Dry Lab · Hardware

Hardware

Building a reader around light, airflow and living samples

Abstract

AeroSense combines four replaceable 100 µL wells, frequency-tagged blue LEDs, a filtered photodiode, ESP32 acquisition and a local phone interface. An eight-part enclosure fixes the optical geometry and separates sample gas from electronics cooling. Optical and airflow simulations guided filter selection and the inlet shelf; assembly feedback informed the manual. Fluorescein measurements resolve four frequency channels, quantify cross-well responses and test calibration under isolated and simultaneous loading.

Introduction

Monitoring volatile changes around stored produce requires a practical link between receptor biology and an interpretable measurement. Our sensing strategy pairs insect odorant receptors with the calcium indicator GCaMP6f so that receptor activation can be read optically. [13] [14] The hardware task is to expose several cell samples to the same gas input and record their optical responses in a compact reader with replaceable wells.

A difference between channels should reflect the samples being compared, rather than differences in gas delivery or light collection. We therefore need to control the gas path, keep each well in a fixed position and limit excitation light reaching the detector. Using one photodiode simplifies the detector layout, but requires distinct LED modulation frequencies and digital lock-in processing to recover the excitation channels. These requirements connect the enclosure, optics, electronics and acquisition software.

The exploded view below shows how the gas lid, wells, optical path and electronics fit together. The following sections connect each part to its design question and supporting figures, then examine cell recovery, system verification and the manuals revised after assembly feedback.

Airflow

All four wells share one gas input, so a difference in exposure could be confused with a difference in receptor response. We compared the open lid, perforated plate and integral shelf to ask how each directs the inlet jet and delivers gas toward the liquid surfaces.

The printed gas lid carries an inlet and outlet on opposite sides of one shared chamber. Its integral shelf redirects the incoming stream above the four well openings. A removable compression plate presses the sealing stack around the wells, while flexible couplers connect the gas ports to tubing through the side openings.

Inlet filter→Gas lid and shelf→Shared well headspace→Outlet filter→Controlled suction

In the video, first compare jet direction and speed across the three geometries, then follow the concentration front as the shelf flow rate changes. The arrival-time calculation uses concentration beside each liquid surface, connecting the flow pattern to sample exposure.

Video 1. Simulated airflow. The first view compares three geometries at 3 mL/min, colored by speed; the second compares shelf flow rates, colored by gas concentration. At the 0.4 mm grid spacing, the shelf reduces the downward-momentum proxy by 31% relative to the open lid. Particle markers show advection; the simulation clock restarts between views.

At 3 mL/min with 100 µL in each well, the model predicts gas concentration beside the liquid surface reaching 10% of the inlet concentration in 5.30–6.29 s and 90% in 47.80–49.42 s.

At matched conditions (3 mL/min, 100 µL/well and D = 10 mm²/s), the open chamber, 12-hole plate and current shelf had normalized downward-momentum proxies of 1.00, 0.57 and 0.69. Their slowest-well t90 values were 51.67, 48.36 and 49.42 s, with interwell spreads of 4.64, 2.83 and 1.62 s, respectively. We retained the shelf: it reduced the momentum proxy by 31% and brought all four wells to t90 within approximately 50 s, meeting our provisional 60 s gas-transport target. Reusing the completed integrated enclosure also avoided another fabrication cycle. The plate reduced momentum more and remains a candidate for the next version.

Here, t90 is the time for mean gas concentration beside each liquid surface to reach 90% of the inlet value; it excludes external tubing and cellular response. Similar velocities do not guarantee matched concentration arrival, because diffusion also affects transport. The fixed-surface model describes gas delivery beside the medium. A coupled fluid–fluid model would need to solve both phases and their interfacial force balance, as described by COMSOL [24].

Gas lid and wells

Source CAD drawing of the gas lid, integral inlet shelf and opposite gas ports
Figure 1. The gas lid integrates a 7.2 × 8 × 1.2 mm inlet shelf within the nominal 14.8 × 12 mm chamber. The shelf spreads the incoming jet above the well openings; dimensions are from the CAD model.
Source CAD section showing the four replaceable wells, compression plate and sealing stack
Figure 2. Separate liquid samples sit beneath the shared gas chamber. The section shows the light-blocking stack and nominal 100 µL liquid surface calculated from the inner well cross-section.

Each well is operated at 100 µL (approximately 120 µL brim capacity). Detailed gas-delivery and liquid-transport simulations are presented in Model →.

Optics

Well illumination

Our design uses one photodiode to read four wells, with LED modulation distinguishing the excitation channels. Light can still reach neighboring wells, so a frequency channel may contain fluorescence from more than its intended sample. The optical model therefore asks how well the geometry preserves each well’s contribution.

In the transfer matrix below, compare the diagonal contributions from the intended well with off-diagonal contributions from neighboring wells. Detailed simulations and the optical-path animation are presented in Model →.

Canonical floor-layer simulation: row-normalized response percentages for D1–D4
Figure 3. The canonical optical design model, also used in Model and the measured comparison below. Each row is normalized to 100%; rows are LED / readout channels and columns are fluorescent source wells. The original 20 µm floor layer, 0.5 mm⁻¹ absorption coefficient and three-seed mean are unchanged. Intended-well shares are 88.6–90.9%.

The intended well contributes 88.6–90.9% of each modeled readout. The same matrix is used in Model and in verification Figure 15, which compares the displayed response distributions with measured single-well controls. These percentages describe response distribution, not a fraction of all emitted fluorescence.

Filter selection and spectral alignment

The photodiode responds to both blue excitation and green reporter light. Filter selection therefore balances rejection of excitation leakage against retention of the signal band. The comparison below holds the source and geometry assumptions fixed: read the LED-derived and proxy-green currents separately, then compare the resulting raw peak.

Simulation comparing no filter with a single LEE 101 layer: LED-derived current 100 versus 2.987 percent; proxy-green current 100 versus 70.43 percent; raw peak 104.36 versus 4.09 nanoamps.
Figure 4. With the same geometry and assumed source spectra, one LEE 101 layer retains 2.987% of LED-derived current and 70.43% of the proxy-green current. The simulated peak falls from 104.36 to 4.09 nA. These are optical-model comparisons, rather than measured cell fluorescence. Full calculation.

A single LEE 101 layer retains 70.43% of proxy-green current while reducing LED-derived current to 2.987% of the unfiltered model. The filter sits over the collection aperture, leaving the excitation path open.

The filter comparison depends on where excitation, emission and detector sensitivity overlap. In the first panel, compare the LED spectrum with GCaMP6f excitation and locate its green emission. In the second, compare that emission band with filter transmission and photodiode responsivity to follow the collection path.

Two-panel spectral comparison. The LED model is centered at 468 nm beside published calcium-bound GCaMP6f excitation and emission curves. LEE 101 transmission increases across the green region, while the VEMD5060X01 photodiode responds to blue and green light.
Figure 5. Spectral alignment. GCaMP6f curves are digitized from the calcium-bound traces in Dana et al., Supplementary Figure 1; excitation is shown over 406–510 nm and emission over 500–646 nm, ending at the digitized limits. [20] The LED is the existing 468 nm Gaussian model with an assumed 25 nm full width at half maximum. [9] LEE 101 uses manufacturer transmission samples; the two-layer line is calculated as T². [12] Photodiode responsivity comes from the manufacturer's typical response curve and its 950 nm normalization. Spectral shapes are normalized; filter transmission and detector responsivity retain the separate scales shown on the axes. [10]

Wells and light paths

Four removable clear wells sit in a black printed carrier. Their flat floors, fixed spacing and open lower optical band connect the cell samples to the LEDs and central photodiode.

Excitation: the four sideview LEDs illuminate the samples through the lower optical band. Collection: green fluorescence passes through the clear well floors and the receiver filter to the photodiode. The filter covers the collection aperture while all four excitation paths remain open.

