Abstract
This page presents three laboratory workflows: fly-brain imaging, bacterial cloning and plasmid preparation, and HEK293 expression and odor-response testing. Each track documents its preparation, acquisition and protocol steps.
Six stages, three tracks
The core wet-lab record follows colony plates → DNA electrophoresis → transfection efficiency assessment → odor testing → G418 selection → post-selection odor testing. The fly-brain study remains an independent supporting track. Follow the same six stages in Results.
Fly-brain imaging provides biological context in a separate preparation. Cloning supplies the DNA. HEK293 assays assess the engineered sensor.
Read a track
Track A
Fly brain imaging
A01
Light-sheet imaging
Measure regional fluorescence during timed odor presentation.
TH-GFP records membrane-labeled neuronal fluorescence. GH146 × UAS-GCaMP6f records calcium-sensitive fluorescence in the GH146 expression domain. The GH146 analysis covers two recordings from one female fly: a three-odor sequence and a four-odor sequence.
Flies and preparation
The TH preparation was 5137;TH, using the UAS-mCD8::GFP reporter. The GH146 preparation used UAS-GCaMP6f. Female flies were approximately 7–14 days old, starved for 2–6 h and kept on ice under cold anesthesia for 2 h.
Acquisition and odor timing
Each recording spans approximately 132 s and contains 696 five-plane stacks acquired with 488 nm excitation and approximately 33 ms exposure per plane. Analyses use the elapsed time in the original TIFF filenames. A stack interval is approximately 0.190 s.
The table gives team-confirmed stimulus windows. Web recordings follow acquisition time; the source AVI playback rate of 6 frames/s is not the acquisition clock.
| Recording | Odor presentation |
|---|---|
| TH-GFP and GH146: three odors | 20–40 s · isopentyl acetate; 60–80 s · 3-octanol; 100–120 s · 4-methylcyclohexanol |
| GH146: four odors | 20–35 s · ethyl propionate; 55–70 s · benzaldehyde; 90–105 s · 2-heptanone; 120–132 s · valeric acid |
ROI analysis
GH146 regional means were measured from maximum-intensity projections of the five-plane stacks, using the saved ImageJ rectangular ROIs. Whole-image means include the complete field. No bleaching correction or background subtraction was applied to GH146.
TH ΔF/F₀ curves use the supplied ImageJ bleaching-corrected ROI values. The correction used an exponential fit to the tissue-reference ROI named background. Original and corrected measurements remain in the source data.
For each odor, F₀ is the mean fluorescence over the 8 s window ending 2 s before stimulus onset. Local ΔF/F₀ (%) = 100 × (F − F₀) / F₀. The common protocol therefore uses baselines of 10–18, 50–58 and 90–98 s. The four-odor GH146 sequence uses 10–18, 45–53, 80–88 and 110–118 s.
For the full 132 s curves, each ROI uses its own mean over 8 ≤ t < 16 s as F₀, with ΔF/F₀ (%) = 100 × (F − F₀) / F₀. TH uses bleaching-corrected measurements and GH146 uses the original regional means. These full-recording baselines stay fixed; the odor-aligned panels use the local baselines above. All curves are unsmoothed.
Track B
Cloning and plasmid preparation
01 · Colony plates
Colony plates and plasmid preparation
Maintain sample identity from a bacterial colony to the plasmid preparation used downstream.
Three colony-plate photographs and four DNA gels document the bacterial preparation track.
- 01TransformIntroduce the plasmid and retain plate identity
- 02MiniprepRecover plasmid DNA from a selected colony
- 03DNA gelInspect preparation quality with a documented lane map
Transformation and colony selection
The reference procedure uses E. coli DH5α, LB and ampicillin (Protech, supplier) selection (100 µg/mL), with overnight propagation at 37 °C and 180 rpm. The supplied photographs retain the original labels and marked colonies. Sample-specific competent-cell preparation, DNA input, transformation/recovery settings and negative-control records are unavailable. View the three original plate photographs.
