Skip to content
ENCORE

Hardware

3D-Aid: Printable Hardware for Gel Electrophoresis

The finished 3D-Aid casting tray, printed in black carbon-fiber filament, with both end plates and their red silicone cord seals attached The same tray taken apart, with the two end plates lying on either side
Figure 1. 3D-Aid. Our printed gel casting tray in its final version, assembled and taken apart. Hover over the image or tap it to see its parts.

Lab Bottleneck

As the first iGEM team at Karlsruhe Institute of Technology (KIT), we had to establish parts of our experimental infrastructure while simultaneously developing our main project. During routine molecular biology work, we identified a simple but recurring limitation. We did not have enough gel electrophoresis casting trays and combs to efficiently run experiments in parallel, which slowed down our workflow.

Instead of purchasing additional components, we explored whether commonly used electrophoresis accessories could be reproduced locally using accessible FDM 3D printing. Existing laboratory components served as dimensional references. However, we quickly learned that directly reproducing commercial laboratory equipment was not sufficient. Material properties, printer resolution, sealing, dimensional tolerances and usability all had to be considered.

What initially appeared to be a simple 3D-printing task therefore developed into an iterative hardware engineering project.

Engineering Approach

We followed the iterative engineering process: Design → Build → Test → Learn → Redesign. Rather than considering unsuccessful prototypes as failures, each version helped us identify new design requirements and improve the following iteration.

Our final objective was to develop:

  • a functional gel electrophoresis casting tray for daily lab use
  • reusable, removable sealing components
  • compatible electrophoresis combs
  • combs with different well sizes
DesignBuildTestLearn3iterations
Figure 2. Our engineering cycle. We went through Design, Build, Test and Learn in three iterations before reaching the final version.
CAD models of our three tray iterations side by side: iteration 1 in teal, iteration 2 in yellow and iteration 3 in orange The same three iterations as exploded views, each with its comb and end plates
Figure 3. Three iterations side by side. CAD models of our three iterations, from left to right: iteration 1 (teal), iteration 2 (yellow) and iteration 3, our final design (orange). Hover over the image or tap it to see each one with its comb and end plates. The final design is shown in detail in Figure 8.

Key Lessons

Measuring is not Designing

Commercial components are optimized for their original manufacturing method, so a direct one-to-one reproduction performed poorly when printed.

Manufacturing Constraints are Design Constraints

Printer resolution, support structures and dimensional tolerances became design constraints in their own right.

Material is a functional Parameter

Material choice proved to be a functional parameter rather than a detail. Moving the tray from PLA to a carbon-fiber reinforced filament at higher infill, for example, gave it the weight and stability a usable tray needs.

Gray PLA casting tray from our second iteration, with a red silicone cord seal along its edge
Figure 4a. Iteration 2: the tray printed in PLA.
Black casting tray from our third iteration, printed in GreenTEC Pro Carbon Fiber, with both end plates attached
Figure 4b. Iteration 3: the final tray in GreenTEC Pro Carbon Fiber.

If you want to see exactly what we designed, built, tested, and learned in each iteration, including the ideas that did not make it into the final design, visit our Engineering Cycles page.

Engineering Cycles →

A modular, validated Tool

The final version is more than a prototype. We validated it across the complete workflow rather than through a single fit test, confirming that:

  1. Assemblythe components assemble correctly
  2. Sealingno leakage during casting
  3. Castingagarose gels pour cleanly
  4. Wellsthe combs form usable wells
  5. Runa successful electrophoresis from start to finish
The printed tray with its two removable end plates taken off and lying on either side The same tray assembled, with both end plates attached
Agarose gel poured into the sealed tray, with a comb inserted in the middle slot The set gel in the tray seen from above, with a row of wells across the middle
The tray with the gel placed in the electrophoresis chamber, seen from above The electrophoresis chamber with the tray inside and the electrodes connected
Gel image after electrophoresis under blue light, with a DNA ladder and a row of bands A second gel image after electrophoresis, with bands in several lanes
Figure 5. Validation across the workflow. (a) The tray with its two removable end plates and their red silicone cord seals, taken apart and assembled. (b) Agarose gel poured into the sealed tray with a comb in place, and the set gel from above. (c) The gel in the electrophoresis chamber, and the chamber connected for the run. (d) Gel images after electrophoresis runs with gels cast in our tray. Hover over a photo or tap it to see the second view.