CAD plan of four clear wells in a black carrier, four sideview LEDs and the central photodiode, with labelled well-floor and optical-band dimensions.
Figure 6. The carrier positions four wells above the central detector and leaves a 1.7 mm lower optical band open. The photodiode rectangle is a placement reference; package orientation, optical heights and clearances require the component drawing and assembly check. The nominal 100 µL liquid surface follows the well geometry. Gas and liquid transport.

Electronics

Signal and control

To compare optical channels, we must convert photocurrent into a timed electrical record. The diagram traces the main readout path and distinguishes the LED-control and power branches.

Photodiode → TIA → ADC → ESP32, with separate LED-control and power branches.
Figure 7. Main readout path: photocurrent is amplified by the TIA, digitized by the ADC and recorded by the ESP32. Separate branches show LED control and power; component and pin details are given in the schematic below.

Schematic

Photodiode → TIA → ADC → ESP32

Four independently driven LEDs and one shared photodiode reduce the number of detector channels while preserving a route to frequency-resolved readout. This choice places the burden on stable acquisition and measured optical calibration.

The acquisition specification uses a 4 kHz timebase, two-second I/Q blocks and 41/67/89/113 Hz LED channels.

The acquisition retains in-phase and quadrature components at each excitation frequency, following Scofield’s frequency-domain treatment of lock-in detection. [1]

Reader schematic sheet 1: battery, charging, USB and boost supply
Figure 8. Reader schematic. Select a sheet to inspect component values.

Component calculations

To check whether the optical-model output fits the readout range, we propagated its predicted current through the nominal TIA gain and ADC reference. Read the first table from current to voltage to digital code; the component table then identifies the values that set gain, bandwidth and timing.

Optical power to ADC

P(λ) is incident spectral power density, T(λ) is filter transmission, and RPD(λ) is photodiode responsivity.

Table 1. Selected-filter simulation · nominal DC conversion
StageCalculationSelected-filter value
Raw detector currentIlight=∫P(λ)T(λ)RPD(λ)dλIraw=Ilight+Idark+Inoise4.087 nA simulated raw peak
TIA outputV≈Iraw×100MΩ0.4087 V
Ideal ADC codeN≈V2.5V×65,53610,714 codes

Spectral model and source conditions →

Table 2. Calculated from the released schematic, manufacturer specifications and firmware settings.
StageSource valuesCalculationDesign meaning
Photocurrent → voltageR27 = 100 MΩ; C21 = 2.2 pFVout≈Iphoto×100MΩfp=12π×100MΩ×2.2pF=723.43HzThe nominal feedback gain is 100 mV per nA. At the four excitation frequencies, its amplitude is 99.84%, 99.57%, 99.25% and 98.80% of DC gain. The photodiode polarity gives a positive output. No additional voltage-gain stage appears in the schematic.[7]
Voltage → digital codeU7 = 2.5 V; U9 = 16 bit2.5V216=38.147µV/code38.147µV/code100MΩ=0.38147pA/codeThese are ideal code scales. The reference span corresponds to 25 nA; usable signal range also depends on baseline, amplifier headroom and offsets. The LTC6655 B-grade reference specifies ±0.025% initial accuracy, equivalent to ±625 µV at 2.5 V.[8] [6]
Sampling and modulationFs = 4000 Hz; N = 8000Ts=250µsNFs=2s16bits2MHz=8µs10µs+8µs=18µsThe ADS8866 supports 100 kSPS; this firmware configures 4 kSPS. A 10 µs conversion wait exceeds the specified 8.8 µs maximum, followed by an 8 µs nominal SPI transfer plus software overhead. The 41/67/89/113 Hz channels complete 82/134/178/226 cycles per block, enabling coherent sine/cosine estimation.[8]
Excitation currentR16/R21/R24/R30 = 100 ΩI≈(5−3.4)V100Ω=16mAPR≈0.016A2×100Ω=25.6mWThis estimate uses the LED’s typical 3.4 V forward drop at its 20 mA datasheet test point. Default 25% duty gives approximately 4 mA average per channel and 6.4 mW average resistor dissipation. Actual current follows the LED I–V curve, rail voltage and tolerances.[9] [17]

The input interface and supply filters serve different purposes. C24 spans both ADC inputs, so R29 + R36 = 200 Ω gives a nominal 795.8 kHz differential pole. R34 is in series with shunt C28, while REF is fed directly. Neither network is a 2 kHz anti-alias filter. The 723 Hz TIA feedback pole attenuates higher frequencies, but unwanted light, switching transients and harmonics still require an aliasing and settling check. [8]

Full component audit and inventory

PCB design

The layout places signal paths beside power circuitry, LED drivers and the optical aperture. Compare front and back views to locate acquisition paths and connectors before assembly.

Front · component layout
Reader PCB front component layout from the supplied board source
Figure 9. Reader PCB, component layout and copper routing. Switch sides to follow power, excitation and signal acquisition. Open full-resolution source ↗.

Enclosure and cooling

Eight source parts

The enclosure fixes optical alignment while allowing sample replacement and electronics servicing. Compare assembled access with the exploded lid, wells, carrier and base; the following drawings specify nominal dimensions and clearances.

The 108 × 108 × 66 mm body contains an approximately 100 × 90 mm PCB and a 40 × 40 × 10 mm cooling fan. Four replaceable Aqua Clear Plus wells sit within a black PC/GF-like base, hood, gas lid and compression plate.

Black square AeroSense enclosure and exploded assembly of eight CAD parts, with a monochrome raised fly emblem and AEROSENSE lettering on the lid.
Figure 10. Assembled and exploded views rendered from the eight manufacturing CAD meshes. The monochrome fly emblem and AEROSENSE lettering are a proposed raised relief, about 0.8 mm high, shown only in the visualization; the printable STLs are unchanged. The base integrates the fan seat; the gas lid integrates the inlet shelf.

Dimensioned engineering drawings

These twelve sheets locate and dimension the enclosure, its internal features and all eight printed parts from the retained CAD and STL geometry. The assembly and principal part views use aligned, equal-scale third-angle projections; enlarged sections and details identify their relative scale. All dimensions are in millimetres. The CAD provides nominal dimensions and design clearances for fabrication and assembly checks.

Assembly and shared datums: dimensioned black-line CAD projections and feature coordinate schedule, sheet A01.

A01 — Assembly and shared datums. Open full-size lossless WebP.

Lower case / base — B01–B03
Lower case: shell and service apertures: dimensioned black-line CAD projections and feature coordinate schedule, sheet B01.

B01 — Lower case: shell and service apertures. Open full-size lossless WebP.

Lower case: floor, fan and mounting pattern: dimensioned black-line CAD projections and feature coordinate schedule, sheet B02.

B02 — Lower case: floor, fan and mounting pattern. Open full-size lossless WebP.

Lower case: fasteners and locating tongue: dimensioned black-line CAD projections and feature coordinate schedule, sheet B03.

B03 — Lower case: fasteners and locating tongue. Open full-size lossless WebP.

Upper case / black hood — H01–H03
Upper case: shell, bridges and hose slots: dimensioned black-line CAD projections and feature coordinate schedule, sheet H01.

H01 — Upper case: shell, bridges and hose slots. Open full-size lossless WebP.

Upper case: well carrier and guide details: dimensioned black-line CAD projections and feature coordinate schedule, sheet H02.

H02 — Upper case: well carrier and guide details. Open full-size lossless WebP.

Upper case: optical seat and joint section: dimensioned black-line CAD projections and feature coordinate schedule, sheet H03.

H03 — Upper case: optical seat and joint section. Open full-size lossless WebP.

Gas lid — L01–L03
Gas lid: cover and reinforcement: dimensioned black-line CAD projections and feature coordinate schedule, sheet L01.

L01 — Gas lid: cover and reinforcement. Open full-size lossless WebP.

Gas lid: chamber, spigots and deflector: dimensioned black-line CAD projections and feature coordinate schedule, sheet L02.