Plasmid miniprep
Link each picked colony to its construct and preparation identifier. Record the purification method, culture input, elution volume, concentration and purity. The current source does not identify the miniprep kit or provide a measured yield. The earlier endotoxin-free maxiprep specification remains a separate requirement for mammalian transfection-quality DNA.
02 · Electrophoresis
DNA gel electrophoresis
Inspect DNA preparations alongside the PROTECH Bio-1Kb™ Mass DNA Ladder.
DNA gel electrophoresis used BIOMAX agarose (Hispanagar) and Bio-1Kb™ Mass DNA Ladder (PROTECH; Protech Technology Enterprise Co., Ltd.). The supplied product sheet lists PT-M5-1KB (500 µL) and SA-M5-1KB (50 µL). Undigested plasmids were loaded as paired sample lanes for ORCO, GCaMP6f and ten OR labels. All four gels used this ladder; the figures retain the original left-to-right lane order.
The ladder contains 11 linear fragments: 0.5, 1, 1.5, 2, 2.5, 3, 4, 5, 6, 8 and 10 kb. The manufacturer recommends loading 5 µL per lane, which contains 120 ng of the higher-mass 3 kb reference fragment. Manufacturer product specifications.
The original TIFFs are retained. Each panel uses a contiguous crop and a small rigid rotation to align the wells. The original 0–255 intensity range is unchanged; no lanes are spliced and no bands are retouched. Crop coordinates and rotation angles accompany the figure downloads.
Undigested plasmid migration depends on both conformation and length; the linear ladder therefore provides a migration reference rather than a direct measure of intact plasmid size. Lane maps and source records are retained in the Notebook. [1]
Track C
HEK293 cell experiments and odor testing
C01
Mammalian cell culture
Maintain the mammalian host under defined growth, wash, and detachment conditions.
Executed, details incomplete. HEK293 cultures were imaged; passage, seeding history and a dated operator record for this batch remain unresolved.
Key parameters
| Parameter | Condition |
|---|---|
| Cell line | HEK293 (team-confirmed mammalian experimental host) |
| Growth medium | DMEM (high glucose) + 10 % FBS + 1 % penicillin–streptomycin |
| Incubation | 37 °C, 5 % CO2 |
| Passaging | At ~80 % confluence |
| Detachment | TrypLE Express — 1 mL per T-25, 3 mL per T-75; 37 °C, 3–5 min |
| Wash before detachment | Ca2+/Mg2+-free PBS |
Method
- Grow cells in high-glucose DMEM (Gibco; cat. 11995-065) with 10 % FBS and 1 % penicillin–streptomycin at 37 °C, 5 % CO2.
- Passage at about 80 % confluence.
- Before detachment, wash with Ca2+/Mg2+-free PBS.
- Lift with TrypLE Express (1 mL per T-25, 3 mL per T-75; 37 °C, 3–5 min).
Read the corresponding evidence →
For the printed-well recovery assay, wells contained 100 µL of a nominal 2 × 10⁶ HEK293–GCaMP6f cells/mL suspension and were incubated for 7 h at 37°C and 5% CO₂. Recovery and viability results are reported on Hardware.
C02
Transfection
Deliver OR and Orco plasmids into HEK293 cells and preserve the delivery record for each image set.
Acquired record. HEK293 transfection produced the WL-MIC-01 image set. The source workbook and earlier electroporation specification do not identify its delivery method.
Key parameters
| Parameter | Value |
|---|---|
| Earlier protocol specification | Electroporation in suspension; execution for the imaged samples is unconfirmed |
| Cells per reaction | 1–5 × 106, log phase, > 90 % viability |
| DNA per reaction | ~5 µg total. OR plasmid : Orco plasmid DNA mass ratio 2:1 (not protein stoichiometry) |
| Buffer | Low-conductivity electroporation/nucleofection buffer, 100 µL |
| Recovery | Pre-warmed DMEM + 10 % FBS, no antibiotics |
Method
The earlier specification calls for suspension electroporation using the DNA mass, plasmid DNA mass ratio, buffer volume, and recovery medium in the table. Pulse voltage, pulse number, instrument programme, cuvette, and named buffer are not specified in this source.
