3D-Aid: Printable Hardware for Gel Electrophoresis
Lab Bottleneck
As the first iGEM team at Karlsruhe Institute of Technology (KIT), we had to establish parts of our experimental infrastructure while simultaneously developing our main project. During routine molecular biology work, we identified a simple but recurring limitation. We did not have enough gel electrophoresis casting trays and combs to efficiently run experiments in parallel, which slowed down our workflow.
Instead of purchasing additional components, we explored whether commonly used electrophoresis accessories could be reproduced locally using accessible FDM 3D printing. Existing laboratory components served as dimensional references. However, we quickly learned that directly reproducing commercial laboratory equipment was not sufficient. Material properties, printer resolution, sealing, dimensional tolerances and usability all had to be considered.
What initially appeared to be a simple 3D-printing task therefore developed into an iterative hardware engineering project.
Engineering Approach
We followed the iterative engineering process: Design → Build → Test → Learn → Redesign. Rather than considering unsuccessful prototypes as failures, each version helped us identify new design requirements and improve the following iteration.
Our final objective was to develop:
- a functional gel electrophoresis casting tray for daily lab use
- reusable, removable sealing components
- compatible electrophoresis combs
- combs with different well sizes
Key Lessons
Measuring is not Designing
Commercial components are optimized for their original manufacturing method, so a direct one-to-one reproduction performed poorly when printed.
Manufacturing Constraints are Design Constraints
Printer resolution, support structures and dimensional tolerances became design constraints in their own right.
Material is a functional Parameter
Material choice proved to be a functional parameter rather than a detail. Moving the tray from PLA to a carbon-fiber reinforced filament at higher infill, for example, gave it the weight and stability a usable tray needs.
If you want to see exactly what we designed, built, tested, and learned in each iteration, including the ideas that did not make it into the final design, visit our Engineering Cycles page.
Engineering Cycles →A modular, validated Tool
The final version is more than a prototype. We validated it across the complete workflow rather than through a single fit test, confirming that:
- Assemblythe components assemble correctly
- Sealingno leakage during casting
- Castingagarose gels pour cleanly
- Wellsthe combs form usable wells
- Runa successful electrophoresis from start to finish
The red cord visible in Figure 5 is a silicone round cord of 5 mm diameter, seated between the tray and each end plate. We added it after the first iteration, when we found that the printed tray was not leak-tight on its own. It is an off-the-shelf part that we bought online, sold by the meter and cut to length, so other teams can source it easily.
The comb system is modular by design. Different combs can be printed with varying numbers and widths of teeth, so the same casting tray can be adapted to different experimental needs without reprinting the tray itself.
| Comb | Teeth | Well volume | Length | Tray-fit edges |
|---|---|---|---|---|
| Small wells | 12 | approx. 160 µl | 100 mm | 3.54 mm |
| Large wells | 8 | approx. 300 µl | 110 mm | 3.54 mm |
Both are printed in PLA at 20% infill and need about 7 g of filament each, roughly 2.5 m. A comb is therefore cheap enough to print a new one whenever an experiment needs a different well size, rather than working around the combs you happen to own.
Final Configuration
| Parameter | Value |
|---|---|
| Printer | Prusa i3 MK3 |
| Nozzle diameter | 0.4 mm |
| CAD software | Autodesk Fusion |
| Tray dimensions | 150 × 110 × 30 mm |
| Tray material | GreenTEC Pro Carbon Fiber |
| Tray infill | 50% |
| Tray filament used | 132 g, approx. 40 m |
| Comb material | PLA |
| Comb infill | 20% |
| Comb filament used | 7 g each, approx. 2.5 m |
| Tray printing temperature | 230 °C |
| Comb printing temperature | 215 °C |
| Seals | Silicone round cord, ø 5 mm |
What it costs
The whole point of the project was to stop buying trays we could make. Looking at generic casting trays online, prices start above 100 € and climb from there depending on the quality and on how complete the system is. Our tray takes 132 g of filament, which comes to roughly 8 €.
That number is the filament alone. It does not include the printer, the electricity or the silicone cord. Those are one-off costs that are spread across every tray and comb we print, and a lab that already owns an FDM printer pays none of them again. For a team that needs several trays at once, which is exactly the bottleneck we started from, the difference is what made the project worth doing.
An open-source Contribution
What began as a fix for our own bottleneck became an alternative piece of laboratory hardware that other teams may be able to use as well. To make it useful beyond our lab, we are releasing the complete design package so that any iGEM team or small lab with access to an FDM printer can reproduce the system or adapt it to their own equipment. In our repository you can find our final printable STL files for the tray, its end plates and both combs, together with every dimension we used, the printing parameters and assembly notes.
3D-Aid design package: printable STL files for the casting tray, its end plates and both combs. The print settings, every dimension and the assembly notes are in the README beside them.
Open the design files ↗The files are released under the Creative Commons Attribution 4.0 International license (CC BY 4.0), the same license as the rest of this wiki. You may use them, adapt them and build on them, including commercially, as long as you credit iGEM Team KIT 2026. That is the whole point: we would rather someone improved this design than started again from an empty room.
Limitations, and what to change
Our tray is built around how we work. It takes two combs at the same time, which doubles the number of wells per gel, because we were running many samples in parallel and that was the whole reason the project started. For a lab with different habits that is not necessarily the right choice, and it does not have to be.
What we publish is the printable geometry rather than a parametric CAD source, so adapting the design means rebuilding it rather than editing a parameter. We have tried to make that as easy as we can: every dimension is in the tables above and in the README, and these are simple, angular parts. An STL can be scaled in any slicer, imported and reshaped in a free browser tool such as Tinkercad, or used as a reference to model the part again in an hour. The one dimension to check before anything else is the outer size of the tray, because that is what has to fit your own electrophoresis chamber.
The combs are the easy part. They are closer to templates than to finished parts: a different number of teeth or a different well volume is a simple shape to rebuild, and 7 g of filament to try. A team can match the wells to what they actually pipette instead of adapting their experiment to the comb they happen to own.