L02 — Gas lid: chamber, spigots and deflector. Open full-size lossless WebP.

Gas lid: locating lip and screw clearances: dimensioned black-line CAD projections and feature coordinate schedule, sheet L03.

L03 — Gas lid: locating lip and screw clearances. Open full-size lossless WebP.

Four mirrored sample wells — W01
Four mirrored sample wells: dimensioned black-line CAD projections and feature coordinate schedule, sheet W01.

W01 — Four mirrored sample wells. Open full-size lossless WebP.

Gas compression plate — P01
Removable gas compression plate: dimensioned black-line CAD projections and feature coordinate schedule, sheet P01.

P01 — Removable gas compression plate. Open full-size lossless WebP.

Assembled reader

Assembled reader with tubing and inline filters
Figure 11. Assembled reader. Printed enclosure with the lid fastened, tubing connected and inline filters fitted.

Keep cooling separate from sample gas

A fan below the PCB cools the electronics separately from the sample headspace supplied by the gas lid. Separating these paths keeps fan speed from becoming a sample-delivery variable; supplementary CAD views show the fan cradle and service openings.

Interface and firmware

Current AeroSense phone interface with sample naming, acquisition controls, history and measurement results; DEMO data
Figure 12. Local reader interface. Captured from the current software with its built-in demonstration data.

One local interface runs measurements, displays channel signals and saves named records. History reopens earlier results; JSON and CSV exports retain the acquisition settings and raw data.

  1. Connect

    Connect to the reader locally and confirm the active data source. The same interface supports device acquisition and a clearly labeled demonstration mode.

  2. Acquire

    Name the sample and choose a single measurement or scheduled acquisition. Advanced settings control block count, interval, LED duty and active LEDs. Each block records two seconds; stopping turns the LEDs off.

  3. Inspect

    Read the sample count, ADC voltage, channel signals and quality flags together. ADC diagnostics compare hardware SPI with slow GPIO sampling and test the data-line response to weak pulls.

  4. Reopen and export

    Open a selected history record, reload JSON or CSV, and export named files. Browser storage retains available raw samples for later inspection.

Reference-pattern analysis

The reader interface sends saved acquisition and calibration records to a local Python backend. For reference analysis, D1–D4 hold Or22a, Or35a, Or85b and Or98a, respectively; the backend reconstructs their ΔF/F₀ curves, averages all four over the same selected response window and returns separate ranked matches against laboratory and DoOR references. The bars show 100 × max(0, Pearson r), a pattern-similarity score rather than an odor probability or validated identification.

Manuals and feedback

Read the current English manuals and the anonymized assembly feedback below.

One participant’s assembly record identified unclear PCB fastening and nut placement. The developer manual combines this feedback with team-confirmed corrections for M2.5 PCB screws, M3 × 36 mm enclosure screws, fan clearance and the directional middle layer’s photodiode alignment.

The record documents the issues found and the resulting assembly instructions.

See the assembly photographs, annotated manual and feedback record →

User manual

Developer manual

Assembly feedback · English

Bill of materials and cost estimate

The table combines the current assembly specification with the retained materials workbook. Prices are reference purchase costs, with their pack sizes shown; they are not added into a misleading single-device total.

Table 3. Current assembly materials and recorded cost basis. All amounts in USD; shipping and tax excluded.
ComponentQuantity for the readerRecorded cost referenceSpecification and price basis
Assembled PCB · NTHU_TEST_BOARD1 boardApprox. USD 4,024 / two-board development projectOne-time PCB setup: approx. USD 2,453. Estimated average cost per board: approx. USD 613.
Printed enclosure and wells8 parts · 4 black + 4 clearUSD 35.07Recorded materials estimate for three printing batches; printer excluded.
LEE 101 receiver filter1 trimmed layerUSD 12.00 / sheet21 × 24 in sheet; the price is not for one 5.2 × 4.2 mm cut piece.
Gas tubingCut to inlet and outlet lengthsUSD 7.99 / metreReference silicone tubing, 3 mm ID × 5 mm OD; fit and VOC compatibility need checking.
Gas-line filters2 filtersUSD 95.06 / pack of 50Candidate specification: PTFE, 0.22 µm, 25 mm.
Cooling fan1 fanUSD 7.99 / fanComparable 40 × 40 × 10 mm two-wire fan; the existing fan is labeled 5 V, 0.2 A.
18650 battery1 cellUSD 6.99 / cellMolicel INR-18650-P30B retail reference.
PCB fasteners · M2.54 screws, 4 nuts, 4 suitable washersNot quotedAssembly specification. Select screw length for the actual PCB/support stack.
Fan fasteners and spacers2 screws, 2 nuts, 2 washers; matched spacersNot quoted after revisionM3. Recheck the earlier 16 mm length after adding spacers; confirm free blade rotation.
Outer-enclosure fasteners · M3 × 36 mm2 screws and 2 M3 nutsNot quotedAssembly specification; the workbook’s older 35 mm screw price is not applied.
Seals, couplers and fan power leadTo match the assembled stackNot quotedVerify seal compression, tubing fit and fan connector polarity during assembly.

PCB currency conversion: USD 1 = NTD 32.615, the midpoint of Bank of Taiwan’s USD spot buying and selling rates in the recorded conversion snapshot. Converted amounts are rounded to the nearest dollar. Exchange-rate source

Reusable build-tool references are USD 9.95 for a 3.3 V USB–UART adapter, USD 4.70 for an abrasive-paper set, USD 3.00 for polishing compound and USD 19.95 for a soldering iron. Existing computers, printers, cleaning/curing equipment, incubators and microscopes are outside this materials estimate.

The PCB project total is retained from the historical two-board development record. The setup and per-board amounts are approximate team estimates, rather than an itemized reconciliation of that total. Current M2.5 PCB fasteners, M3 × 36 mm enclosure screws and fan spacers require updated quotations.

Materials workbook and editable sources →

Printed wells

Cell viability

A living-cell cartridge needs recoverable viable cells as well as suitable optical geometry. We examined cells recovered from the medium and well bottom after the same incubation. Panel A compares viability in these paired fractions; Panel B converts bottom counts to nominal live and dead recovery density. Read the individual well points alongside the mean and spread.

Two panels after 7 hours: medium and bottom-fraction viability, and nominal live and dead bottom-recovery density. Three paired wells are shown as individual points with mean and sample SD.
Figure 13. Cell viability and recovery density. (A) Trypan blue viability in paired medium and trypsin-recovered bottom fractions. (B) Nominal bottom-recovery density: live (4.43 ± 0.43) × 10⁵ and dead (12.90 ± 0.65) × 10⁵ cells/cm², calculated with 100 µL recovery volume and a 0.112 cm² well floor. Each point is one well; bars show the mean and sample SD (n = 3).

After 7 h in the printed wells, recovered HEK293–GCaMP6f cells showed 60.0 ± 7.5% viability in the medium and 25.5 ± 0.9% in the bottom recovery fraction (mean ± SD, three paired wells).

The estimated live bottom recovery was 49,667 ± 4,805 cells per well. Recovery estimates use nominal volume and area; bottom recovery can include settled cells and does not measure attachment.

Verification

System verification

Verification follows the measured signal from frequency separation through cross-talk comparison, multi-well additivity and concentration checks. Raw records, calibration calculations and residuals link each result to its acquisition conditions.

All wells contained 100 µL. The study includes uniform standards spanning 0–20 µM, five unequal-well loadings and four isolated-well controls. All 58 acquisitions are retained, including both 2.5 µM loadings examined in quality control.

1. Separate the four frequency-tagged signals

The LEDs label D1–D4 at 41, 67, 89 and 113 Hz. Digital lock-in amplification (DLIA) extracts their in-phase and quadrature components from one photodiode waveform. A fresh illuminated-water blank is subtracted once in complex I/Q space; the corresponding single-well control sets each channel’s phase.