The reference study used FuGene HD in HEK293 stable lines [2]. AeroSense retains its own delivery specification and sample record.
03 · Transfection
Assessing transfection and expression
Confirm cassette transcription and Orco/GCaMP function before VOC assays.
Acquired measurements. Fixed-cell endpoint images and Test 1 fluorescence time courses are archived in Results.
Current microscopy record
The supplied HEK293 images were acquired approximately 44–48 h after transfection and before G418 selection. Cells were fixed with 2% paraformaldehyde, without permeabilization, then stained with DAPI. Blue images show DAPI; matching field numbers identify the same field across channels. Exposure was held constant within each channel across groups, and blank calibration was performed. Fixation duration, numerical exposure settings and the complete acquisition log were not supplied.
Fields are technical observations from one batch. Segmentation and thresholds below define selected-object fractions; their interpretation and limits distinguish them from cell-population or functional measurements.
Image analysis, sample-level estimates and limitations
Key parameters
| Parameter | Value |
|---|---|
| mCherry microscopy | ex ~587 nm / em ~610 nm |
| GCaMP6f rest fluorescence | ex ~485 nm / em ~510 nm; dim at rest is expected, not a failure |
| VUAA1 | Synthetic Orco agonist used as a functional positive-control stimulus [3] |
| Western blot | No measurement in this dataset |
| Ionomycin | No measurement in this dataset |
Method
- Record mCherry (ex ~587 nm / em ~610 nm) versus mock / untransfected cells. Red fluorescence is consistent with reporter expression from the OR cassette. It does not report OR folding or plasma-membrane trafficking.
- After an expression gate, compare vehicle versus VUAA1 on cells that should contain Orco–GCaMP6f. VUAA1 is a synthetic Orco agonist used as a functional positive-control stimulus [3] Jones et al., 2011 — full entry in References. . A VUAA1 response supports functionality of the Orco/GCaMP sensing chain. It does not, by itself, prove membrane trafficking.
03 · Image analysis
Endpoint microscopy and image analysis
Reproduce the DAPI-object fluorescence screen reported in Results.
Executed analysis. All 25 fields and all 2,624 eligible objects are retained in the published data tables and reproducibility package.
Source data and experimental context
The host was HEK293, as confirmed by the team. Imaging occurred approximately 44–48 h after transfection and before G418. The team fixed cells with 2% paraformaldehyde (PFA), performed no permeabilization step, and then stained with DAPI. The source workbook contains plate-layout and reagent-calculation records, including DAPI/G418 layouts and lipid-reagent mixtures. These records do not independently establish experimental execution or an assay outcome; the VUAA1/odor grids contain labels without response measurements. The workbook does not supply objective, numerical aperture, camera model, gain, acquisition software, DAPI concentration, fixation duration, or field coordinates. Neither the source TIFF tags nor the workbook establishes these missing acquisition details. The biological preparation and transfection record is linked from Experiments; undocumented settings are not reconstructed from a literature protocol.
Conservative DAPI segmentation
The BLUE plane was extracted from each exported TIFF. A 3 × 3 median filter was used for segmentation only, followed by a common threshold of >128 on the 0–255 export scale and 8-connected labeling. The bottom-right annotation rectangle (x ≥85% of image width and y ≥90% of image height) was excluded equally in every field. Eligible objects had area 200–2200 pixels, bounding-box aspect ratio ≤2.5, and object area/bounding-box area ≥0.4. Objects within 49 pixels of a frame edge or whose measurement neighborhood intersected the annotation exclusion were rejected. No watershed split was imposed on merged objects.
