concentration in the media.

Discussion & Troubleshooting

M9* has proven itself as a potential media for real time screening under growth by showing cell growth in this first test. In general, the growth media are not good for quantitative determination for low fluoride concentrations; M9 is not usable at all, while M9* can be used in a qualitative way above the LOD. This would allow you to evolve/screen enzymes at the same time as growing them. 20mM HEPES and tap water are usable for fluoride quantification. The media can have different purposes depending on the research question and are summarised in Figure 12.

Comparison of the different calibrated media. Depending on the metrics, a media can be selected for an application.
Figure 12. Comparison of the different calibrated media. Depending on the metrics, a media can be selected for an application.


A difference in incubated spiked HEPES against the (non-incubated) HEPES calibration line is found over the total linear range of HEPES. A parallel shift from the calibration line for the fit through the incubated spiked samples, means that there is a general loss of signal by incubation instead of a concentration dependent competing species in the complex reaction. Real incubated can therefore be re-zeroed on the negative control/blank for enzymatic tests.

Interference by PFAS is present for all PFAS in roughly the same order and likely competing with fluoride in a concentration dependent mechanism. This could decrease the limit of detection. However, defluorinating enzymes with a production rate outside of the linear range will need dilutions, minimising the total influence of PFAS in the total signal measured. As this is a limitation of the measurement, future optimisation could focus on making a calibration curve with hard to break down PFAS taken into account for the best LOD. For easy to degrade PFAS, the assumption is made that the fluoride concentration is high enough for dilutions and breaking down of PFAS changes its concentration in the medium, minimising the interference.

Real enzyme tests and bigger picture

Context

With a working mechanism, visual screening, quantifiable media and mapped interferences for our measurement, we advanced towards enzyme defluorination testing. Here, two well working enzymes for small fluorochemicals, H1ML (BBa_26GE2RMX) (E. coli) and H1 (BBa_26FK7Q5U) (P. putida), in different bacterial hosts are grown, washed, incubated, screened, diluted, and quantified for defluorination on four fluorochemicals. Bygd et al., (2022) also looked at these hosts qualitatively above 30 µM. Being able to determine and quantify very low fluoride concentrations of hard to degrade TFA, would be very valuable.

Aim

To be able to screen enzymes for defluorination activity on four fluorochemicals, quantify in different bacterial hosts and differentiate between different fluoride production rates for wide range screening of potential enzyme targets and comparison of evolved enzymes.

Materials & Equipment
Reagents
  1. 5 mM Alizarin-3-methyliminodiacetic acid (Ali);

  2. 5 mM Lanthanum(III) nitrate hexahydrate (Lan);

  3. Demineralised water;

  4. 1.68 M acetate buffer (pH 5.2);

  5. Acetone (>99.9%);

  6. 1 mM sodium fluoride (NaF) in MilliQ;

  7. M9 media (0.1 mM cumate, 50 µg/mL kanamycin, 20 mM glycerol);

  8. E. coli MG1655 WT;

  9. E. coli MG1655 expressing H1ML (BBa_26GE2RMX);

  10. P. putida KT2440 empty vector;

  11. P. putida KT2440 expressing H1 (BBa_26FK7Q5U).

Consumables
  1. 96 microwellplate;

  2. P200, P1000 pipette tips;

  3. 1.5 mL microcentrifuge tubes;

  4. 40 mL Falcon tube;

  5. Multichannel pipette reservoirs;

  6. Aluminium foil.

Equipment
  1. Micropipette (P200, P1000);

  2. Multichannel pipette (P200);

  3. Plate reader (400-700 nm spectral analysis);

  4. pH meter;

  5. Analytical balance.

Protocol
Incubation enzyme testing
  1. Grow E. coli MG1655 in M9 overnight.

  2. Spin down the overnight culture at 3000 g for 10 minutes

  3. Resuspend the pellet in 3 mL 20 mM HEPES. This step is performed to wash the cell pellet and minimises interference from the growth media.

  4. Spin down the overnight culture at 3000 g for 10 minutes

  5. Resuspend the pellet in 3 mL 5 mM fluorochemical (MFA, DFA, TFA, PFOA) 20 mM HEPES. Incubate the E. coli MG1655 for 93 hours.