Measured waveform, modulation-frequency peaks, DLIA integration and blank correction.
Figure 14. DLIA separates the four modulation channels from one photodiode recording. A shows the first 150 ms of the first uniform 10 µM acquisition; the complete record contains 8,000 samples at 4 kHz. B resolves the D1–D4 carriers at 41, 67, 89 and 113 Hz. These are modulation frequencies, not optical wavelengths. C shows the extracted carrier amplitudes as the integration window grows to 2 s. D compares raw carrier magnitude with the signed response after subtracting the same-session water blank in I/Q space and applying the phase of each channel’s own single-well 10 µM control. The raw carrier includes excitation leakage and other modulated background; it is not fluorescence alone. Recomputing I/Q directly from all 30 original ADC recordings reproduces the stored values to within 6.53 × 10⁻¹⁰ V.

2. Compare simulated and measured cross-talk

A is the optical design simulation and B is the measured single-well response. Columns identify the fluorescent source well; rows identify the LED-frequency/readout channel. C pairs the same 16 matrix positions, with A on the x-axis and B on the y-axis.

A and B compare simulated and measured responses. C plots their 16 corresponding matrix entries with Pearson r 0.810.
Figure 15. A/B show row-normalized responses. A is the floor-layer model used in the optics section and Model. B averages three technical reads after complex blank subtraction and own-well phase projection: one well contains 100 µL of 10 µM fluorescein and the other three each contain 100 µL DDW. C uses all 16 matching A/B entries, with Pearson r = 0.810; the dashed line is B = A. These are dependent matrix entries, so the correlation describes the overall pattern, not independent-sample significance or concentration accuracy.

The overall Pearson correlation is r = 0.810 across all 16 corresponding matrix entries; the 12 off-diagonal cross-talk entries give r = 0.257.

D4 has the largest own-well share in both matrices: 90.9% simulated and 69.5% measured. Its strongest non-own source is D3 in both: 8.6% and 30.6%. This means source D3 contributes to readout D4. The reverse direction is different: source D4 contributes 8.5% to readout D3 in the simulation but only 0.3% in the measurement.

These comparisons describe response patterns. Section 3 checks a separate requirement for cross-talk correction: whether four isolated-well responses can be added to predict a simultaneous four-well measurement.

3. Multi-well signal additivity

The previous figure shows which source wells affect each readout channel. To subtract those contributions, we first need to check whether the signals measured one well at a time can predict the signal when all four wells contain fluorescein.

We added the four separate single-well measurements to form a prediction. We then measured all four wells together, each at 10 µM. In the chart, each channel’s prediction is set to 100%; the bar is the actual simultaneous measurement. If the signals add up as assumed, the bar should reach the dashed line.

All-well measured response reaches 68.4, 69.4, 48.3 and 65.6 percent of the summed single-well prediction for D1 through D4; the dashed line marks the 100 percent prediction.
Figure 16. Summed single-well predictions and simultaneous four-well measurements. Each bar is 100 × (measured all-well response) ÷ (sum of the four isolated-well responses), after blank subtraction and the same channel phase projection. Every well contains 100 µL. For each isolated control, one well contains 10 µM fluorescein and the other three contain DDW; for simultaneous loading, all four contain 10 µM fluorescein. Each condition averages three technical reads. These percentages compare signals, not concentration accuracy.

All four measured responses fall below the sum of the isolated-well predictions; D3 reaches only 48.3%. This comparison tests the additivity required to use isolated-well responses for simultaneous loading.

4. Check calibrated concentrations against prepared standards

The figure shows per-channel curve calibration after blank subtraction and phase projection, using the same color and symbol for each channel. Unequal-well loadings were excluded from calibration; uniform-concentration checks exclude all readings at the tested concentration. This analysis does not explicitly invert the cross-well mixing matrix.

Curve-calibrated concentrations in A, and residual violins grouped by prepared concentration in B, with all 32 points and labeled group means.
Figure 17. Per-channel concentration checks. A compares predicted with prepared concentration; the dashed line marks equality. B groups prediction-minus-prepared residuals by prepared concentration, with equal spacing between categories. Each violin shows all original channel-colored points and the group mean. The 0, 0.1, 1, 5 and 10 µM groups contain 5, 5, 9, 4 and 9 means. The 32 channel/loading means come from 22 acquisitions, not independent preparations. Descriptive R² = 0.886 and RMSE = 1.415 µM are unchanged and describe curve calibration, not matrix unmixing.

The checks compare calibrated estimates with prepared concentrations across loading patterns. Deviations remain at low concentrations and in unequal-well loadings.

Matrix-unmixing estimates and residuals

Biology · Hardware · Measured data

Living-cell integration

We followed four cell-containing wells before and after adding VUAA1 (1:100 dilution).

D1 contained Or35a + ORCO, D2 Or85b + ORCO, D3 Or98a + ORCO, and D4 ORCO only.

Four hardware channel time courses. Cells-plus-medium and VUAA1 series each start at one, using their own first amplitude as the denominator. Each panel annotates the difference between the two late-window changes in percentage points.
Figure 18. Each ten-point trace is normalized to its first reading on the nominal 0–1 min schedule; the first VUAA1 record uses the supplied replacement CSV. Late-window contrast is the VUAA1 minus cells-plus-medium change in the last-three-point mean ratio, expressed in percentage points.

What changes during the readout?

D2 and D3 increase relative to their first post-addition readings, D4 rises modestly after a brief dip, and D1 decreases; the cells-plus-medium series change only slightly. D3’s large ratio depends on its low first post-addition reference.

The final absolute signals remain 7.5%, 6.1%, 5.6% and 2.7% below the initial three-reading means for D1–D4, respectively.

Comparison with wet-lab kinetics

We compared the VUAA1 trajectories with the corresponding Test 2 wells, using each acquisition’s first reading as its own reference. D2 shows the clearest shared rise, while D1 changes in the opposite direction.

Four matched hardware and wet-lab VUAA1 time courses, each normalized to its own first reading, with five-point Pearson and Spearman correlations.
Figure 19. Ten hardware points (0–54 s) and six Test 2 points (0–60 s) are shown as F/Fstart on nominal sampling times. Pearson r and Spearman ρ use only the five native points at 0, 12, 24, 36 and 48 s; the hardware trace includes the supplied first-record replacement.

For D1–D4, the matched Pearson correlations are −0.859, +0.972, +0.961 and +0.948; Spearman correlations are −0.80, +0.90, +0.90 and +0.90. The corresponding first-difference correlations are −0.529, +0.650, +0.761 and −0.125. D2 and D3 share positive level and step-to-step trends with their matched wet-lab traces; D1 trends in the opposite direction, while D4 shares less step-to-step variation than its level correlation suggests.

Four-by-four Pearson and Spearman maps compare the same five signal-level observations, with intended construct matches outlined.
Figure 20. All 16 hardware–wet-lab pairings at 0, 12, 24, 36 and 48 s are shown. Left: Pearson r; right: Spearman ρ of the same five signal levels, each divided by its own first reading. Outlines mark intended construct matches; strong off-diagonal correlations show that similar trends are not pairing-specific. The five serial observations are not biological replicates.

Discussion

The assembled reader combines replaceable cell wells, a shared gas chamber and frequency-tagged optical readout in a serviceable enclosure. Fixed optical registration supports repeated calibration, while separate cooling and sample-gas paths avoid using fan operation to set sample delivery. The tested acquisition workflow and revised assembly manual provide a practical starting point for replication.

The reader separates four modulation channels, but its isolated-well calibration does not directly transfer to simultaneous loading: uniform 10 µM responses are only 48.3–69.4% of the predicted sum. Likewise, strong same-well responses drive much of the simulation–measurement correlation (r = 0.810); the 12 off-diagonal entries give r = 0.257. The floor-layer simulation and liquid-volume standard have different source conditions, and the measured signal includes electronics. Arbitrary-mixture recovery therefore needs matched single-source and mixed-source controls across concentrations.