The rules select an image-analysis subset, not a validated nuclear census. Visual inspection does not replace independent boundary annotations: touching nuclei or fragments may pass the filters, and accepted objects may not equal cells. Fixation and absence of permeabilization are confirmed, but DAPI labeling completeness and uniformity were not assessed. Fixed images cannot recover pre-fixation viability. [4]
Reporter measurement and positivity threshold
The full Measurement record adds an explicit statistical specification, all three same-batch blank-field holdouts and the area-versus-object audit. These additions leave the original segmentation and primary thresholds unchanged.
For each eligible object, the signal region contains its nuclear mask plus pixels within 12 pixels, assigned to the nearest blue connected component to limit sharing between adjacent objects. The local background is sampled 24–48 pixels from that object, excluding blue foreground; at least 100 background pixels are required. For RED and GREEN separately, the score is the signal region’s 90th percentile intensity minus the background median. Measurements use the original exported channel values, before the display adjustments.
This nearest-DAPI expansion does not segment cell boundaries. Excluding blue foreground from the background ring also does not exclude every neighboring cell’s cytoplasm. Fluorescent cytoplasm within that ring can raise the estimated background and reduce an object's score, particularly in dense fields.
The primary positivity threshold is the maximum of the three blank-field 99th percentiles of these object scores: RED >16.60 DN and GREEN >11.95 DN. DN denotes the 8-bit exported digital value. Alternative P95 and P99.5 blank thresholds are reported for sensitivity analysis. Thresholds are calibrated and evaluated using the same blank set, so the blank fraction is a calibration diagnostic, not an independent false-positive validation.
The reported fraction is 100 × positive eligible objects / all eligible objects. Condition summaries pool these counts across fields; they do not average percentages with different denominators. “Both” indicates both channel criteria within one object's measurement region. It does not prove two-plasmid delivery, protein localization, or co-expression in a viable cell.
Image processing and presentation
Original TIFF and workbook files were opened read-only and verified by SHA-256 checksums. New figures use fixed linear display windows across all conditions: DAPI 0–255, GREEN 0–120 and RED 0–180; brightfield remains 0–255. The enhanced display can clip bright exported values and is used only for viewing. Original-level views remain available. No sharpening, deconvolution, content removal, generative enhancement or reconstruction of missing detail was applied. The same-field RGB merges combine the original channel positions; no registration correction was imposed. [5] [6]
Web previews are resized with Lanczos resampling and saved as lossless WebP. The original scale bars remain within the full field; they were not used to infer pixel size. Object overlays mark classifications on a dimmed merge for legibility and have full-resolution PNG versions for audit.
Microscopy results and all-field gallery · Scripts, data and source files · Reproduce the control comparison
04 · Pre-selection
Odor-test acquisition and analysis
Reproduce the fluorescence time courses and descriptive well metrics reported in Results.
Executed; critical metadata incomplete. Runs 904–935 comprise the recorded acquisition sequence. This is a manual-addition 96-well pilot, separate from the unrun 384-well specification that follows.
Samples, experimental unit and preparation
The team confirmed that HEK293 cells were measured approximately 72 h after transfection, without G418 selection. Only this plate is currently available. The plate contains one well per condition × stimulus; longitudinal readings are technical timepoints, not replicate wells or independent cultures. The labels identify intended construct conditions and do not establish stable clonal lines. Their identity is not automatically matched to the earlier 44–48 h fixed-cell microscopy cultures.
Columns 1–10 are Or7a, Or19a, Or42a, Or98a, Or22a, Or67a, Or67b, Or59b, Or85b and Or35a, respectively, each with ORCO. Column 11 is ORCO only; column 12 is untransfected HEK293. Rows A–H receive VUAA1, ethyl propionate, isopentyl acetate, 3-octanol, benzaldehyde, 2-heptanone, valeric acid and buffer alone, respectively.
The confirmed assay buffer contains 140 mM NaCl, 5 mM KCl, 2 mM CaCl2, 1 mM MgCl2, 5 mM HEPES-Na and 20 mM glucose, at pH 7.3 and 340 mOsm. The team reports initially aqueous VOC preparations subsequently diluted into this buffer. Final VOC concentrations are unknown. VUAA1 uses DMSO (Protech, inventory supplier); the exact Test 1 bottle and lot, final DMSO fraction and an equivalent DMSO-only vehicle condition are not documented. The source report lists nominal final VUAA1 at 50 µM, without an independently verified preparation record.