  6. Take 250 µL samples at 3, 20, 46, and 93 hours and store them at -20: Spin down the overnight culture at 3000 g for 10 minutes and harvest the cell supernatant.

  7. Load the samples in duplicate in a well plate with triplicates.

  8. Measure the samples between 400-700 nm.

  9. Follow the data analysis protocol for well determination.

  10. Dilute if the calculated concentrations are out of the linear range.

Data analysis

Each well is reduced to a single absorbance ratio; that ratio is converted to fluoride through a media dependent calibration line. For this part, the time points are re-zeroed on the negative controls and the HEPES calibration were used. Error bars are the standard deviation of the replicate wells.

Building the dataframe
  1. Read every plate into a table with wavelength (nm) as the index and one column per well (A1…H12\mathrm{A1}\ldots\mathrm{H12}), taking the well names from row 1 of the sheet (can change depending on the platereader type).

  2. Clean the values: convert comma decimals to points, blank cells to missing, and keep only rows whose wavelengths fall inside the measured wavelength window (400-700 nm).

LOD and LOQ
  1. With bb the slope of the calibration line and σ\sigma the standard deviation:

    LOD=3.3σbandLOQ=10σb\mathrm{LOD} = \frac{3.3\,\sigma}{b}\qquad\text{and}\qquad \mathrm{LOQ} = \frac{10\,\sigma}{b}
  2. When there are replicates for the blank, take σ\sigma as the standard deviation of the 0μ0~\muM wells. This needs more than two blank wells (>2).

  3. When there are no replicates (one well per condition, as in the reagent/time optimisation), take σ\sigma as the residual standard deviation sys_y of the fit. The limit is then a property of the curve rather than of repeated measurement.

  4. Carry forward the LOD and LOQ of the reference (fresh HEPES) curve over its validated range, and draw them uncorrected as flat horizontal lines in every figure.

Well quantification
  1. Pick the calibration line that matches the well’s media: HEPES for every enzyme in this case.

  2. Divide the two locked wavelengths of that curve to get the well’s ratio:

    y=AnumAdeny = \frac{A_{\mathrm{num}}}{A_{\mathrm{den}}}
  3. Fill that ratio into the calibration line y=ax+by = ax + b and solve for the concentration, using the re-zeroed (from blank) intercept:

    x=y−bax = \frac{y - b}{a}
  4. Multiply by the well’s dilution factor (1 if undiluted) to get the concentration in the original sample.

  5. Wells that are outside of the linear range of the calibration are rediluted and measured till matching the range.

Controls & Purpose

Negative E. coli MG1655 control (WT)

Should not have any defluorination activity; therefore, useful as a negative biological control and to re-zero at incubated time points.

Positive E. coli MG1655 control (H1ML)

Should have defluorination activity on monofluoroacetate, therefore useful as a positive biological control.

Negative P. putida control (Empty vector)

Should not have any defluorination activity; therefore, it is useful as a negative biological control and to re-zero at incubated time points.

Positive P. putida control (Defluorinase)

Should have defluorination activity on monofluoroacetate, therefore useful as a positive biological control for showing host versatility.

Positive defluorination control (Monofluoroacetate)

Monofluoroacetate should be broken down by fluoroacetate dehalogenases and defluorinases for short chain fluorochemicals.

Negative defluorination control (PFOA)

PFOA should not be broken down by fluoroacetate dehalogenases and defluorinases for short chain fluorochemicals.

Results

After expressing H1ML in E. coli MG1655, washing and resuspending the cells in 20 mM HEPES containing 5 mM PFAS, the taken and analysed samples from the incubation medium are shown in Figure 13. H1ML FAcD increased fluoride in the medium for MFA and DFA, with the steepest increase in the fluoride concentration in the first 2 days. Samples for the 46 and 93 hours, were diluted 50x for MFA/DFA to be inside the linear range of the measurement. TFA and PFOA do not get above the LOD threshold and is similar to the negative control.