The red cord visible in Figure 5 is a silicone round cord of 5 mm diameter, seated between the tray and each end plate. We added it after the first iteration, when we found that the printed tray was not leak-tight on its own. It is an off-the-shelf part that we bought online, sold by the meter and cut to length, so other teams can source it easily.

The comb system is modular by design. Different combs can be printed with varying numbers and widths of teeth, so the same casting tray can be adapted to different experimental needs without reprinting the tray itself.

Four printed combs lying one above the other, one gray, one blue and two green, with different numbers and widths of teeth The same four combs turned over, showing the other side of the teeth
Figure 6. The comb variants. The two green combs are the current versions. The gray and the blue one are from earlier iterations, kept here because they show how the design changed. Hover over the image or tap it to see them from the other side.
CombTeethWell volumeLengthTray-fit edges
Small wells12approx. 160 µl100 mm3.54 mm
Large wells8approx. 300 µl110 mm3.54 mm

Both are printed in PLA at 20% infill and need about 7 g of filament each, roughly 2.5 m. A comb is therefore cheap enough to print a new one whenever an experiment needs a different well size, rather than working around the combs you happen to own.

Final Configuration

ParameterValue
PrinterPrusa i3 MK3
Nozzle diameter0.4 mm
CAD softwareAutodesk Fusion
Tray dimensions150 × 110 × 30 mm
Tray materialGreenTEC Pro Carbon Fiber
Tray infill50%
Tray filament used132 g, approx. 40 m
Comb materialPLA
Comb infill20%
Comb filament used7 g each, approx. 2.5 m
Tray printing temperature230 °C
Comb printing temperature215 °C
SealsSilicone round cord, ø 5 mm
Our Prusa i3 MK3 3D printer in its enclosure in the workshop, printing a part, with filament spools on top Close-up of the print head and print bed while a part is being printed, with the printer display in front
Figure 7. Printing in our workshop. Our Prusa i3 MK3 producing a part during one of the design iterations. Hover over the image or tap it for a close-up of the print bed.

What it costs

The whole point of the project was to stop buying trays we could make. Looking at generic casting trays online, prices start above 100 € and climb from there depending on the quality and on how complete the system is. Our tray takes 132 g of filament, which comes to roughly 8 €.

That number is the filament alone. It does not include the printer, the electricity or the silicone cord. Those are one-off costs that are spread across every tray and comb we print, and a lab that already owns an FDM printer pays none of them again. For a team that needs several trays at once, which is exactly the bottleneck we started from, the difference is what made the project worth doing.

An open-source Contribution

What began as a fix for our own bottleneck became an alternative piece of laboratory hardware that other teams may be able to use as well. To make it useful beyond our lab, we are releasing the complete design package so that any iGEM team or small lab with access to an FDM printer can reproduce the system or adapt it to their own equipment. In our repository you can find our final printable STL files for the tray, its end plates and both combs, together with every dimension we used, the printing parameters and assembly notes.

Exploded view of the final CAD model: the casting tray with two comb slots, a comb above it and the two removable end plates on either side The same CAD model assembled: the comb inserted in the tray and both end plates in place
Figure 8. The final CAD model. Exploded view of the casting tray, its two removable end plates and a comb. Hover over the image or tap it to see the assembled system.

3D-Aid design package: printable STL files for the casting tray, its end plates and both combs. The print settings, every dimension and the assembly notes are in the README beside them.

Open the design files ↗

The files are released under the Creative Commons Attribution 4.0 International license (CC BY 4.0), the same license as the rest of this wiki. You may use them, adapt them and build on them, including commercially, as long as you credit iGEM Team KIT 2026. That is the whole point: we would rather someone improved this design than started again from an empty room.

Limitations, and what to change

Our tray is built around how we work. It takes two combs at the same time, which doubles the number of wells per gel, because we were running many samples in parallel and that was the whole reason the project started. For a lab with different habits that is not necessarily the right choice, and it does not have to be.

What we publish is the printable geometry rather than a parametric CAD source, so adapting the design means rebuilding it rather than editing a parameter. We have tried to make that as easy as we can: every dimension is in the tables above and in the README, and these are simple, angular parts. An STL can be scaled in any slicer, imported and reshaped in a free browser tool such as Tinkercad, or used as a reference to model the part again in an hour. The one dimension to check before anything else is the outer size of the tray, because that is what has to fit your own electrophoresis chamber.

The combs are the easy part. They are closer to templates than to finished parts: a different number of teeth or a different well volume is a simple shape to rebuild, and 7 g of filament to try. A team can match the wells to what they actually pipette instead of adapting their experiment to the comb they happen to own.