Acquisition and additions
The report identifies a BMG CLARIOstar Plus with SMART Control/MARS and a GREINER 96 F-bottom plate. It records bottom optics, a −4.1 mm focal height, LP 504 dichroic, 20 flashes per well per cycle, 0.5 s settling, enhanced dynamic range off, temperature control off and no shaking during the response read. These are source-reported settings. The instrument CSVs directly confirm the 482–16 excitation / 530–40 emission filter labels and 12 s read spacing; they do not contain numeric gain values.
- At the reported initial 100 µL volume, acquire F0a: five readings at 0, 12, 24, 36 and 48 s.
- Add 50 µL plain buffer and acquire the first response, labeled
blank: fifteen readings from 0 to 168 s, every 12 s. The well now contains 150 µL. - Acquire F0b in the same wells: five readings with the same timing.
- Add 50 µL of the reported 4× working preparation, or buffer for row H, to reach 200 µL. Acquire the
testresponse: fifteen readings from 0 to 168 s.
The reported manual sequence uses an eight-channel pipette for columns 1–8 followed by 9–12. Injection timestamps and addition-to-first-read delays are unavailable. Each CSV restarts its time coordinate at its first reading; no interpolation bridges the unmeasured gaps. Header run-start times span approximately 144 minutes. The final recorded response contains a further 168 s of within-run observations. The headers do not encode a timezone.
Independent calculation
Each F0 is the mean of five readings. For each 15-reading response, calculate all thirteen contiguous three-reading means and take their maximum, Fpeak3. This reduces the influence of an isolated point but still selects a maximum; it is a descriptive summary rather than an unbiased effect estimator. The reported peak time is the center of the earliest maximizing window.
ΔFbuffer = (Fpeak3, buffer − F0a) / F0a
ΔFtest = (Fpeak3, test − F0b) / F0b
ΔFcorr = ΔFtest − ΔFbuffer
The two terms use different baselines and well volumes. Their difference is dimensionless; multiplied by 100 it is a difference in percentage points, not a single test-run fluorescence increase. Time-course plots show every measured test point relative to the paired F0b; raw-RFU panels are also provided. No cell-free-background subtraction, gain rescaling, interpolation, fitted response or significance test is applied.
Quality control, exclusions and normalization
The source report identifies gain mismatch in both baseline–response pairs of row B and in the first pair of row C. Coherent shifts across wells are consistent with this concern. All 24 corrected metrics from B/C are excluded; C’s second-pair arithmetic remains available as exploratory data. The remaining 72 corrected metrics are retained conditionally. Without recorded numeric gain, this does not certify the other pairs as gain-matched. Changing detector gain changes the RFU scale, so comparisons require acquisition consistency rather than a guessed correction from untransfected wells. [7]
The independent QC table reports each untransfected stage’s first, mean and median RFU, plus the median across the twelve wells of response-mean / paired-baseline-mean ratios. These formulas are explicit. The original QC field named median_col12_F0a equals the first reading in each row; its published ratio algorithm is unspecified. The original table is retained for audit, not silently relabeled.
All 96 well summaries retain unmasked arithmetic in clearly labeled *_computed columns. The *_retained columns leave excluded values empty, never zero. The source’s VUAA1 normalization is not used: it includes a negative Or7a denominator and otherwise depends on single, unvalidated reference wells. The recalculated six-well 3 SD value is an audit of the source report only; it is not a limit of detection or a response cutoff. No technical-read error bars, confidence intervals or p values are used to imply biological replication.
Acquisition limits
Numeric gains, final VOC concentrations, exact addition delays and a matched DMSO-only condition are unavailable. The source records partial detachment without quantifying it. Test 1 contains one well per condition × stimulus and no independent culture repeats; Results explains the consequences for interpretation.