Viable cells were recovered from printed wells, but the lower viability in the bottom fraction cannot be assigned to resin, culture conditions or recovery without matched controls. Resin choice and post-processing can affect compatibility [21]. Machined PMMA and poly-D-lysine are candidates supported by separate HEK293 studies [22][23]; comparison with standard culture plastic should measure viability, retained cells, optical background and reporter response under matched conditions.

Airflow calculations describe gas delivery beside fixed liquid surfaces, not cell forces or uniform cellular exposure. Nominal CAD dimensions still require fit and alignment checks. An independent timed assembly and operation trial should test the revised manual, recording assistance and errors.

The single cell run does not separate receptor responses from liquid addition or optical mixing; the fluorescein calibration was not applied to it. Its replacement first VUAA1 record matches ADC values in an earlier file described as synthetic, while the archived firmware export differs. The supplied remeasurement description does not resolve that conflict, so these traces remain descriptive evidence of the supplied records.

Cross-platform similarity is not specific to matched constructs: D2 gives r = 0.9719 with Or85b and 0.9671 with Or98a, and D3 exceeds its strongest off-diagonal match by only 0.0055. D2 and D3 remain positive after omitting the first point (r = 0.970 and 0.904), whereas restoring original record 19 changes D4 from +0.948 to −0.290. The original series spans 69.186 s, while the replacement lacks a verified timestamp. Control subtraction preserves some shared trends, but preceding hardware-medium readings and separate wet-lab DMSO wells are different references. Serial time points do not establish synchronized kinetics, biological replication or receptor-specific activation.

Conclusion

AeroSense integrates removable wells, filtered optical collection, shared-photodiode electronics and a local phone interface. Fluorescein measurements demonstrate four-frequency separation and quantify source mixing. The measured I/Q matrix, concentration checks and cell-containing-well traces provide a reproducible basis for selecting the next calibration and biological-control experiments.

Data availability

Research archive: Zenodo · DOI: 10.5281/zenodo.23469212. Public release pending.

The fluorescein measurement archive and current analysis are downloadable below. The official repository provides code, documents and the linked numerical tables; the separate archive note records the publication status of the complete research package.

The archived raw-data ZIP preserves all 58 acquisitions from the initial study, expanded study and additional 2.5 µM loading, including records excluded from the primary fit. Each original file is checked against its SHA-256 manifest. These files are also preserved in the complete research archive prepared locally for Zenodo; that full archive has no verified public deposit.

Build resources

Electronics

CAD and print files

Original build photographs · 10 WebP files
Printed parts · eight operating STLs
Table 5. Current print inventory — one of each part
PartMaterial / functionPrint file
01 · Integral baseRigid PC/GF-like BlackSupport PCB, fan, captive nuts and device. AssemblyZ0–43.3.STL file — archive availability
02 · Black hood with integral well carrierRigid PC/GF-like BlackEnclose well sides, register wells and support filter. AssemblyZ40.8–64; carrierZ44.2–55.5; lower optical band open.STL file — archive availability
03 · Gas-path lidRigid PC/GF-like BlackOpen for well replacement, deflect inlet and form shared gas path. AssemblyZ57.9–66; inlet-X/outlet+X, axisZ60.8.STL file — archive availability
04 · Clear well D3Aqua Clear PlusIndividually removable liquid vessel; gas phase shared. Center(3.3, 2.65); outer floorZ42.5, inner floorZ43, topZ55.5; flange outward.STL file — archive availability
05 · Clear well D4Aqua Clear PlusIndividually removable liquid vessel; gas phase shared. Center(-3.3, 2.65); outer floorZ42.5, inner floorZ43, topZ55.5; flange outward.STL file — archive availability
06 · Clear well D1Aqua Clear PlusIndividually removable liquid vessel; gas phase shared. Center(-3.3, -2.65); outer floorZ42.5, inner floorZ43, topZ55.5; flange outward.STL file — archive availability
07 · Clear well D2Aqua Clear PlusIndividually removable liquid vessel; gas phase shared. Center(3.3, -2.65); outer floorZ42.5, inner floorZ43, topZ55.5; flange outward.STL file — archive availability
08 · Four-hole compression plateRigid PC/GF-like BlackDistribute cap load onto four well-mouth gaskets. Above lower four-hole gasket/below upper ring, assemblyZ56.1–57.3.STL file — archive availability

Firmware and interface

Model data and scripts

Cell viability data

System verification

Preview the focused validation protocol

Electrical, optical and acquisition checks with settings and raw-data records.

Read fluorescence calibration and cell-test instructions

The guide separates dye calibration from live-cell testing and specifies acquisition settings, controls and analysis records.

Verification protocol

The protocol organizes electrical, optical, transport and workflow verification. Current records include both fluorescein studies, single-well controls, concentration-level holdouts and an assembly-feedback record.

Table 6. Ten checks defined by the verification protocol
GateSpecified recordCurrent evidence and next measurement
1 · Electrical bring-upRails, power sequencing, board mapping and safe ADC range58 complete fluorescence acquisitions are archived; the expanded ADC records are varying and contain no rail samples.
2 · Dark noiseLED-off recordings with acquisition settingsRecord an LED-off baseline alongside the illuminated-water references.
3 · Excitation leakageCell-free illuminated blanks; spectral leakageIlluminated-water references recorded; next separate the spectral contributions to the background.
4 · Standard calibrationKnown fluorescein concentration response and repeatabilityA shared exponential family covers D1–D4; concentration-level holdout RMSE is 1.28 µM for the uniform-standard calibration.
5 · FDM recoveryKnown single-well conditions recovered at each modulation frequencyThe four LED frequencies are resolved; all 12 single-well technical holdouts identify the filled well.
6 · CrosstalkMeasured four-by-four transfer matrixA measured single-well I/Q matrix is available; uniform-loading transfer and regularization are assessed separately from source identification.
7 · DLIA improvementMatched raw / demodulated SNR under controlled interferenceCompare raw and demodulated SNR under controlled interference.
8 · Biological fluorescenceControlled cell response and a same-sample reference measurementExtend the dye-tested readout to controlled cells and a same-sample reference measurement.
9 · Battery stabilityAcquisition stability over a documented discharge / power conditionMeasure acquisition stability across a documented battery-discharge series.
10 · User workflowTask completion, elapsed time and user observationsOne assembly-feedback record informed the manual; evaluate the instructions in an independent timed trial.

Next calibration sequence

The future developer pilot begins with electrical-rail and ADC-headroom checks, an LED-off baseline (DARK), stop-control checks and observed D1–D4 mapping. Its sodium-fluorescein sequence in PBS is: illuminated PBS calibration blank → independent PBS check blank → four single-well 100 nM conditions (PBS in the other wells) → all wells at 50 nM → all wells at 100 nM → repeat all wells at 50 nM. This sequence contains nine illuminated analytical conditions, each with at least five distinct valid two-second technical blocks; the LED-off baseline is saved separately.

This 50/100 nM PBS pilot is a separate future developer study from the water-based fluorescein calibration reported above. Its approximately five-second record spacing calls for a separate timing assessment before recording fast GCaMP6f cell responses.

Optical output

Concentration response, a 4 × 4 single-well matrix and held-out blank before/after correction. The analysis rejects DEMO, duplicate or incompatible records.

Independent assembly

The manual-evaluation specification covers 2–3 non-designers and records completion, time, assistance, orientation and seal errors.

User testing and performance records

Blank feedback form — PDF reader

The downloadable performance ledger is a blank recording template for electronics, LED mapping, standards, fit, reassembly, water/flow and manual feedback.