The defluorination by E. coli expressing dehalogenase H1ML on 4 fluorochemicals was shown. Mono- and difluoroacetate were defluorinated, while trifluoroacetate and PFOA were not.
Figure 13. The defluorination by E. coli expressing dehalogenase H1ML on 4 fluorochemicals was shown. Mono- and difluoroacetate were defluorinated, while trifluoroacetate and PFOA were not.

Usability of different bacterial hosts is visible in Figure 13 (E. coli) and Figure 14 (P. putida). The gene H1 expressed in P. putida is compared in the media containing 4 fluorochemicals against a putida strain with the same, but empty vector. The gene was able to defluorinate both MFA and DFA effectively, although less than E. coli. A drop in defluorination between 46 and 93 hours is observed for MFA. TFA defluorination stays below the LOD, as do all data points for PFOA, except the 20 hours measurement, which is likely an artifact.

The defluorination by P. putida expressing dehalogenase H1 on 4 fluorochemicals was shown. Mono- and difluoroacetate were defluorinated, while trifluoroacetate and PFOA were not.
Figure 14. The defluorination by P. putida expressing dehalogenase H1 on 4 fluorochemicals was shown. Mono- and difluoroacetate were defluorinated, while trifluoroacetate and PFOA were not.

Discussion & Troubleshooting

The positive control MFA has shown defluorination, while PFOA, the negative control, was not defluorinated by the enzymes. It is demonstrated that, with the appropriate dilution factor, even high fluoride concentrations can be measured. The fact that DFA is harder to break down than MFA (and TFA is even harder) was reiterated by our results. In parallel, the DFA defluorination proves that low concentrations of fluoride (3/20h samples for both hosts) can be measured as well. Therefore, differences in defluorination rate between enzymes and substrates can be tested.

Additionally, the ability to use other bacterial hosts in this measurement technique was demonstrated. P. putida is able to break down MFA and DFA. Unexplainable noise signals going up (and decreasing again) in fluoride are visible for the defluorinase and empty vector P. putida strain in MFA, TFA and PFOA. This is likely noise, but since the method was optimised for E. coli, the method could still be less stable for other hosts. In the case of PFOA, it even peaks above the LOD (20h), but the same level is not reached in later time points. One possible explanation can be the use of the HEPES calibration curve, which was proven not to be affected by incubation of E. coli too much, but might be affected by P. putida incubation.

Conclusion

We developed and documented a colourimetric/absorbance assay using alizarin dye and lanthanum for fluoride quantification. Starting from a method that reported qualitative measuring, we fixed the 1:1 reagent concentration to 100 µM, used 1 hour incubation time, and made a rational data analysis for the ratio’s and interferences. The denominator sits at the isosbestic point to cancel out small changes in reagent concentrations and path length. This gave us a linear range between 0-40 µM fluoride in demi water, 20 mM HEPES and tap water.

Equally important is where the method does not work. M9 media is unusable, while the M9* we formulated supports growth in E. coli and allows for qualitative screening above the LOD. This could allow for skipping an extra incubation step by growing and screening/evolving at the same time, which is an important improvement compared to the previous study. Therefore, the next step would be to investigate this further to obtain real time sensing. 5 mM concentrations do interfere with the measurement, resulting in lower total recovery of the signal, which can be solved with dilution of the fluorochemicals. The technique was applied to H1ML and fluorinase in E. coli and P. putida. The assay could distinguish between the defluorination rates between the enzymes and on various fluorochemicals. Measurements could be performed in parallel, and high (>20 µM) concentrations could be screened by eye. The protocol, controls, and data analysis can provide the basis for any team working on C-F bond cleavage and be adopted, instead of completely rebuilding it.

References

Show all references
[1] Bygd, M. D., Aukema, K. G., Richman, J. E., & Wackett, L. P. (2022). Microwell Fluoride Screen for Chemical, Enzymatic, and Cellular Reactions Reveals Latent Microbial Defluorination Capacity for −CF3_{\textrm{3}} Groups. Applied and Environmental Microbiology, 88(9), e00288–22. https://doi.org/10.1128/aem.00288-22
[2] Belcher, R., & West, T. S. (1961). A comparative study of some lanthanon chelates of alizarin complexan as reagents for fluoride. Talanta, 8(12), 863–870. https://doi.org/10.1016/0039-9140(61)80212-9