Read the Test 1 result → · Test 1 executed methods · PDF
Test 2: two-acquisition odor panel
Test 2 contains 72 wells and 1,440 readings. Rows A, D, E, F, G and H contain VUAA1, isopentyl acetate, 2-heptanone, valeric acid, pentyl acetate and DMSO control, respectively. VUAA1 was diluted 1:100; the odor solutions were diluted 1:1,000, with 0.1% DMSO reported. Stock concentrations were not supplied.
Positions 1–12 are Or7a + ORCO, Or19a + ORCO, Or42a + ORCO, Or98a + ORCO, Or22a + ORCO, Or67a + ORCO, Or67b + ORCO, Or59b + ORCO, Or85b + ORCO, Or35a + ORCO, ORCO only and blank. F0 contains five reads over 0–48 s; post-addition contains fifteen reads over 0–168 s, both at 12 s intervals. Each acquisition starts its own clock. Results defines the endpoint correction and VUAA1-reference normalization.
post-addition fluorescence dynamic reanalysis
The supplied post-addition fluorescence workbook retains 1,080 readings: six conditions × twelve channels × fifteen readings, at 12 s intervals from 0 to 168 s. For each well, E is the mean at 0, 12 and 24 s and L is the mean at 144, 156 and 168 s. D = L/E − 1; the vehicle-referenced contrast is C = Dodor − DDMSO for the same channel. We report 100C in percentage points. F0 does not enter this calculation; VUAA1 remains a reference record and is not a denominator or exclusion criterion.
The delivered channel order is Or7a, Or19a, Or42a, Or98a, Or22a, Or67a, Or67b, Or59b, Or85b, Or35a, ORCO only and blank. Analysis follows the workbook’s channel labels and source-cell provenance. DMSO fraction and addition volume were matched as confirmed by the experimenter. Pearson and Spearman correlations are calculated separately for each odorant across the ten tuning ORs, using DoOR consensus minus each receptor’s spontaneous firing rate. The missing Or42a–valeric acid reference is omitted from that correlation only. Controls are retained separately, and all signed contrasts are preserved. Reproduction and validation code · Results.
05 · G418 selection
G418 selection record
Select G418-resistant mammalian lines after transfection.
Completed activity record. The team confirmed the recorded G418 activities as completed; no selection trajectory or endpoint dataset accompanies the record.
The Test 1 pilot at approximately 72 h post-transfection was also measured without G418. Neither Test 1 nor the 44–48 h microscopy batch supplies a post-selection outcome.
The G418 sulfate stock is recorded at 50 mg/mL. The schedule below is the documented literature-informed reference protocol. The acquired 44–48 h microscopy and approximately 72 h Test 1 records both precede G418 selection.
Reference selection schedule
Reference schedule. Day 0: transfection and antibiotic-free recovery. At 24–48 h: G418 at 100 µg/mL for 2 weeks, followed by 150 µg/mL for 1 week and maintenance at approximately 75 µg/mL.
- Day 0 Transfection Suspension transfection. Recovery in DMEM + 10 % FBS, no antibiotics for 24 h.
- 24–48 h G418 100 µg/mL Selection start, 2 weeks. Not added at seeding. Days 1–2: little visible effect; 3–5: bulk death; 10–14: resistant colonies.
- After 2 weeks G418 150 µg/mL Selection raised, 1 week.
- Thereafter G418 ~75 µg/mL Maintenance.
Key parameters
| Stage | G418 concentration | Duration |
|---|---|---|
| Selection start (24–48 h post-transfection) | 100 µg/mL | 2 weeks |
| Selection raised | 150 µg/mL | 1 week |
| Maintenance | ~75 µg/mL | Maintenance phase |
Method
- Before transfection, run a G418 kill curve on untransfected host cells (draft range 0–800 µg/mL) for that cell lot, confluence, and G418 batch. Do not treat the table below as a confirmed effective dose.
- After transfection, allow 24–48 h recovery. Do not add G418 at seeding.