Table 7. Prepared measurement-record templates
RecordWhat it captures
Electronics · LED mappingElectrical checks, actual channel identity and acquisition conditions
Optical standardsIndependent blanks, single-well standards, correction and repeat readings
Fit checks · Reassembly · Water and flowMechanical compatibility, repeated assembly and liquid/gas-path checks
Manual tasks · RatingsAssembly tasks, completion time and participant feedback

Source library

Manuals and assembly feedback

Calibration data and reproducible analyses

Instrument calibration source — GitLab

Matrix calculation and solver checks
Signal comparison data · CSV

Cross-talk calibration analysis source — GitLab
Full complex mixing matrix · CSV
All matrix-derived estimates · CSV
Residual means by source well · CSV
Solver comparison · CSV
Curve checks after matrix inversion · CSV
Full method, assumptions and data provenance

Per-channel calibration analysis source — GitLab
All 32 plotted means · CSV
All 88 individual well/read estimates · CSV
Calibration functions and coefficients · CSV
Three shared standards · CSV
Shared-shape model comparison · CSV
Complete method and validation limits

Historical residual-correction comparison · CSV

Residual means by prepared concentration · CSV

Canonical A/B matrices · CSV
Corresponding A/B entries · CSV
Overall and cross-talk-only correlations · CSV
Sources and reproducible calculation

Supplementary data

Cross-platform source and timing sensitivity

Pearson r is shown for four matched pairs. Selected, original-first and control-adjusted comparisons use five native nominal points; first-point omission uses four points, and the original-device-clock comparison interpolates hardware to six wet-lab times from 0 to 60 s.

Pearson r
Channel and constructSelected first pointOriginal first pointFirst point omittedOriginal device clockControl-adjusted
D1 · Or35a + ORCO-0.859-0.859-0.218-0.796-0.852
D2 · Or85b + ORCO+0.972+0.972+0.970+0.981+0.975
D3 · Or98a + ORCO+0.961+0.961+0.904+0.957+0.960
D4 · ORCO only+0.948-0.290+0.941-0.182+0.966
Measured matrix correction and its limits

For z = 1000(I + iQ) mV, subtract the same-set DDW mean. Each source column is the corresponding isolated 10 µM net response divided by 10; stack its four I rows and four Q rows into M. Solve ĉ = (MᵀM + λσmax²I)⁻¹Mᵀy with λ = 0.0001, selected from 15 single-source/blank technical leave-one-acquisition-out folds. Matrix condition number: 10.613. No uniform target concentration or mixed-loading truth is used to tune λ.

The measured all-well 10 µM responses are only 68.4%, 69.4%, 48.3% and 65.6% of the summed isolated-source predictions. A fixed additive source matrix must satisfy that sum at its own calibration concentration regardless of the later scalar curve family. Nonnegative constraints reduce technical-control error but set failed components to the boundary; they do not demonstrate correct concentration recovery.

Set A has no independent four-source matrix. Its two unequal concentration patterns and uniform standards span only three source-design directions, so an unrestricted 4-source matrix is not identifiable there. A per-channel complex affine map is fitted to the three shared standard means (0, 1, 10 µM), then the set-B matrix is applied without tuning against mixed-loading truth. Set-B uniform 1/5/10 µM loadings are absent from matrix training; isolated 10 µM controls remain, so these are loading holdouts rather than exclusion of every 10 µM record. Technical repeats, channels and violin density are not independent biological replicates.

Each isolated-source control contains 100 µL of 10 µM fluorescein in one well and 100 µL DDW in each of the other three. All-well standards also contain 100 µL per well. The observed discrepancy therefore occurs with neighboring wells filled in both configurations; these measurements do not isolate its physical cause.

Per-channel curves and quality checks

Figure 17 fits each readout separately without explicitly inverting a cross-well matrix. Its descriptive R² of 0.886 describes the per-channel calibration display. The matrix-derived concentration estimates are reported separately below.

The preceding display used separate calibrations and had not aligned all three shared standards. This revision reconstructs the raw I/Q records, applies the same current-blank and signed-phase processing to both calibration sets, and compares the candidate functions using whole-concentration holdouts.

DDW, 10C and 100C correspond to 0, 1 and 10 µM. All three shared levels were audited. After blank subtraction, the 1-to-10 µM response ratio differs between calibration sets: for D4 it is 0.86% versus 16.18%, so one gain factor cannot align both. A flexible curve can be made to pass through three points, but those same points cannot independently validate it. The main result uses fresh within-set blank correction and the selected inverse curve to obtain common concentration units; no common three-point signal warp is applied.

A joint exponential shape with a separate offset/gain for each calibration set gives an interpolation RMSE of 1.036 µM, but only 19/20 inverses remain in range. The retained functions give 1.043 µM and 20/20 in-range estimates. That small RMSE difference is not a demonstrated statistical advantage. Entire concentrations are removed from both sets and transfer/phase fitting in the joint checks; endpoint extrapolation is reported separately.

Use z = 1000(I + iQ) mV, subtract the current calibration set's DDW mean, then take y = Re[(z − blank) conjugate(phase)]. Set A uses its highest retained training standard as the phase reference; set B uses separate isolated-source controls. The linear form is y = a + b(c/10). The exponential form is y = a + b expm1(kc/20)/expm1(k). All channel coefficients, raw hashes, folds and inverse-validity flags are downloadable.

The original 2.5 µM loading is unstable in D3; the additional loading predates a matching blank. Both remain in the archive and sensitivity analyses. Calibration anchors are not presented as new holdout predictions. Selection and QC are exploratory, channels in one loading are dependent, and signed phase projection does not establish arbitrary-mixture interference removal.

The 0–10 µM set contains 12 independently prepared uniform standards and five withheld unequal-well loadings. The 0–20 µM set supplies three tested interior levels, each averaged over three technical reads. Main-figure means therefore combine 20 withheld mixed-loading estimates with 12 concentration-holdout estimates. Their combined R² and RMSE describe this display, not one independent accuracy specification. Low-concentration evidence remains limited to two held standard levels in the first calibration set.

A historical sensitivity check compared extra residual corrections after the exponential calibration: nested whole-level RMSE was 1.28 µM without an added correction, 1.34 µM with a constant offset and 19.57 µM with a linear adjustment. That extra adjustment is not part of the retained method; the original diagnostic is preserved.

Test interference removal before trusting it

The isolated 10 µM controls supply a measured I/Q response matrix. Leaving out one read from every source and the blank identifies the active well in 12/12 technical checks, with concentration RMSE 0.51 µM. Applying that same matrix to simultaneous equal loading produces a negative D3 estimate; ridge regularization chosen from source-only holdouts does not remove the mismatch. The matrix therefore supports the single-well check, while uniform concentration conversion uses the measured uniform-standard curves.

Measured row-normalized response matrix and direct matrix estimates for uniform standards.
Supplementary figure 1. Interference-transfer check. A: the measured single-well matrix, using the same signed row normalization as Figure 15. B: direct matrix estimates for uniform standards, retaining negative estimates. At actual 10 µM the estimates are 9.09, 8.72, −2.37 and 10.51 µM. This layout mismatch is why the matrix-before-curve route is evaluated but not adopted for arbitrary unknown mixtures.

Compare both 2.5 µM loadings

The additional loading is more repeatable: D1–D4 raw-amplitude CVs are 0.20%, 0.18%, 0.29% and 0.42%. The original D3 readings jump by about 81% between the first and second read. However, the additional loading was acquired earlier and its D3/D4 signals fall below the later water blank, so it lacks a matching baseline. Both loadings remain in the archive and sensitivity analyses; neither enters the primary fit.

Technical repeatability and background comparison of both 2.5 micromolar loadings.
Supplementary figure 2. Original and additional 2.5 µM loadings, mean ± technical SD of three reads. A: raw carrier amplitude. B: signed response after subtraction of the expanded-session water blank. The additional loading is stable within its three reads but differs in baseline; stability alone does not make it interchangeable with the original loading. All-original and replacement analyses are included in the downloadable results.

Complete raw data, calculations and reproducible scripts

Concentration estimates after matrix unmixing

These diagnostic estimates explicitly apply the measured I/Q mixing matrix. They are 10 µM-reference-equivalent values, not the per-channel curve estimates in Figure 17. The full signed results are retained here.