- Start selection at 100 µg/mL G418 for 2 weeks. Refresh selection medium every 2–3 days.
- Raise G418 to 150 µg/mL for 1 week.
- Maintain lines at ~75 µg/mL G418 thereafter.
The reference protocol describes little visible effect at days 1–2, bulk cell death at days 3–5 and resistant colonies at days 10–14. These intervals describe the reference method.
06 · Post-selection
Post-selection odor testing
Evaluate fluorescence responses after a documented selection stage.
Completed activity record. The team confirmed completion of the post-G418 odor test. Its measurements are not archived. Test 1 is a separate pre-selection pilot.
The 384-well design below is a literature-based reference method, separate from both Test 1 and the recorded post-selection activity.
06 · Assay specification
Calcium imaging assay
Measure VOC-evoked GCaMP6f fluorescence in a 384-well plate with a matched vehicle injection.
384-well assay design. For the acquired 96-well Test 1 pilot, see Results.
The calcium-response assay was adapted from Zboray et al. [2] Zboray et al., 2023 — full entry in References. . AeroSense-specific construct, transfection, and validation procedures are documented separately above. Plate geometry, buffer volumes, and the two-round injection sequence below follow that published assay design.
Key parameters
| Parameter | Value |
|---|---|
| Plate format | Black-wall, clear-bottom 384-well |
| Seeding | 12,000–18,000 cells per well, 48–72 h before measurement |
| Assay volume | 20 µL assay buffer per well |
| Vehicle injection | 7 µL plain VOC buffer |
| Odorant injection | 9 µL of VOC at 4× final concentration |
| Final well volume | 36 µL (20 + 7 + 9) |
| VOC concentrations | 1, 10 and 100 µM final |
| VUAA1 reference | 50 µM on every plate for every cell line |
| Instrument | Multimode plate reader with a two-channel injector (model not specified) |
| Readout | ex 485 nm / em 535 nm |
| Sampling | F0 for 5 s; then 30 s at 1 read/s after each injection |
Buffers
| Component | Assay buffer | VOC buffer |
|---|---|---|
| NaCl | 140 mM | 140 mM |
| KCl | 5 mM | 5 mM |
| CaCl2 | 2 mM | 2 mM |
| MgCl2 | 1 mM | 1 mM |
| HEPES-Na | 5 mM | 5 mM |
| Glucose | 20 mM | 31 mM |
| DMSO | — | 0.1 % |
| pH / osmolarity | 7.3 / 340 mOsm | 7.3 / 340 mOsm |
Method
Read on a multimode plate reader with a two-channel injector, ex 485 nm / em 535 nm.
- Seed cells at 12,000–18,000 per well in black-wall, clear-bottom 384-well plates, 48–72 h before measurement.
- On the day of measurement, replace growth medium with 20 µL assay buffer per well.
- Round 1 — vehicle: record baseline F0 for 5 s, inject 7 µL plain VOC buffer, record 30 s at 1 read/s.
- Round 2 — odorant: into the same wells, inject 9 µL of VOC at 4× final concentration, same recording protocol.
- Layout: one compound per row; up to eight cell lines per 24-column plate. Use three wells per cell line and measure each VOC–OR pair on at least three separate days.
The volumes are self-consistent: 20 + 7 = 27 µL, then + 9 = 36 µL; 9/36 = 1/4, so a 4× stock dilutes exactly to 1× final. The vehicle injection simultaneously establishes the volume for the second round and measures the solvent and injection artefact alone.
Well volume and two-round sequence
Three well volumes. First: 20 microlitres assay buffer. Second: 20 plus 7 microlitres vehicle equals 27. Third: 27 plus 9 microlitres of 4 times VOC equals 36. Nine divided by 36 equals one quarter, so a 4 times stock becomes 1 times final.