After actual cross-talk correction, integrated matrix-equivalent estimates and well residual violins with eight points and a mean per well.
Supplementary figure 3. After matrix unmixing. A: matrix-derived estimates versus prepared concentration, with descriptive R² = −0.276 and RMSE = 4.738 µM. B: residual = estimate − prepared concentration; the violin shows a smoothed distribution of eight means per well, with every point and its well mean retained. Ten negative estimates remain visible. The 32 means come from 22 acquisitions, not independent preparations. For the calibration set lacking isolated-source controls, a complex affine transfer uses DDW, 10C and 100C before the measured matrix is applied; unchanged matrix shape across sets remains an unverified assumption.

Mean residuals are +0.03, −0.84, −4.41 and −0.02 µM for D1–D4. A mean near zero does not remove the wide spread. The negative D3 estimates and poor combined fit show that this fixed matrix has not achieved reliable concentration recovery for these simultaneous loadings.

Matrix solvers and concentration-curve checks

We now explicitly remove the measured cross-well contribution before interpreting concentration. Four isolated 10 µM controls define an 8 × 4 matrix from the I and Q responses; the other three wells contain 100 µL DDW in each control. After subtracting the current water blank, a signed ridge solution estimates the four source contributions. Its strength is chosen using only single-source and blank technical holdouts, without fitting the tested mixed or uniform loadings.

Measured cross-talk correction: technical holdout solver errors and disagreement between summed isolated-source controls and all-well measurements.
Supplementary figure 4. Actual matrix correction checks. A: 15 acquisition-level technical holdouts, including blank controls, give component RMSEs of 0.331 µM (unregularized), 0.328 µM (signed ridge, relative penalty 0.0001) and 0.212 µM (nonnegative LS). Ridge is selected among signed solvers to retain negative values as model diagnostics; nonnegative LS is a separately evaluated constrained optimization. B: adding the four isolated 10 µM signals overpredicts the all-well 10 µM measurement. The fixed matrix distinguishes isolated controls but has not transferred reliably to simultaneous loading.

Simply adding a scalar curve after this fixed-matrix inversion does not repair its transfer. The inferred D3 source component decreases across part of the uniform series. None of the tested post-matrix linear, quadratic, exponential, power or PCHIP families provides increasing, valid in-range inverses for all four components. Signed matrix estimates are therefore retained as diagnostics. This does not rule out other nonlinear forward models, which these data have not validated.

Optical comparison: corresponding matrix entries

Hardware and Model use the same floor-layer simulation, displayed as row-normalized response percentages. The canonical CSV preserves the raw simulation in nW and the measured signed response in mV as well as the displayed percentages.

For C, pair Aᵢⱼ with Bᵢⱼ at the same readout i and source j, using full-precision values before display rounding. Pearson r is 0.8097999981 across all 16 entries and 0.2570218322 across the 12 off-diagonal entries. Signed negative responses are retained. The shared strong diagonal and the dependence between cells must be considered when interpreting the overall correlation; no independent-sample significance test is claimed.

The four readouts are LED-frequency channels of one photodiode, not four physical detectors. Source-column patterns include excitation and electronic-sensitivity differences. Neither matrix direction directly measures the fraction of all photons leaving a well.

Optical specifications

Match the spectra

The optical chain needs both excitation and separation: blue light excites GCaMP6f, while the receiver filter passes green fluorescence to a photodiode that responds to both colors. The spectra connect the biological reporter to the filter and detector choices. [14]

Heterologous insect-receptor assays and GCaMP6 calcium indicators provide the biological basis for the specified optical readout. [13] [14]

Filter excitation leakage

Alignment and calibration

The collection aperture fixes the filter and detector position. Separate well calibrations account for unequal collection and optical mixing; the reader retains raw complex I/Q, illuminated blanks and source-well reference records.

The calculation uses published LEE transmission, while spatial photometry supports fixed registration. The full assumptions and comparisons are documented in Model. [12] [11]

Geometry, illumination maps and well mixing → Spectra, alternative filters and noise analysis →

The filter calculation uses fixed Monte Carlo mean powers and published transmission. Geometry and material uncertainty are not propagated. The synthetic current trace omits analog transfer and ADC clipping; the plotted ratios describe optical rejection and retention.

Readout electronics and controls.

Table 4. Well and optical assembly dimensions
Optical partNominal CAD dimensionsRole in the light path
Clear wells · Aqua Clear Plus4 × 2.8 mm interior; 5 × 3.8 mm exterior; 0.5 mm flat floorHold four separate samples and provide a transparent collection surface.
Black hood and carrier · PC/GF-like Black20 × 18 mm carrier; 5.3 × 4.1 mm slots; minimum shared wall 1.2 mmFix well spacing and shield the upper sidewalls, while keeping the lower optical band open.
Filter seat and detector sealSeat up to 5.2 × 4.2 × 1.0 mm; seal aperture 3 × 3.4 mmPlace the receiver filter over the photodiode and close paths around its edges.
RegistrationWell pitch 6.6 × 5.3 mm; photodiode active-area offset (0, −0.4) mmPreserve the well, LED and detector relationship during assembly.

The dimensions and material names are the current build specifications.

Component sources and pin mapping

Circuit components

The power path uses TPS61023 boost conversion [2] and BQ24074 charging [3]. ESP32-WROOM-32E coordinates acquisition [4]. LP5907 and LTC6655 provide regulation and a 2.5 V reference [5] [6]; an LMP7721 TIA and ADS8866 16-bit SAR ADC form the analog readout [7] [8]. Four Würth blue LEDs and a VEMD5060X01 photodiode provide the optical path [9] [10].

Firmware assigns LED gates to GPIO 25/26/27/32 and ADC CONVST/SCLK/DOUT to 33/18/19. Its 4 kHz timebase specifies 8,000 samples per two-second I/Q block at 41/67/89/113 Hz. The firmware configuration retains BOARD_VALIDATED=false.

Mechanical drawings and assembly checks
Technical drawings · exploded assembly, fan and ports
Current base with integral fan seat and connector access
Figure 21. The base integrates the fan cradle and service openings. The PCB is hidden to expose the existing cradle and screw locations; no additional fan bracket is introduced. Check populated component heights, connector bodies, solder and wiring during assembly.
Current PCB orientation and service-port plan
Figure 22. Service access and board orientation from the CAD design. J2 USB-C supplies power only; J3 exposes 3.3 V UART, with manual reset and boot controls. Check the actual board against mirrored bottom-view drawings.

Assembly checkpoints

The assembly specification protects the electronics and keeps the well vessels replaceable.

  1. With power disconnected, inspect all eight prints and measure PCB, mounting holes, component heights and antenna insertion clearance.
  2. Mount the PCB with M2.5 screws, nuts and suitable washers. Mount the fan with M3 fasteners and matched spacers, checking blade clearance; never force a tight part into place with screws.
  3. Install the filter and optical seals, locate the four wells by their flanges, then fit the well seals and removable compression plate.
  4. Connect gas tubes to the gas lid and route them through the side openings; close the lid without displacing or pinching seals.
  5. Perform an off-board cell-free liquid leak check, then verify electrical rails, raw ADC headroom, stop control and actual LED orientation before acquisition.

The CAD includes an antenna aperture with about 0.66 mm nominal insertion clearance against the supplied drawing.

Interface records and firmware details
Current measurement results and channel signals, using demonstration dataSelected demonstration history record with JSON and CSV export controls

Signal correction

The correction routine subtracts an independent illuminated blank in complex I/Q space and uses a separate fluorescence phase reference. Exports preserve raw I/Q, unsmoothed corrected values, signed projection, calibration identifiers and quality flags. This path is verified by software tests; a living-cell baseline remains a separate quantity.

Firmware and interface

J2 supplies USB-C power. A 3.3 V USB–UART adapter connects through J3; hold SW3 while resetting with SW2 to enter upload mode. The ESP32 firmware and LittleFS interface form one software installation.

References

Research papers
  1. Scofield, J. H. (1994). Frequency-domain description of a lock-in amplifier. American Journal of Physics, 62(2), 129–133. DOI: 10.1119/1.17629. Author-hosted paper.
    Source highlights
    • Classic frequency-domain description of lock-in amplification.
    • Foundational context for digital lock-in recovery of weak periodic optical signals.
  2. Texas Instruments. TPS61023 3.7-A Boost Converter With 0.5-V Ultra-Low Input Voltage. Manufacturer datasheet, revision B. Original TPS61023 datasheet.
    Source highlights
    • Texas Instruments. TPS61023 3.7-A Boost Converter With 0.5-V Ultra-Low Input Voltage. Manufacturer datasheet, revision B. Original TPS61023 datasheet. Back to citation
  3. Texas Instruments. BQ2407x Standalone 1-Cell 1.5-A Linear Battery Chargers with Power Path. Manufacturer datasheet, revision N. Original BQ24074 datasheet.
    Source highlights
    • Texas Instruments. BQ2407x Standalone 1-Cell 1.5-A Linear Battery Chargers with Power Path. Manufacturer datasheet, revision N. Original BQ24074 datasheet. Back to citation
  4. Espressif Systems. ESP32-WROOM-32E & ESP32-WROOM-32UE Datasheet. Manufacturer datasheet, version 2.1. Original ESP32 module datasheet.
    Source highlights
    • Espressif Systems. ESP32-WROOM-32E & ESP32-WROOM-32UE Datasheet. Manufacturer datasheet, version 2.1. Original ESP32 module datasheet. Back to citation
  5. Texas Instruments. LP5907 250-mA, Low-Noise, Low-IQ LDO. Manufacturer datasheet, revision Q. Original LP5907 datasheet.
    Source highlights
    • Texas Instruments. LP5907 250-mA, Low-Noise, Low-IQ LDO. Manufacturer datasheet, revision Q. Original LP5907 datasheet. Back to citation
  6. Analog Devices. LTC6655/LTC6655LN 0.25ppm Noise, Low Drift Precision References. Manufacturer datasheet, revision J. Original LTC6655 datasheet.
    Source highlights
    • Analog Devices. LTC6655/LTC6655LN 0.25ppm Noise, Low Drift Precision References. Manufacturer datasheet, revision J. Original LTC6655 datasheet. Back to citation
  7. Texas Instruments. LMP7721 3-Femtoampere Input Bias Current Precision Amplifier. Manufacturer datasheet, revision E. Original LMP7721 datasheet.
    Source highlights
    • Texas Instruments. LMP7721 3-Femtoampere Input Bias Current Precision Amplifier. Manufacturer datasheet, revision E. Original LMP7721 datasheet. Back to citation
  8. Texas Instruments. ADS8866 16-Bit, 100-kSPS, Serial Interface, microPower, Miniature, SAR ADC. Manufacturer datasheet, revision C. Original ADS8866 datasheet.
    Source highlights
    • Texas Instruments. ADS8866 16-Bit, 100-kSPS, Serial Interface, microPower, Miniature, SAR ADC. Manufacturer datasheet, revision C. Original ADS8866 datasheet. Back to citation
  9. Würth Elektronik. WL-SMSW SMT Mono-color Sideview Waterclear: 155124BS73200. Manufacturer datasheet, revision 001.011. Original excitation LED datasheet.
    Source highlights
    • Würth Elektronik. WL-SMSW SMT Mono-color Sideview Waterclear: 155124BS73200. Manufacturer datasheet, revision 001.011. Original excitation LED datasheet. Back to citation
  10. Vishay Semiconductors. VEMD5060X01 Silicon PIN Photodiode. Manufacturer datasheet, revision 1.2. Original VEMD5060X01 datasheet.
    Source highlights
    • Vishay Semiconductors. VEMD5060X01 Silicon PIN Photodiode. Manufacturer datasheet, revision 1.2. Original VEMD5060X01 datasheet. Back to citation
  11. Burton, A. et al. (2020). Wireless, battery-free subdermally implantable photometry systems for chronic recording of neural dynamics. PNAS, 117(6), 2835–2845. DOI: 10.1073/pnas.1920073117.
    Source highlights
    • Published photometry-system engineering used as design context; its device performance is not an AeroSense result.
  12. LEE Filters. 101 Yellow. Source ↗
    Source highlights

    Manufacturer transmission information for the sampled LEE101 curve. Layer stacking, source spectra and complete-reader performance are separate modeling assumptions.

  13. Zboray, K. et al. (2023). High-throughput ligand profile characterization in novel cell lines expressing seven heterologous insect olfactory receptors for the detection of volatile plant biomarkers. Scientific Reports, 13, 21757. DOI: 10.1038/s41598-023-47455-4 (opens in a new tab).
    Source highlights
    • Heterologous insect olfactory receptors expressed in mammalian cell lines for odorant / volatile profiling.
    • Demonstrates a living-cell sensor route for plant-related VOC biomarker detection (literature context, not AeroSense data).
  14. Chen, T.-W. et al. (2013). Ultrasensitive fluorescent proteins for imaging neuronal activity. Nature, 499, 295–300. DOI: 10.1038/nature12354 (opens in a new tab).
    Source highlights
    • Introduces ultrasensitive GCaMP6 fluorescent calcium indicators for imaging activity.
    • Supports calcium / fluorescence reporting as an optical readout modality (contextual literature).
  15. AeroSense · source library — Code and documents — GitLab · Hardware analysis and device source — GitLab.
    Source highlights
    • AeroSense · hardware source library.
  16. AeroSense · simulation report (PDF).
    Source highlights
    • AeroSense · simulation report.
  17. Alpha & Omega Semiconductor. AO3400A: 30 V N-Channel MOSFET. Manufacturer datasheet, revision 3.1, July 2023, p. 2. Original AO3400A datasheet.
  18. E-CMOS. EC50117: 1 A Low Dropout Positive Voltage Regulator. Manufacturer datasheet, revision F004, pp. 2–3, 5–6. Original EC50117 datasheet.
  19. Espressif Systems. ESP32 Hardware Design Guidelines: Schematic Checklist. Power supply; chip power-up and reset timing. Original ESP32 hardware guidelines.
  20. Dana, H. et al. (2019). High-performance calcium sensors for imaging activity in neuronal populations and microcompartments. Nature Methods, 16, 649–657. DOI: 10.1038/s41592-019-0435-6 (opens in a new tab).
    Source highlights

    Supplementary Figure 1 reports excitation and emission spectra for GCaMP6 and jGCaMP7 sensors, with one measurement per sensor. The calcium-bound GCaMP6f traces supply the chart-derived spectra shown here.

    Supplementary information ↗

  21. Kreß, S., Schaller-Ammann, R., Feiel, J., Priedl, J., Kasper, C., & Egger, D. (2020). 3D Printing of Cell Culture Devices: Assessment and Prevention of the Cytotoxicity of Photopolymers for Stereolithography. Materials, 13(13), 3011. DOI: 10.3390/ma13133011.
  22. Obuliraj, S. et al. (2014). Surface characteristics and electrical properties of PMMA chips for incubation-type planar-patch-clamp biosensors. Colloids and Surfaces B: Biointerfaces, 116, 193–200. DOI: 10.1016/j.colsurfb.2013.12.055.
  23. Faussner, A., Deininger, M. M., Weber, C., & Steffens, S. (2022). Direct addition of poly-lysine or poly-ethylenimine to the medium: A simple alternative to plate pre-coating. PLOS ONE, 17(7), e0260173. DOI: 10.1371/journal.pone.0260173.
  24. COMSOL. About the Fluid Interface Boundary Conditions. COMSOL Multiphysics 6.3 documentation. Official documentation.
    Source highlights
    • The fluid–fluid interface formulation solves both phases and couples their velocity and interfacial force balance.
Datasheets Other sources