9 / 36 = 1/4 → 4× stock dilutes to 1× final
- Round 1 F0 5 s Inject 7 µL vehicle Record 30 s at 1 read/s
- Round 2 Same wells Inject 9 µL 4× VOC Record 30 s at 1 read/s
20 µL assay buffer + 7 µL vehicle = 27 µL; then + 9 µL of 4× VOC = 36 µL. 9/36 = 1/4, so a 4× stock becomes 1× final concentration. Geometry adapted from Zboray et al. [2].
Reagent preparation VUAA1 stock and VOC handling
VUAA1. Dissolve in DMSO to a 50 mM stock (18.4 mg/mL). Aliquot and store at −20 °C protected from light. Prepare the 4× working dilution fresh in VOC buffer.
VOCs. Dilute directly in VOC buffer, or first in DMSO for water-insoluble compounds. Prepare working solutions fresh daily. Store concentrated stocks in amber glass with PTFE-lined caps.
Reproducibility metadata
- Technical replicates
- Three wells per cell line, one compound per row, triplicate
- Day replicates
- Each VOC–OR pair measured on at least three separate days
- Instrument
- Multimode plate reader with a two-channel injector (model not specified)
- Excitation / emission
- 485 nm / 535 nm
- Sampling rate
- 1 read/s for 30 s after each injection; F0 recorded for 5 s before injection
- Analysis
- R then Rnorm; see Data analysis
06 · Analysis specification
Data analysis
Convert raw fluorescence into a VUAA1-normalised relative response for each receptor–stimulus pair.
Reference analysis. These equations describe the cited relative-response method. The acquired Test 1 data use the Test 1 analysis; fixed-cell images use the microscopy pipeline.
Relative-response definition, the ≤ 0.05 floor, plate/day VUAA1 mean normalisation, and Grubbs’ outlier removal follow Zboray et al. [2] Zboray et al., 2023 — full entry in References. . Zboray et al. denote the relative response R as ΔF. This page uses R and Rnorm to avoid mixing ΔF with ΔF/F0.
Key parameters
| Parameter | Value |
|---|---|
| Raw fluorescence | Plate-reader intensity at ex 485 nm / em 535 nm |
| F0 | Baseline recorded for 5 s before the relevant injection |
| Fmax | Peak fluorescence in the post-injection window |
| Raw relative response | R = (Fmax − F0) / F0 |
| VUAA1-normalised response | Rnorm = RVOC / mean(RVUAA1, corresponding cell line, same plate/day) |
| Zero floor | Rnorm ≤ 0.05 treated as zero, following [2] |
| Outliers | Removed by Grubbs’ test, following [2] (α not specified here) |
Method
- Record raw fluorescence at ex 485 nm / em 535 nm. Take F0 as the 5 s baseline before the relevant injection.
- Take Fmax as the peak in the post-injection window. Compute R = (Fmax − F0) / F0.
- Normalise each VOC response to the mean 50 µM VUAA1 response of the corresponding OR cell line measured on the same plate/day: Rnorm = RVOC / mean(RVUAA1, corresponding cell line, same plate/day). Do not use a same-well VUAA1 injection as the denominator.
- Following [2], treat Rnorm ≤ 0.05 as zero, and remove outliers by Grubbs’ test.
R = (Fmax − F0) / F0
Rnorm = RVOC / mean(RVUAA1, line, plate/day)
Reproducibility metadata
- Excitation / emission
- 485 nm / 535 nm
- Sampling
- F0 for 5 s before the relevant injection
- Analysis method
- R, then plate/day mean 50 µM VUAA1 Rnorm; ≤ 0.05 floor and Grubbs’ test following [2]
- Downstream
- Multi-receptor vectors are intended for a sparse-coding classifier based on [8] Dasgupta et al., 2017 — full entry in References. . That classifier is not specified on this page. See Model.
Safety
All work is contained. E. coli DH5α is Risk Group 1 (BSL-1); HEK293 is handled at BSL-2 in a Class II biosafety cabinet. Aromatic VOC ligands are flammable and toxic and are handled in a fume hood per their Safety Data Sheets. Nothing is released outside the laboratory.
Full biosafety, chemical handling, and training records are on the Safety & Security page.